Why it was recalled

The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.

FDA root cause category: Software design

What the company is doing

On September 8, 2022, the company notified affected customers via phone call and emailed letter. Customers were instructed to do the following: 1. If a high priority Pump Problem alarm occurs, the LVP should be powered down by holding the power button for 10 seconds then releasing the power button so that the screen goes dark, then pressing the power button again to re-start the pump. Alternatively, if you have another LVP available, you may use it to resume infusion. 2. Inform potential users of the product within your organization of this notification. If your facility further distributes or transfers products amongst satellite sites or other locations, please disseminate this information accordingly. Fresenius Kabi has updated the software to rectify the condition. The updated software is currently available. Please schedule a service call at your earliest convenience to have it installed. Customer support: [email protected]; 1-978-775-8100

Recall details

Recall number
Z-0002-2023
Status
Terminated
Initiated by the company
Sep 8, 2022
Posted by the FDA
Oct 6, 2022
Recalling firm
Fresenius Kabi USA, LLC
Quantity
4 instances
Product code
PHC
FDA submissions
K183311
Distribution
Domestic distribution to NJ and WI.

Product description

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004 version 5.2.0

Code information (models, lots, serial numbers)

UDI-DI: 00811505030122 version 5.2.0

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.