Berkeley VC-10 Vacuum Curettage System, Model No. VC-10.… recallZ-0078-2023 · Olympus Corporation of the Americas · initiated Sep 16, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Update to the Instruction for Use (IFU): Olympus is adding a new warning against the use of accessories which are not listed in the IFU. There have been reported adverse events involving the use of collection canisters which were not designed for use with the VC-10, which can lead to serious injuries as it does not fit properly with the VC-10.
FDA root cause category: Labeling design
What the company is doing
On September 16, 2022, the recalling firm notified affected customers through "URGENT Medical Device Field Corrective Action" letters. Customers were informed that the following new warning will be added to the IFU: Do not use replacement parts or accessories that are not compatible with the VC-10 system. Use of noncompatible accessories could lead to reduced system functionality and/or patient injury such as perforation, bleeding, and may require additional intervention. Refer to Section 11 for all accessories compatible with the VC-10 system. Customers should notify their users about risks associated with using non-compatible collection canisters and/or accessories and remind them to follow the list of compatible accessories outlined in Section 11 of the IFU. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. " Go to https://olympusamerica.com/recall. Enter the recall number 0414 " Enter your name and mailing address. Olympus will supply a copy of the revised IFU at a later date If you have distributed these devices outside your facility, please notify your customers of this matter immediately. If you have any questions, call (647) 999-3203 or email: [email protected] for any additional information or support concerning this matter.
Recall details
- Recall number
- Z-0078-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 16, 2022
- Posted by the FDA
- Oct 12, 2022
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 4184 US; 191 OUS
- Product code
- HHK
- FDA submissions
- K171440
- Distribution
- Domestic distribution nationwide. Foreign distribution to Canada and Japan.
Product description
Berkeley VC-10 Vacuum Curettage System, Model No. VC-10. For transcervical aspiration of the uterine cavity.
Code information (models, lots, serial numbers)
UDI-DI: 00821925010475 All units are included.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BERKELEY VC-10Gyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)1 recall8 reports
Other devices under the same FDA submission
6 devices cleared under K171440 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BERKELEY SAFE-TOUCH 003853-902Gyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)8 reports
- Gyrus ACMI 23116Gyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)3 reports
- SAFETOUCH 003984-901Gyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)2 reports
- Gyrus ACMI 54298Gyrus ACMI, LLC · CURETTE, SUCTION, ENDOMETRIAL (AND ACCESSORIES)2 reports
- Gyrus ACMI 003997-901Gyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)1 report
- Berkeley Vacurette 21593AGyrus ACMI, LLC · Curette, Suction, Endometrial (And Accessories)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page