Skip to content

Speed recallZ-0099-2023 · Synthes (USA) Products LLC · initiated Sep 27, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

FDA root cause category: Error in labeling

What the company is doing

On September 27, 2022, the firm notified affected customers via urgent medical Device Recall letters. Customers were instructed to examine their inventory to determine if they have affected product and to quarantine them immediately - Customers should not use the recalled products. Customers should contact their DePuy Synthes Sales Consultant or call the customer support services at [email protected] to coordinate the return/credits of the subject products.

Recall details

Recall number
Z-0099-2023
Status
Open, Classified
Initiated by the company
Sep 27, 2022
Posted by the FDA
Oct 18, 2022
Recalling firm
Synthes (USA) Products LLC
Quantity
94 US; 30 OUS
Product code
JDR
FDA submissions
K142292
Distribution
US Nationwide distribution.

Product description

SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 00810633020166 Lot/Expiry MSE220445 7/1/2027 MSE210184 8/1/2026

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 90918).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page