Why it was recalled
The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.
FDA root cause category: Error in labeling
What the company is doing
On September 27, 2022, the firm notified affected customers via urgent medical Device Recall letters. Customers were instructed to examine their inventory to determine if they have affected product and to quarantine them immediately - Customers should not use the recalled products. Customers should contact their DePuy Synthes Sales Consultant or call the customer support services at [email protected] to coordinate the return/credits of the subject products.
Recall details
- Recall number
- Z-0101-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 27, 2022
- Posted by the FDA
- Oct 18, 2022
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 196
- Product code
- JDR
- FDA submissions
- K191463
- Distribution
- US Nationwide distribution.
Product description
DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001
Code information (models, lots, serial numbers)
UDI-DI (GTIN): 00810633022047 Lot/Expiry MHC200114 2/9/2026 MHC220117 3/2/2027
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N/A StandardSynthes GmbH · Staple, Fixation, Bone2 recalls16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 90918).
- Z-0098-2023 · Open, ClassifiedSPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kit, Product No. SE-2020-10. Used with DePuy Synthes SPEEDSHIFT Continuous Compression Staple Implants for bone fixation and managemen…
- Z-0099-2023 · Open, ClassifiedSPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512
- Z-0100-2023 · Open, ClassifiedSPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page