Skip to content

LineSider recallZ-0421-2023 · Integrity Implants Inc. · initiated Aug 3, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integrity Implants Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for intraoperative screw head separation at weld location.

FDA root cause category: Under Investigation by firm

What the company is doing

Integrity Implants notified consignees on about 08/03/2022 via emailed letter. The letter instructed consignees to examine inventory and quarantine any affected product, arrange for the return of affected products, and complete and return the response form.

Recall details

Recall number
Z-0421-2023
Status
Terminated
Initiated by the company
Aug 3, 2022
Posted by the FDA
Dec 5, 2022
Recalling firm
Integrity Implants Inc.
Quantity
50 screws
Product code
NKB
FDA submissions
K203367
Distribution
US distribution to California, Indiana, New York, Oklahoma, Washington DC

Product description

LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as tray replenishments. Used in spinal fixation surgery.

Code information (models, lots, serial numbers)

Part/Model Number LS-N6SG6545; UDI: (01)00818613027815(10)10210067; Lot Number: 10210067

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

68 devices cleared under K203367 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates