SPiN Thoracic Navigation System recallZ-1103-2023 · Olympus Corporation of the Americas · initiated Dec 13, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A grainy image is observed when using a SPiN Vision Single-Use Flexible Bronchoscope (models INS-7100 or INS-7130) with the SYS-4000 system due to an incorrect DVI cable supplied may result in prolonged procedure/surgery due to additional time to troubleshoot the device
FDA root cause category: Under Investigation by firm
What the company is doing
Olympus issued URGENT - Medical Device Field Corrective Action letter on 12/13/22. Letter states reason for recall, health risk and action to take: 1. Notify your users about this issue and stop using the single-use bronchoscopes (models INS-7100 or INS-7130) with SYS-4000 system until cable is replaced. Your users can continue to use the SYS-4000 using a reusable bronchoscope, with its respective video processor, without affecting their electromagnetic navigation (EMN) procedural experience. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter the recall number 0418 Provide your contact information as indicated in the portal. An Olympus Field Service Representative will contact you to schedule a site visit to replace the DVI cable. 3. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints, including patient injuries and adverse events, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 to report complaints. Contact (647) 999-3203 or at [email protected] for any additional information or support.
Recall details
- Recall number
- Z-1103-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 13, 2022
- Posted by the FDA
- Feb 7, 2023
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 13 units
- Product code
- JAK
- FDA submissions
- K170023
- Distribution
- FL, GA, IN, MD, MS, NY, OH, PA, RI, SC, VA
Product description
SPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
Code information (models, lots, serial numbers)
UDI: 00815686020101 Serial Numbers: 85330200430, 87896220519, 87899220524, 87379210921, 87835220505, 87832220502, 87767220420, 87777220415, 87827220427, 87736220412, 87891220517, 87864220509, 87738220412
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SPiN Thoracic Navigation System, 120V, 60Hz SYS-4000Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls1 death28 reports
Other devices under the same FDA submission
11 devices cleared under K170023 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Always-On Tip Tracked Triple Needle Brush, 12mm L, 1.8mm OD INS-5300Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls16 reports
- SPiN Access Catheter, 90 Degree For Use With Olympus Scopes INS-5900Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall3 reports
- SPiN Access Catheter, 180 Degrees For Use With Olympus Scopes INS-5905Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall2 reports
- 0.016" Stylet, 1235mm L INS-5411Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall1 report
- SPiN Access Catheter, 90 Degrees For Use With Olympus 190 Scopes INS-5910Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall1 report
- SPiN Access Catheter, 180 Degrees For Use With Olympus 190 Scopes INS-5915Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall1 report
- SPiN Perc Biopsy Needle Guide Kit - 20cm INS-5620Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall0 reports
- SPiN Perc Localization Needle 19ga x 105mm, 1cm marking zone INS-5610Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall0 reports
- SPiN Perc Localization Needle 19ga x 105mm, 2cm marking zone INS-5611Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall0 reports
- SPiN Access Catheter, 90 Degrees For Use With Pentax Scopes INS-5920Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall0 reports
- SPiN Access Catheter, 180 Degrees For Use With Pentax Scopes INS-5925Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page