SPiN Access Catheter, 180 Degrees For Use With Olympus Scopes INS-5905Recalls & reported problems
Veran Medical Technologies Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815686021269), received between Apr 13, 2021 and Feb 10, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
14 other recalls cover devices cleared under the same FDA submission (K170023); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 10, 2023Class IIOpen, Classified
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Z-2660-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
- Insufficient Device Problem Information1
- Material Separation1
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2022Material Separation
OUT OF THE BOX, THE TELESCOPING SHAFT CONNECT TO THE HANDLE OF THE SPIN ACCESS CATHETER BROKE OFF WHEN PUTTING THE ADAPTER, GUIDEWIRE, AND CATHETER TOGETHER. A NEW INS-5905 WAS OPENED AND THE PROCEDURE WAS COMPLETED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 13509737 ↗MalfunctionApr 13, 2021Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
THE PHYSICIAN WAS UNABLE TO PUSH THE CATHETER THROUGH THE ADAPTER. THERE WAS A SNAP AND A CREASE FOR FORMED. THE PHYSICIAN WAS UNABLE TO FEED INSTRUMENTS DOWN THE CATHETER. A SECOND CATHETER WAS OPENED AND BROKE IN WHILE THE STAFF WAS ATTEMPTING TO FEED IT THROUGH THE SCOPE. A THIRD CATHETER WAS OPENED AND THE PROCEDURE CONTINUED AS PLANNED. THE BROKEN INSTRUMENT WAS NOT USED WITH A PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11655956 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170023. One submission often covers many models, so check the product and lot details in each record.
- Z-2661-2023 · Aug 10, 2023 · Open, ClassifiedVeran : Percutaneous Models: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" INS-5611 SPiN Perc Localization Needle - 2cm, 5/Bx INS-5610 SPiN
- Z-2659-2023 · Aug 10, 2023 · Open, ClassifiedVeran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50H
- Z-2662-2023 · Aug 10, 2023 · Open, ClassifiedVeran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC INS-5411 0.016
- Z-2664-2023 · Aug 10, 2023 · Open, ClassifiedVeran : Endobronchial Always on Track Models: INS-0392 AOTT 21ga Needle, 12mm L, 1.8mm OD, 5/BX INS-0382 AOTT 19ga Needle, 12mm L, 1.8 OD, 5/Bx INS-0372 AOTT Serrated For
- Z-2663-2023 · Aug 10, 2023 · Open, ClassifiedVeran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "
- Z-2262-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anatomical structures where computed tomograph
- Z-2268-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) 22ga SPiN Flex Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where co
- Z-2265-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) 19ga ANSO Histology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures whe
- Z-2266-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures wher
- Z-2263-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8 mm OD, Oval Cup-Intended for use in clinical interventions and for anatomical structures where computed tomo
- Z-2267-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where comp
- Z-2269-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) SPiN Xtend 2.0mm OD, 21ga Intended for use in clinical interventions and for anatomical structures where computed tomogr
- Z-2264-2023 · May 15, 2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical interventions and for anatomical structures where computed t
- Z-1103-2023 · Dec 13, 2022 · Open, ClassifiedSPiN Thoracic Navigation System Model: SYS-4000 (SYS-4000D)
Other System, X-Ray, Tomography, Computed devices
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
- Spectral CTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed11 recalls3 reports
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
- Brilliance iCTPhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed8 recalls6 reports
- Revolution Apex EliteGE Medical Systems, L.L.C. · System, x-ray, tomography, computed8 recalls0 reports
- Brilliance CT Big BorePhilips Medical Systems Nederland B.V. · System, x-ray, tomography, computed7 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page