SynCardia Total Artificial Heart System (70CC): 500101-001 recallZ-1022-2023 · SynCardia Systems LLC · initiated Nov 2, 2022
Class II Open, Classified All recalls posted in January 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from SynCardia Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to tears reported tears in the cannula due to wear and tear stress.
FDA root cause category: Device Design
What the company is doing
On 11/02/2022, the firm send an "Urgent Medical Device Correction" Letter via email alerting customers that there is the potential for hole or tear that may occur in the pneumatic cannulae of the temporary Total Artificial Heart (TAH-t). Customers are instructed to: Prior to a hole/tear occurring, train patients to properly inspect the cannulae for holes/ tears and to recognize the signs of a tear/hole. If a hole/tear occurs, the patient should immediately call the hospital to arrange for repair. For Clinicians Only: While SynCardia has not validated the below instructions, there have been literature articles[i][ii] on this method to repair a cannula tear by removing the damaged cannula segment. The repair requires a backup Companion 2 or Freedom portable driver, a secondary driver driveline connector, zip ties, and high-leverage cable cutters. The repair should be implemented quickly by at least two operators following the instructions below. One operator will prepare and operate the backup driver. A second operator will cut and reconnect the cannula. 1. Inspect cannulae for any sign of a tear, hole, or cut, which can be visually detected or detected by listening for a hissing sound or whooshing of air. 2. Prepare a backup driver with an appropriate secondary connector attached to the affected driver driveline. Power on the selected backup driver and keep it on standby for the next repair steps. 3. Using a high-leverage cable cutter cut the cannula through the coil wire slightly distal to the tear to minimize the loss of cannula length. Insert a secondary connector barb (connected to the powered backup driver) into the cut cannula. Secure the cannula to the new connector with zip ties. 4. Disconnect the other undamaged cannula from the primary driver and reconnect it to the backup driver. 5. Change backup driver from standby to run with the correct settings for the patient. For questions contact Customer Support Manager at 1-866-480-1122. In addition to custome
Recall details
- Recall number
- Z-1022-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Nov 2, 2022
- Posted by the FDA
- Jan 27, 2023
- Recalling firm
- SynCardia Systems LLC
- Quantity
- 1,732 systems
- Product code
- LOZ
- FDA submissions
- P030011
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, FL, IA. IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, PA, TN, TX, UT, VA, WA, and WI. The countries of Australia, Austria, Canada, Croatia, Finland, France, Germany, Greece, Israel, Italy, Kuwait, Lebanon, Lithuania, Macedonia, Poland, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Product description
SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000
Code information (models, lots, serial numbers)
SynCardia Total Artificial Heart System (70CC): 500101-001 SynCardia Total Artificial Heart System (50CC): 570500-000
Other devices under the same FDA submission
9 devices cleared under P030011 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SynCardia Freedom® Driver 595000-001Syncardia Systems, LLC · Artificial Heart1 recall7 deaths809 reports
- SynCardia Companion 2 Driver 397002-001Syncardia Systems, LLC · Artificial Heart2 deaths414 reports
- SynCardia temporary Total Artificial Heart System 500101-001Syncardia Systems, LLC · Artificial Heart141 deaths215 reports
- SynCardia Freedom® Onboard Battery 295025-001Syncardia Systems, LLC · Artificial Heart62 reports
- SynCardia 50cc TAH-t, Clinical 570500-001Syncardia Systems, LLC · Artificial Heart1 recall5 deaths5 reports
- SynCardia Companion Hospital Cart 397003-001Syncardia Systems, LLC · Artificial Heart9 reports
- SynCardia Companion Rechargeable LI-ION Battery 293001-001Syncardia Systems, LLC · Artificial Heart8 reports
- SynCardia Companion Driver Caddy 397001-001Syncardia Systems, LLC · Artificial Heart3 reports
- SynCardia Hand Pump 397004-001Syncardia Systems, LLC · Artificial Heart2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.