Skip to content

Sterile Procedural Trays, labeled as the following recallZ-1068-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 17, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

FDA root cause category: Device Design

What the company is doing

Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.

Recall details

Recall number
Z-1068-2023
Status
Open, Classified
Initiated by the company
Nov 17, 2022
Posted by the FDA
Feb 6, 2023
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
622 cases (1,619 units)
Product code
OFF
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Product description

Sterile Procedural Trays, labeled as the following: a. PORT/TRACH INSERTION CDS-3 b. RESIDENT DERM SURG PACK c. THROMBECTOMY PACK-LF d. AV FISTULA PACK e. AV SHUNT PACK-LF f. RR-VP SHUNT PACK-LF g. VEIN HARVESTING PACK-LF h. AV FISTULA PACK-LF i. PK, GEN-BARIATRIC j. VEIN PACK k. VENOUS ACCESS PACK-LF l. VENOUS ACCESS PACK m. VASCULAR ACCESS PACK n. ARTERIOGRAM PACK o. AV FISTULA/GRAFT INSER PK p. VEIN ABLATION PACK q. PORT PACK r. MINOR VASCULAR s. FISTULA PACK t. AV FISTULA-LF u. AV FISTULA v. PORT VENOUS ACCESS w. AV FISTULA

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a. PORT/TRACH INSERTION CDS-3, Item Number: CDS983004D, Case UDI/GTIN: 40889942000464, Unit UDI/GTIN: 10889942000463, Lot Number: 22IBF922,22IBE764; b. RESIDENT DERM SURG PACK, Item Number: DYNDA2997, Case UDI/GTIN: 40195327235353, Unit UDI/GTIN: 10195327235352, Lot Number: 22JBR611; c. THROMBECTOMY PACK-LF, Item Number: DYNJ0160348B, Case UDI/GTIN: 40889942211594, Unit UDI/GTIN: 10889942211593, Lot Number: 22IDB621; d. AV FISTULA PACK, Item Number: DYNJ01966N, Case UDI/GTIN: 40193489383769, Unit UDI/GTIN: 10193489383768, Lot Number: 22IMA429; e. AV SHUNT PACK-LF, Item Number: DYNJ0781844W, Case UDI/GTIN: 40193489226530, Unit UDI/GTIN: 10193489226539, Lot Number: 22IMG652; f. RR-VP SHUNT PACK-LF, Item Number: DYNJ0785917M, Case UDI/GTIN: 40193489412827, Unit UDI/GTIN: 10193489412826, Lot Number: 22GMH382; g. VEIN HARVESTING PACK-LF, Item Number: DYNJ0826732I, Case UDI/GTIN: 40193489377973, Unit UDI/GTIN: 10193489377972, Lot Number: 22IMF839; h. AV FISTULA PACK-LF, Item Number: DYNJ0826753O, Case UDI/GTIN: 40193489384292, Unit UDI/GTIN: 10193489384291, Lot Number: 22JMH548,22IMC908; i. PK, GEN-BARIATRIC, Item Number: DYNJ33890B, Case UDI/GTIN: 40193489213455, Unit UDI/GTIN: 10193489213454, Lot Number: 22JBV659,22HBW898; j. VEIN PACK, Item Number: DYNJ36309F, Case UDI/GTIN: 40195327078035, Unit UDI/GTIN: 10195327078034, Lot Number: 22IMC227; k. VENOUS ACCESS PACK-LF, Item Number: DYNJ36725D, Case UDI/GTIN: 40195327148714, Unit UDI/GTIN: 10195327148713, Lot Number: 22HBO043; l. VENOUS ACCESS PACK, Item Number: DYNJ42694A, Case UDI/GTIN: 40193489983341, Unit UDI/GTIN: 10193489983340, Lot Number: 22JBK167,22IBJ914; m. VASCULAR ACCESS PACK, Item Number: DYNJ45071C, Case UDI/GTIN: 40889942943303, Unit UDI/GTIN: 10889942943302, Lot Number: 22IBO914; n. ARTERIOGRAM PACK, Item Number: DYNJ53095A, Case UDI/GTIN: 40195327148721, Unit UDI/GTIN: 10195327148720, Lot Number: 22IBM491,22HBY446; o. AV FISTULA/GRAFT INSER PK, Item Number: DYNJ57152, Case UDI/GTIN: 40889942721888, Unit UDI/GTIN: 10889942721887, Lot Number: 22IBT472; p. VEIN ABLATION PACK, Item Number: DYNJ62889B, Case UDI/GTIN: 40195327030958, Unit UDI/GTIN: 10195327030957, Lot Number: 22JDB544; q. PORT PACK, Item Number: DYNJ64675, Case UDI/GTIN: 40193489212144, Unit UDI/GTIN: 10193489212143, Lot Number: 22IBP112; r. MINOR VASCULAR, Item Number: DYNJ66148A, Case UDI/GTIN: 40195327096329, Unit UDI/GTIN: 10195327096328, Lot Number: 22JBL900,22HBP630,22HBG264; s. FISTULA PACK, Item Number: DYNJ66502, Case UDI/GTIN: 40193489358507, Unit UDI/GTIN: 10193489358506, Lot Number: 22HBO053; t. AV FISTULA-LF, Item Number: DYNJ901747K, Case UDI/GTIN: 40193489482165, Unit UDI/GTIN: 10193489482164, Lot Number: 22JDC010,22JDA876; u. AV FISTULA, Item Number: DYNJ903417F, Case UDI/GTIN: 40193489418218, Unit UDI/GTIN: 10193489418217, Lot Number: 22KBF261,22KBG353; v. PORT VENOUS ACCESS, Item Number: DYNJ903426F, Case UDI/GTIN: 40193489418447, Unit UDI/GTIN: 10193489418446, Lot Number: 22JBR462,22JBB174,22HBS253; w. AV FISTULA, Item Number: DYNJ903726B, Case UDI/GTIN: 40193489966849, Unit UDI/GTIN: 10193489966848, Lot Number: 22KBD044,22IBO370,22HBX959;

Devices named in this recall

23 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 91386).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page