Sterile Procedural Trays, labeled as the following recallZ-1088-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 17, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
FDA root cause category: Device Design
What the company is doing
Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.
Recall details
- Recall number
- Z-1088-2023
- Status
- Open, Classified
- Initiated by the company
- Nov 17, 2022
- Posted by the FDA
- Feb 6, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 6,782 cases (15,253 units)
- Product code
- OJG
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
Product description
Sterile Procedural Trays, labeled as the following: a. AV SHUNT CDS b. ZALE V-P SHUNT CDS c. NEURO CDS-LF d. MAJOR NEURO e. NEURO PACK-LF f. CUSTOM NEURO PACK-LF g. NEURO PACK-LF h. NEURO PACK-LF i. NEURO PACK-LF j. NEURO PACK-LF k. NEURO/LAMINECTOMY BASIN PK-LF l. U H S NEURO PACK SC m. NEURO PACK-CELEBRATION n. NEURO PACK o. NEURO BASIN PACK-LF p. NEURO PACK q. MAJOR NEURO PK-LF r. NINOR NEURO PK-LF s. NEURO VP SHUNT/VAGAL NERVE-RF t. NEURO PACK-LF u. NEURO PACK v. CRANIOTOMY w. PEDIATRIC NEURO PACK x. NEURO MATRIX PACK y. ACOUSTIC NEUROMA PACK-LF z. MAJOR NEURO PACK-LF aa. SCRIPPS SW NEURO PACK-LF bb. NEURO PACK cc. GENERAL PACK dd. NEURO PK ee. NEURO PACK ff. NEURO PACK gg. NEURO PACK hh. NEURO PACK-LF ii. NEURO PACK jj. NEURO PACK kk. NEURO PACK ll. NEURO PACK mm. CRANIOTOMY PACK nn. PK, NEURO-MINOR oo. NEURO PACK pp. NEURO PACK qq. NEURO PACK rr. PK, NEURO-MINOR ss. NEURO BASIC tt. NEURO BASIC uu. CRANIOTOMY PACK (CRSJE)227-LF vv. NEURO PACK ww. NEURO PACK xx. NEURO CRANI PACK yy. NEURO PACK zz. NEURO TRAY-LF aaa. NEURO PACK bbb. WH NEURO PACK ccc. RFT ISC-N NEURO PACK ddd. NEURO-LF eee. NEURO PACK fff. NEURO LUMBAR PACK ggg. NEURO FUSION PACK-LF hhh. NEURO CERVICAL PACK iii. NEURO PACK jjj. VP SHUNT PACK kkk. NEURO BASIC PACK lll. NEURO PACK mmm. CHRISTUS CHILDRENS NEURO PK nnn. CHRISTUS CHILDRENS VP SHUNT PK ooo. RF NEURO-SHUNT PACK ppp. NEURO PACK qqq. NEURO PACK rrr. NEURO ST DAVIDS MEDICAL CTR sss. GIO NEURO PACK ttt. NEURO KIT uuu. DISCECTOMY PACK vvv. NEURO PACK www. BASIC NEURO PACK-LF xxx. NEURO PACK yyy. NEURO PACK zzz. NEURO aaaa. NEURO bbbb. FHCW NEURO PACK cccc. NEURO FERREE dddd. NEURO eeee. NEURO PACK CW ffff. NEURO gggg. FHT NEURO hhhh. NEURO BASIC iiii. NEURO LUMBAR jjjj. NCH NEURO kkkk. TAVI llll. CHN OR NEURO mmmm. NEURO PACK-LF nnnn. NEURO-CRANI-LF
Code information (models, lots, serial numbers)
Show the full code information (13k characters)
a. AV SHUNT CDS, Item Number: CDS983098D, Case UDI/GTIN: 40195327191307, Unit UDI/GTIN: 10195327191306, Lot Number: 22IBD957; b. ZALE V-P SHUNT CDS, Item Number: CDS983186J, Case UDI/GTIN: 40195327105625, Unit UDI/GTIN: 10195327105624, Lot Number: 22IBT194; c. NEURO CDS-LF, Item Number: CDS985289, Case UDI/GTIN: 40888277050502, Unit UDI/GTIN: 10888277050501, Lot Number: 22JBH363,22IBF823,22IBD861; d. MAJOR NEURO, Item Number: CDS985548N, Case UDI/GTIN: 40195327033256, Unit UDI/GTIN: 10195327033255, Lot Number: 22JBR391,22JBI641,22GBG964; e. NEURO PACK-LF, Item Number: DYNJ0038440B, Case UDI/GTIN: 40193489506885, Unit UDI/GTIN: 10193489506884, Lot Number: 22IME936; f. CUSTOM NEURO PACK-LF, Item Number: DYNJ0190156F, Case UDI/GTIN: 40193489357531, Unit UDI/GTIN: 10193489357530, Lot Number: 22JMA809; g. NEURO PACK-LF, Item Number: DYNJ0367094L, Case UDI/GTIN: 40889942123859, Unit UDI/GTIN: 10889942123858, Lot Number: 22IMI193,22HMF079; h. NEURO PACK-LF, Item Number: DYNJ0374024J, Case UDI/GTIN: 40889942902119, Unit UDI/GTIN: 10889942902118, Lot Number: 22HBW509,22HBX254,22HBV379; i. NEURO PACK-LF, Item Number: DYNJ04596K, Case UDI/GTIN: 40193489301343, Unit UDI/GTIN: 10193489301342, Lot Number: 22JMA062,22IMB264; j. NEURO PACK-LF, Item Number: DYNJ04671Q, Case UDI/GTIN: 40888277743312, Unit UDI/GTIN: 10888277743311, Lot Number: 22HMC512; k. NEURO/LAMINECTOMY BASIN PK-LF, Item Number: DYNJ0781823K, Case UDI/GTIN: 40193489919197, Unit UDI/GTIN: 10193489919196, Lot Number: 22JMB723,22HMD049; l. U H S NEURO PACK SC, Item Number: DYNJ07952J, Case UDI/GTIN: 40889942607953, Unit UDI/GTIN: 10889942607952, Lot Number: 22JBO711; m. NEURO PACK-CELEBRATION, Item Number: DYNJ14032, Case UDI/GTIN: 40080196777953, Unit UDI/GTIN: 10080196777952, Lot Number: 22KBB459,22IBU513; n. NEURO PACK, Item Number: DYNJ14077I, Case UDI/GTIN: 40193489927673, Unit UDI/GTIN: 10193489927672, Lot Number: 22JBJ480; o. NEURO BASIN PACK-LF, Item Number: DYNJ16473K, Case UDI/GTIN: 40193489920865, Unit UDI/GTIN: 10193489920864, Lot Number: 22JDC136,22HDC097; p. NEURO PACK, Item Number: DYNJ19847M, Case UDI/GTIN: 40888277792303, Unit UDI/GTIN: 10888277792302, Lot Number: 22JBY146,22JBK343; q. MAJOR NEURO PK-LF, Item Number: DYNJ21243M, Case UDI/GTIN: 40193489225236, Unit UDI/GTIN: 10193489225235, Lot Number: 22HBY346; r. NINOR NEURO PK-LF, Item Number: DYNJ21244N, Case UDI/GTIN: 40193489225229, Unit UDI/GTIN: 10193489225228, Lot Number: 22JBV649; s. NEURO VP SHUNT/VAGAL NERVE-RF, Item Number: DYNJ21925S, Case UDI/GTIN: 40195327105021, Unit UDI/GTIN: 10195327105020, Lot Number: 22IBS173; t. NEURO PACK-LF, Item Number: DYNJ25097F, Case UDI/GTIN: 40889942871576, Unit UDI/GTIN: 10889942871575, Lot Number: 22JBK228,22JBI289; u. NEURO PACK, Item Number: DYNJ30477K, Case UDI/GTIN: 40193489912679, Unit UDI/GTIN: 10193489912678, Lot Number: 22JMG210,22HMB824; v. CRANIOTOMY, Item Number: DYNJ35349C, Case UDI/GTIN: 40888277858061, Unit UDI/GTIN: 10888277858060, Lot Number: 22HBP277; w. PEDIATRIC NEURO PACK, Item Number: DYNJ35480F, Case UDI/GTIN: 40193489803694, Unit UDI/GTIN: 10193489803693, Lot Number: 22JMB626,22IMF848,22IMC064; x. NEURO MATRIX PACK, Item Number: DYNJ36594F, Case UDI/GTIN: 40193489928991, Unit UDI/GTIN: 10193489928990, Lot Number: 22JBF140,22IBG656,22HBM006; y. ACOUSTIC NEUROMA PACK-LF, Item Number: DYNJ37741B, Case UDI/GTIN: 40195327081332, Unit UDI/GTIN: 10195327081331, Lot Number: 22JMC172,22HMD563; z. MAJOR NEURO PACK-LF, Item Number: DYNJ37751B, Case UDI/GTIN: 40195327082209, Unit UDI/GTIN: 10195327082208, Lot Number: 22JMF991; aa. SCRIPPS SW NEURO PACK-LF, Item Number: DYNJ39836F, Case UDI/GTIN: 40195327176717, Unit UDI/GTIN: 10195327176716, Lot Number: 22JME760,22HMI113,22HMD278,22GMB860; bb. NEURO PACK, Item Number: DYNJ41225B, Case UDI/GTIN: 40193489675673, Unit UDI/GTIN: 10193489675672, Lot Number: 22JBX284; cc. GENERAL PACK, Item Number: DYNJ41249C, Case UDI/GTIN: 40888277882035, Unit UDI/GTIN: 10888277882034, Lot Number: 22IBG204; dd. NEURO PK, Item Number: DYNJ41276F, Case UDI/GTIN: 40195327204021, Unit UDI/GTIN: 10195327204020, Lot Number: 22JMG029,22HMD792; ee. NEURO PACK, Item Number: DYNJ41579A, Case UDI/GTIN: 40888277935274, Unit UDI/GTIN: 10888277935273, Lot Number: 22IBA883; ff. NEURO PACK, Item Number: DYNJ41633B, Case UDI/GTIN: 40888277934918, Unit UDI/GTIN: 10888277934917, Lot Number: 22IBU441,22HBX309,22HBP962; gg. NEURO PACK, Item Number: DYNJ42484, Case UDI/GTIN: 40884389924045, Unit UDI/GTIN: 10884389924044, Lot Number: 22IBH224; hh. NEURO PACK-LF, Item Number: DYNJ42844D, Case UDI/GTIN: 40193489880565, Unit UDI/GTIN: 10193489880564, Lot Number: 22JBT933,22HBZ336; ii. NEURO PACK, Item Number: DYNJ43177B, Case UDI/GTIN: 40889942884392, Unit UDI/GTIN: 10889942884391, Lot Number: 22IMC154; jj. NEURO PACK, Item Number: DYNJ43856C, Case UDI/GTIN: 40195327200238, Unit UDI/GTIN: 10195327200237, Lot Number: 22IBS432,22HBP196; kk. NEURO PACK, Item Number: DYNJ43934A, Case UDI/GTIN: 40193489733601, Unit UDI/GTIN: 10193489733600, Lot Number: 22IMD151; ll. NEURO PACK, Item Number: DYNJ44014B, Case UDI/GTIN: 40889942278917, Unit UDI/GTIN: 10889942278916, Lot Number: 22KMA329,22JMC591; mm. CRANIOTOMY PACK, Item Number: DYNJ44040, Case UDI/GTIN: 40888277085955, Unit UDI/GTIN: 10888277085954, Lot Number: 22HBV694; nn. PK, NEURO-MINOR, Item Number: DYNJ45053C, Case UDI/GTIN: 40195327248759, Unit UDI/GTIN: 10195327248758, Lot Number: 22IBT923; oo. NEURO PACK, Item Number: DYNJ45267B, Case UDI/GTIN: 40193489852425, Unit UDI/GTIN: 10193489852424, Lot Number: 22HBV288; pp. NEURO PACK, Item Number: DYNJ46232A, Case UDI/GTIN: 40889942153450, Unit UDI/GTIN: 10889942153459, Lot Number: 22JBW760; qq. NEURO PACK, Item Number: DYNJ46238C, Case UDI/GTIN: 40888277737304, Unit UDI/GTIN: 10888277737303, Lot Number: 22IBG653,22HBX977,22HBX980; rr. PK, NEURO-MINOR, Item Number: DYNJ46282A, Case UDI/GTIN: 40193489310475, Unit UDI/GTIN: 10193489310474, Lot Number: 22JBO162,22JBO223,22IBJ968; ss. NEURO BASIC, Item Number: DYNJ46424D, Case UDI/GTIN: 40193489710350, Unit UDI/GTIN: 10193489710359, Lot Number: 22IBD742; tt. NEURO BASIC, Item Number: DYNJ46424F, Case UDI/GTIN: 40195327260683, Unit UDI/GTIN: 10195327260682, Lot Number: 22JBB084; uu. CRANIOTOMY PACK (CRSJE)227-LF, Item Number: DYNJ47487A, Case UDI/GTIN: 40889942106241, Unit UDI/GTIN: 10889942106240, Lot Number: 22JMF056,22HMD080; vv. NEURO PACK, Item Number: DYNJ48765, Case UDI/GTIN: 40888277943491, Unit UDI/GTIN: 10888277943490, Lot Number: 22IBQ098,22IBE948,22HME671; ww. NEURO PACK, Item Number: DYNJ49004, Case UDI/GTIN: 40888277810809, Unit UDI/GTIN: 10888277810808, Lot Number: 22JMI090,22JMC180,22IMD860,22HMD361; xx. NEURO CRANI PACK, Item Number: DYNJ49358, Case UDI/GTIN: 40888277847096, Unit UDI/GTIN: 10888277847095, Lot Number: 22JBQ325; yy. NEURO PACK, Item Number: DYNJ49529B, Case UDI/GTIN: 40889942203780, Unit UDI/GTIN: 10889942203789, Lot Number: 22IBS439,22IBE973; zz. NEURO TRAY-LF, Item Number: DYNJ49580C, Case UDI/GTIN: 40193489231206, Unit UDI/GTIN: 10193489231205, Lot Number: 22IBN986; aaa. NEURO PACK, Item Number: DYNJ50292F, Case UDI/GTIN: 40193489834438, Unit UDI/GTIN: 10193489834437, Lot Number: 22JBV778,22JBC215; bbb. WH NEURO PACK, Item Number: DYNJ50569, Case UDI/GTIN: 40889942008453, Unit UDI/GTIN: 10889942008452, Lot Number: 22JBQ343; ccc. RFT ISC-N NEURO PACK, Item Number: DYNJ50852G, Case UDI/GTIN: 40193489565028, Unit UDI/GTIN: 10193489565027, Lot Number: 22JMA472,22HMD933; ddd. NEURO-LF, Item Number: DYNJ52055F, Case UDI/GTIN: 40195327096336, Unit UDI/GTIN: 10195327096335, Lot Number: 22IBD221; eee. NEURO PACK, Item Number: DYNJ52243B, Case UDI/GTIN: 40889942829294, Unit UDI/GTIN: 10889942829293, Lot Number: 22IMF259; fff. NEURO LUMBAR PACK, Item Number: DYNJ53664D, Case UDI/GTIN: 40193489232067, Unit UDI/GTIN: 10193489232066, Lot Number: 22IBH772; ggg. NEURO FUSION PACK-LF, Item Number: DYNJ54360C, Case UDI/GTIN: 40193489815451, Unit UDI/GTIN: 10193489815450, Lot Number: 22IBR798; hhh. NEURO CERVICAL PACK, Item Number: DYNJ55737A, Case UDI/GTIN: 40889942943266, Unit UDI/GTIN: 10889942943265, Lot Number: 22JMB645; iii. NEURO PACK, Item Number: DYNJ57663, Case UDI/GTIN: 40889942763840, Unit UDI/GTIN: 10889942763849, Lot Number: 22JBW468,22IBB120; jjj. VP SHUNT PACK, Item Number: DYNJ59702F, Case UDI/GTIN: 40195327200269, Unit UDI/GTIN: 10195327200268, Lot Number: 22JBM510; kkk. NEURO BASIC PACK, Item Number: DYNJ60054, Case UDI/GTIN: 40888277797988, Unit UDI/GTIN: 10888277797987, Lot Number: 22KBA488,22JBR179,22IBI219; lll. NEURO PACK, Item Number: DYNJ60792C, Case UDI/GTIN: 40195327126965, Unit UDI/GTIN: 10195327126964, Lot Number: 22HBN525; mmm. CHRISTUS CHILDRENS NEURO PK, Item Number: DYNJ61168B, Case UDI/GTIN: 40195327118946, Unit UDI/GTIN: 10195327118945, Lot Number: 22IBR244,22HBY116,22HBV048; nnn. CHRISTUS CHILDRENS VP SHUNT PK, Item Number: DYNJ61174B, Case UDI/GTIN: 40195327119486, Unit UDI/GTIN: 10195327119485, Lot Number: 22JLA318,22ILA429,22HLB020; ooo. RF NEURO-SHUNT PACK, Item Number: DYNJ61387A, Case UDI/GTIN: 40193489444149, Unit UDI/GTIN: 10193489444148, Lot Number: 22IMG094,22IMG100; ppp. NEURO PACK, Item Number: DYNJ61419A, Case UDI/GTIN: 40193489676366, Unit UDI/GTIN: 10193489676365, Lot Number: 22JMH617; qqq. NEURO PACK, Item Number: DYNJ61740F, Case UDI/GTIN: 40195327238811, Unit UDI/GTIN: 10195327238810, Lot Number: 22IBF063; rrr. NEURO ST DAVIDS MEDICAL CTR, Item Number: DYNJ61998, Case UDI/GTIN: 40193489615013, Unit UDI/GTIN: 10193489615012, Lot Number: 22IBF340,22HBX978,22HBX979; sss. GIO NEURO PACK, Item Number: DYNJ62652A, Case UDI/GTIN: 40193489224581, Unit UDI/GTIN: 10193489224580, Lot Number: 22IBK347; ttt. NEURO KIT, Item Number: DYNJ62854C, Case UDI/GTIN: 40193489426985, Unit UDI/GTIN: 10193489426984, Lot Number: 22IBU624; uuu. DISCECTOMY PACK, Item Number: DYNJ67835, Case UDI/GTIN: 40193489444811, Unit UDI/GTIN: 10193489444810, Lot Number: 22IBV281; vvv. NEURO PACK, Item Number: DYNJ68898C, Case UDI/GTIN: 40195327060283, Unit UDI/GTIN: 10195327060282, Lot Number: 22IMI313; www. BASIC NEURO PACK-LF, Item Number: DYNJ81867, Case UDI/GTIN: 40195327153305, Unit UDI/GTIN: 10195327153304, Lot Number: 22HMC380,22HMB884; xxx. NEURO PACK, Item Number: DYNJ82633, Case UDI/GTIN: 40195327199556, Unit UDI/GTIN: 10195327199555, Lot Number: 22HBM512; yyy. NEURO PACK, Item Number: DYNJ83029, Case UDI/GTIN: 40195327228980, Unit UDI/GTIN: 10195327228989, Lot Number: 22JBV413; zzz. NEURO, Item Number: DYNJ900904J, Case UDI/GTIN: 40193489906616, Unit UDI/GTIN: 10193489906615, Lot Number: 22JBY355,22KBC323; aaaa. NEURO, Item Number: DYNJ901680G, Case UDI/GTIN: 40193489308083, Unit UDI/GTIN: 10193489308082, Lot Number: 22KBA252,22JBE473; bbbb. FHCW NEURO PACK, Item Number: DYNJ901990C, Case UDI/GTIN: 40193489932363, Unit UDI/GTIN: 10193489932362, Lot Number: 22JBL441,22IBV599; cccc. NEURO FERREE, Item Number: DYNJ902388K, Case UDI/GTIN: 40195327224753, Unit UDI/GTIN: 10195327224752, Lot Number: 22IBS547; dddd. NEURO, Item Number: DYNJ902585A, Case UDI/GTIN: 40193489985413, Unit UDI/GTIN: 10193489985412, Lot Number: 22IDA639; eeee. NEURO PACK CW, Item Number: DYNJ903091B, Case UDI/GTIN: 40193489961530, Unit UDI/GTIN: 10193489961539, Lot Number: 22IBU575,22IBU579; ffff. NEURO, Item Number: DYNJ903519A, Case UDI/GTIN: 40889942594215, Unit UDI/GTIN: 10889942594214, Lot Number: 22JBU552; gggg. FHT NEURO, Item Number: DYNJ903566C, Case UDI/GTIN: 40193489932103, Unit UDI/GTIN: 10193489932102, Lot Number: 22JBB451; hhhh. NEURO BASIC, Item Number: DYNJ904052G, Case UDI/GTIN: 40195327220922, Unit UDI/GTIN: 10195327220921, Lot Number: 22KDA580,22KDA234; iiii. NEURO LUMBAR, Item Number: DYNJ904111I, Case UDI/GTIN: 40193489363518, Unit UDI/GTIN: 10193489363517, Lot Number: 22IBO732,22IBN438,22IBA887; jjjj. NCH NEURO, Item Number: DYNJ904797B, Case UDI/GTIN: 40195327037988, Unit UDI/GTIN: 10195327037987, Lot Number: 22IBF188; kkkk. TAVI, Item Number: DYNJ904981C, Case UDI/GTIN: 40193489475136, Unit UDI/GTIN: 10193489475135, Lot Number: 22JBV630,22JBB134; llll. CHN OR NEURO, Item Number: DYNJ907876A, Case UDI/GTIN: 40195327011117, Unit UDI/GTIN: 10195327011116, Lot Number: 22HBR766; mmmm. NEURO PACK-LF, Item Number: DYNJC9302F, Case UDI/GTIN: 40195327032754, Unit UDI/GTIN: 10195327032753, Lot Number: 22JMB483; nnnn. NEURO-CRANI-LF, Item Number: DYNJS1001A, Case UDI/GTIN: 40080196665373, Unit UDI/GTIN: 10080196665372, Lot Number: 22IBS931,22HBX420
Devices named in this recall
92 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ42844DMedline Industries, Inc. · Neurological tray5 recalls0 reports
- Medline Industries, Inc. DYNJ902585AMedline Industries, Inc. · Neurological tray4 recalls0 reports
- Medline Industries, Inc. DYNJ21925SMedline Industries, Inc. · Neurological tray4 recalls0 reports
- Medline Industries, Inc. DYNJ60792CMedline Industries, Inc. · Neurological tray4 recalls0 reports
- Medline Industries, Inc. DYNJ35349CMedline Industries, Inc. · Neurological tray4 recalls0 reports
- Medline Industries, Inc. DYNJ04596KMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ62854CMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67835Medline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ50852GMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ61998Medline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ900904JMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ0781823KMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ907876AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. CDS983186JMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ41579AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ48765Medline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ47487AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ0367094LMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ44014BMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ07952JMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJ55737AMedline Industries, Inc. · Neurological tray3 recalls0 reports
- Medline Industries, Inc. DYNJS1001AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ14032Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ62652AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ49580CMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ53664DMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ901680GMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904111IMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61387AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ904981CMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0038440BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ41225BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43934AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ54360CMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ50292FMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ45267BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903091BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJC9302FMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. CDS985548NMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ37741BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ37751BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61168BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61174BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ81867Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ39836FMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ82633Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ902388KMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ83029Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61740FMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ44040Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ46238CMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ49004Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ49358Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ41633BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ903519AMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ57663Medline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ52243BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ25097FMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43177BMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ0374024JMedline Industries, Inc. · Neurological tray2 recalls0 reports
- Medline Industries, Inc. DYNJ43856CMedline Industries, Inc. · Neurological tray1 recall1 report
- Medline Industries, Inc. DYNJ21244NMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ21243MMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ46282AMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ0190156FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ61419AMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ46424DMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ35480FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ30477KMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ16473KMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ14077IMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ36594FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ903566CMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ901990CMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ904797BMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ68898CMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ52055FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. CDS983098DMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ59702FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ41276FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ904052GMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ45053CMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ46424FMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ42484Medline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. CDS985289Medline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ04671QMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ19847MMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ60054Medline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ41249CMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ50569Medline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ46232AMedline Industries, Inc. · Neurological tray1 recall0 reports
- Medline Industries, Inc. DYNJ49529BMedline Industries, Inc. · Neurological tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91386).
- Z-1060-2023 · Open, ClassifiedANCILLARY PACK-LF. Sterile Procedural Tray.
- Z-1061-2023 · Open, ClassifiedTRACH PACK. Sterile Procedural Tray.
- Z-1062-2023 · Open, ClassifiedLOCAL PACK. Sterile Procedural Tray.
- Z-1063-2023 · Open, ClassifiedSterile Procedural Tray, labeled as the following: a. PAIN SERVICE PACK-LF; b. LOCAL BASIC
- Z-1064-2023 · Open, ClassifiedUNIVERSAL CORE DR RUNNELS PACK. Sterile Procedural Tray
- Z-1065-2023 · Open, ClassifiedFEMORAL ARTERY PACK-LF. Sterile Procedural Tray
- Z-1066-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. CARDIAC CATH PACK-LF; b. ANGIOGRAPHY-NS-LF; c. EP BASIC SET UP PACK; d. IMPLANT PACK-LF; e. PORT PACEMAKER INSERTI…
- Z-1067-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. EP LAB PACEMAKER PACK b. OPEN HEART CDS c. OPEN HEART CDS d. AAA CDS e. OPEN HEART CDS f. PACEMAKER CDS g. HEART CDS h. VALV…
- Z-1068-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. PORT/TRACH INSERTION CDS-3 b. RESIDENT DERM SURG PACK c. THROMBECTOMY PACK-LF d. AV FISTULA PACK e. AV SHUNT PACK-LF f. RR-V…
- Z-1069-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. BREAST BIOPSY PACK-LF b. BREAST BIOPSY PACK c. LAPAROTOMY PACK d. BREAST PACK e. SOFT TISSUE PACK f. HEMORRHOID PACK g. RECT…
- Z-1070-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. ORAL SURGERY PACK-LF b. EXTRACTION PACK c. ORAL SURGERY PACK d. DENTAL MINOR PACK e. ORAL
- Z-1071-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. RHINOPLASTY PACK b. OSC OTOLOGY PACK-LF c. RHYZOTOMY
- Z-1072-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. EAR CDS-LF b. SINUS CDS-1 c. ZALE FESS CDS d. ZALE OTOLOGY CDS e. ZALE HEAD & NECK CDS f. CUH PARA/THYROIDECTOMY CDS g. CAM …
- Z-1073-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. MINOR/BREAST BIOPSY-LF b. BREAST BIOPSY-LF
- Z-1074-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. LAPAROTOMY CDS-LF b. LAPAROSCOPY CDS-LF c. BASIC LAPAROSCOPY CDS d. LAPAROSCOPY CDS-LF e. GENERAL LAPAROSCOPY PROCEDURE f. L…
- Z-1075-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: GENERAL ENDOSCOPY
- Z-1076-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. ENDOSCOPY-LF b. SEPTO/ENDO
- Z-1077-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINOPLASTY CDS d. ACH PLASTICS BREAST-LF e. NE BREAST RECON CDS f. P…
- Z-1078-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. ST. ANNE'S MAJOR BASIN SET UP b. SINGLE BASIN SET c. DOUBLE BASIN SET d. LAPAROTOMY BASIN e. ORTHOPEDIC CUSTOM BASIN SET f.…
- Z-1079-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. BASIN SET-UP PACK b. LAP CHOLE CDS c. MAJOR CDS d. UROLOGY MINOR CDS e. UROLOGY MAJOR CDS f. SINGLE BASIC CDS g. ORGAN RECO…
- Z-1080-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. UNIVERSAL PLASTIC b. KNEE ARTHROSCOPY PACK
- Z-1081-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. WOUND CLOSURE TRAY b. WOUND PACK
- Z-1082-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: IM NAIL
- Z-1083-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. MAJOR CDS-LF b. BURN PACK-LF c. BURN PACK d. SMALL BURN
- Z-1085-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: LS MS PACK
- Z-1086-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: DR MOTT DRAPE PACK
- Z-1089-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. GYN LAPAROTOMY HARPER PACK-LF b. OB GYN LAP PACK-LF c. VAGINAL HYSTERECTOMY PACK d. DELTA PACK e. MAJOR VAGINAL-LF f. GYN …
- Z-1090-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. PELVISCOPY PACK b. D-PELVISCOPY
- Z-1091-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. GYN-URO LAPSCP CDS-LF b. ROBOTIC CDS c. GYN LAPAROSCOPY CDS-LF d. GYN LAPAROSCOPY e. GYN LAPAROSCOPY CDS f. UNIVERSAL ROBO…
- Z-1092-2023 · Open, ClassifiedSterile Procedural Trays, labeled as the following: a. C-SECTION OR 30 CDS b. C-SECTION CDS c. C-SECTION CDS d. C-SECTION CDS e. MAJOR ORTHO CDS-LF f. C-SECTION CDS g. C-SECTION…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.