HLS Set recallZ-1188-2023 · Maquet Medical Systems USA · initiated Feb 13, 2023

Class II Open, Classified All recalls posted in February 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Medical Systems USA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm has initiated a removal of the product due to insufficient evidence of packaging sterility. Exposure to a non-sterile medical device may result in inflammation, infection, sepsis, or ischemia.

FDA root cause category: Process control

What the company is doing

On February 13, 2023, the firm notified customers via Urgent Medical Device Correction letters. Customers were informed of the potential packaging issues and the associated risks to health. Customers were instructed to stop use and return all unused and unexpired affected product for credit. If customers have no alternative product, they may choose to continue use of the product, but they must perform a visual inspection of the primary packaging, checking for visible stress marks or damage to the packaging. If there are visible stress marks in the packaging, the customer should not use the product. Please note that there is a possibility that packaging defects may not be detectable by the naked eye. If you have any questions, please contact your Getinge representative or call the Getinge Customer Support at (888) 9GETUSA / (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Recall details

Recall number
Z-1188-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 13, 2023
Posted by the FDA
Feb 28, 2023
Recalling firm
Maquet Medical Systems USA
Quantity
10,587 units (US); 94,987 (worldwide)
Product code
DTZ
FDA submissions
K112360
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czechia, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mauritius, Myanmar, Mexico, Moldova, Morocco, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of North Macedonia, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, The Republic of Korea, The Russian Federation, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Vietnam.

Product description

BEQ-HLS 5050 USA; HLS Set Advanced 5.0, Product Code 70106.9077, BEQ-HLS 7050 USA; HLS Set Advanced 7.0, Product Code 70106.9078. Part of the CARDIOHELP System, a pre-connected set used for extracorporeal respiratory and/or cardiovascular support.

Code information (models, lots, serial numbers)

HLS Set Advanced 5.0, Product Code 70106.9077 UDI-DI 04058863076355 HLS Set Advanced 7.0, Product Code 70106.9078 UDI-DI 04058863080383 All lots/batches affected

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K112360 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates