CardioMEMS HF System PA Sensor and Delivery System, Model… recallZ-1224-2023 · St. Jude Medical · initiated Feb 7, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from St. Jude Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
FDA root cause category: Device Design
What the company is doing
Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to do the following: 1. Prior to the implantation procedure, sensor data should be entered, and the clinician should proceed to the next screen to complete the Error #8 screening, prior to venipuncture. 2. If an Error #8 message occurs, the sensor should not be implanted. 3. Work with your Abbott Sales Representative to exchange the affected sensor. 4. Review Appendix B for labeling and supplemental information related to implantation and monitoring of CardioMEMS PA Sensors.
Recall details
- Recall number
- Z-1224-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 7, 2023
- Posted by the FDA
- Mar 6, 2023
- Recalling firm
- St. Jude Medical
- Quantity
- 1905 units
- Product code
- MOM
- FDA submissions
- P100045
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Product description
CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.
Code information (models, lots, serial numbers)
Show the full code information (23k characters)
UDI/DI 05415067029331: M140100179, M140100180, M140100181, M140100182, M140100183, M140100184, M140100185, M140100186, M140200187, M140200188, M140200190, M140200191, M140200193, M140200196, M140200204, M140200205, M140200213, M140200215, M140200216, M140200218, M140200220, M140200221, M140200223, M140200224, M140200226, M140200228, M140200230, M140200233, M140200234, M140200235, M140200237, M140200238, M140200240, M140200243, M140200245, M140400469, M140400470, M140400471, M140400472, M140400474, M140600522, M140600530, M140600532, M140600536, M140600537, M140600538, M140600541, M140600542, M140600543, M140600545, M140600546, M140700687, M140700688, M140700689, M140700690, M140700691, M140700692, M140700693, M140700694, M140700698, M140700699, M140700700, M140700701, M140700703, M140700705, M140700710, M140700712, M140700713, M140700714, M140700715, M140700717, M140700718, M140700721, M140700722, M140700725, M140700726, M140700799, M140700800, M140700801, M140700804, M140700806, M140700817, M140700818, M140800820, M140800821, M140800822, M140800825, M140800827, M140800828, M140800829, M140800833, M140800835, M140800837, M140800838, M140800839, M140800841, M140800842, M140800846, M140800848, M140800849, M140800850, M140800853, M140800855, M140800856, M140800859, M140800861, M140800862, M140800864, M140800867, M140800868, M140800869, M140800872, M140800874, M140800876, M140800877, M140800879, M140800880, M140800881, M140800883, M140800886, M140800887, M140800890, M140800891, M140800895, M140800896, M140800918, M140800919, M140800922, M140800925, M140800926, M140901157, M140901158, M140901159, M140901160, M140901161, M140901162, M140901163, M140901164, M140901166, M140901169, M140901172, M140901174, M140901175, M140901177, M140901184, M140901185, M140901186, M141001337, M141001339, M141001341, M141001344, M141001345, M141001353, M141001354, M141001355, M141001356, M141001358, M141001359, M141001360, M141001361, M141001362, M141001363, M141001366, M141201849, M141201857, M141201859, M141201861, M141201862, M141201863, M141201865, M141201866, M141201867, M141201869, M141201870, M141201871, M141201873, M141201874, M141201876, M141201879, M141201880, M141201882, M141201883, M141201885, M141201886, M141201887, M141201889, M141201890, M141201891, M141201893, M141201894, M141201895, M141201897, M141201898, M141201899, M141201900, M141201901, M141201902, M141201903, M141201904, M141201906, M150102237, M150102238, M150102239, M150102241, M150102243, M150102245, M150102246, M150102248, M150102253, M150102254, M150102255, M150102257, M150102259, M150102260, M150102261, M150102263, M150102265, M150102266, M150202497, M150202498, M150202499, M150202502, M150202503, M150202504, M150202505, M150202507, M150202509, M150202512, M150202514, M150202516, M150202526, M150302647, M150302649, M150302650, M150302652, M150302655, M150402907, M150402911, M150402912, M150402914, M150402916, M150402918, M150402919, M150402920, M150402921, M150402922, M150402924, M150402925, M150402926, M150402927, M150402929, M150402931, M150402932, M150402934, M150402935, M150402937, M150402938, M150402941, M150402942, M150402943, M150402944, M150402945, M150402946, M150403187, M150403188, M150403189, M150403190, M150403191, M150403192, M150403193, M150403194, M150503377, M150503378, M150503379, M150503380, M150503381, M150503384, M150503385, M150503390, M150503391, M150503393, M150503397, M150503398, M150503400, M150503401, M150503404, M150503407, M150503409, M150503410, M150503411, M150503413, M150503414, M150503415, M150503416, M150503419, M150503420, M150503423, M150503424, M150704378, M150704379, M150704380, M150704381, M150704383, M150704384, M150704385, M150704386, M150704387, M150704389, M150704391, M150704393, M150704397, M150704398, M150704399, M150704401, M150704403, M150704405, M150704407, M150704412, M150704413, M150704414, M150704416, M150704417, M150704418, M150704419, M150704420, M150704421, M150704422, M150704423, M150704424, M150704426, M150804897, M150804899, M150804901, M150804903, M150804904, M150804906, M150804907, M150804908, M150804909, M150804910, M150804911, M150804912, M150804914, M150804915, M150804916, M150804917, M150804918, M150804919, M150804920, M150804921, M150804922, M150804923, M150804924, M150804926, M150804927, M150804929, M150804932, M150804933, M150804934, M150804935, M150804936, M150804937, M150804938, M150804939, M150804943, M150804944, M150804945, M150804947, M150804948, M150804949, M150804950, M150804951, M150804952, M150804953, M150804954, M150804955, M150905408, M150905410, M150905411, M150905412, M150905413, M150905414, M150905415, M150905416, M150905417, M150905418, M150905419, M150905420, M150905424, M150905426, M150905427, M150905428, M150905429, M150905430, M150905431, M150905432, M150905433, M150905435, M150905717, M150905718, M150905719, M150905721, M150905722, M150905724, M150905728, M150905729, M150905730, M150905731, M150905732, M150905734, M150905739, M150905740, M150905741, M150905742, M150905745, M151005983, M151005988, M151005989, M151005994, M151005995, M151106357, M151106359, M151106362, M151106363, M151106364, M151106366, M151106367, M151106368, M151106369, M151106371, M151106372, M151106375, M151106376, M151106381, M151106382, M151106385, M151106387, M151106389, M151106390, M151106391, M151106392, M151106393, M151106394, M151106396, M151106397, M151106398, M151106399, M151106400, M151106401, M151106403, M151106404, M151106405, M151206527, M151206528, M151206530, M151206531, M151206533, M151206537, M151206539, M151206540, M151206541, M151206542, M151206544, M151206545, M151206546, M151206548, M151206549, M151206550, M151206551, M151206554, M151206556, M151206558, M151206559, M151206560, M151206561, M151206563, M151206566, M160307210, M160307211, M160307214, M160307219, M160307221, M160307223, M160307224, M160307226, M160307227, M160307234, M160707625, M160807677, M160807678, M160807679, M160807680, M160807682, M160807683, M160807684, M160807685, M160807687, M160807690, M160807692, M161107927, M161107928, M161107929, M161107930, M161107931, M161107932, M161107933, M161107934, M161107935, M161107936, M161107937, M161107938, M161107939, M161107940, M161107941, M161107942, M161107943, M161107944, M161107945, M161207947, M161207948, M161207953, M161207955, M161207956, M161207957, M161207958, M161207959, M161207960, M161207961, M161207962, M161207963, M161207964, M161207965, M161207968, M161207969, M161207970, M161207972, M161207973, M161207974, M161207975, M161207976, M161207978, M161207980, M161207981, M161207983, M161207984, M161207985, M161208049, M161208065, M161208066, M161208067, M161208068, M161208069, M161208070, M161208071, M161208072, M161208073, M161208074, M161208075, M161208076, M161208077, M161208078, M161208079, M161208080, M161208081, M161208082, M161208084, M170108086, M170108087, M170108088, M170108089, M170108090, M170108091, M170108092, M170108093, M170108095, M170108096, M170108097, M170108098, M170108100, M170108101, M170108102, M170108103, M170108104, M170108106, M170108107, M170108108, M170108109, M170108110, M170108111, M170108112, M170108113, M170108114, M170108115, M170108119, M170108120, M170108122, M170108123, M170108124, M170108125, M170108126, M170108127, M170108130, M170108131, M170108132, M170108133, M170108134, M170108135, M170108136, M170108137, M170108138, M170108140, M170108141, M170108142, M170108144, M170108145, M170108146, M170108147, M170108148, M170108149, M170108150, M170108153, M170108154, M170108155, M170108156, M170108157, M170208158, M170208159, M170208160, M170208161, M170208162, M170208163, M170208164, M170208165, M170208166, M170208167, M170208168, M170208169, M170208170, M170208171, M170208172, M170208173, M170208174, M170208175, M170208176, M170208177, M170208178, M170208179, M170208181, M170208182, M170208183, M170208184, M170208185, M170208187, M170308188, M170308190, M170308191, M170308192, M170308193, M170308194, M170308195, M170308196, M170308197, M170308200, M170308201, M170308202, M170308203, M170308204, M170308205, M170308206, M170308207, M170308208, M170308209, M170308210, M170308211, M170308212, M170308213, M170308214, M170308215, M170408216, M170408217, M170408218, M170408219, M170408220, M170408222, M170408223, M170408224, M170408225, M170408226, M170408227, M170408228, M170408229, M170408230, M170408231, M170408232, M170408233, M170408234, M170408235, M170408240, M170408241, M170408242, M170408243, M170408244, M170408246, M170408249, M170408250, M170408252, M170408253, M170408254, M170408255, M170408256, M170408257, M170408259, M170508260, M170508261, M170508275, M170608293, M170608294, M170608295, M170608296, M170608297, M170708308, M170708309, M170708310, M170708311, M170708313, M170708314, M170708315, M170708316, M170708317, M170708321, M170708323, M170708339, M170808325, M170808326, M170808328, M170808329, M170808336, M170808337, M170808338, M170808340, M170808341, M170808342, M170808344, M170808345, M170808346, M170808347, M170808352, M170808353, M170808357, M170908358, M170908360, M170908361, M170908362, M170908364, M170908366, M170908367, M170908368, M170908369, M170908370, M170908371, M170908372, M170908373, M170908374, M170908377, M170908378, M170908380, M170908383, M170908384, M170908387, M170908389, M170908390, M170908392, M170908393, M170908394, M170908395, M170908396, M171108397, M171108398, M171108399, M171108400, M171108402, M171108403, M171108406, M171108407, M171108408, M171108409, M171108410, M171108412, M171108413, M171108414, M171108415, M171108416, M171108417, M171108419, M171108422, M171108423, M171108424, M171108425, M171108426, M171108427, M171108428, M171108429, M171108430, M171108431, M171108432, M171108433, M171108436, M171108437, M171108438, M171108439, M171108440, M171108441, M171108442, M171108443, M180108444, M180108445, M180108451, M180108453, M180108455, M180108456, M180108457, M180108459, M180108460, M180108462, M180108464, M180108465, M180108467, M180108468, M180108469, M180108470, M180108471, M180108472, M180108474, M180208475, M180208476, M180208479, M180208480, M180208481, M180208482, M180208483, M180208484, M180208485, M180208488, M180208489, M180208491, M180208492, M180208493, M180208494, M180208495, M180208497, M180208498, M180208499, M180208501, M180208502, M180208503, M180208504, M180208505, M180208506, M180208507, M180208509, M180208510, M180208511, M180208512, M180208514, M180208517, M180308526, M180308527, M180308529, M180308535, M180308537, M180308539, M180408540, M180408543, M180408544, M180408545, M180408546, M180408549, M180408551, M180408552, M180408553, M180408554, M180408555, M180408556, M180408557, M180408562, M180408574, M180408575, M180408576, M180408577, M180408578, M180408579, M180408580, M180408583, M180408584, M180408585, M180408586, M180408587, M180408588, M180408589, M180408590, M180408591, M180508593, M180508596, M180508598, M180508600, M180508601, M180508602, M180508603, M180508604, M180508605, M180508606, M180508607, M180508608, M180508611, M180508617, M180508619, M180508620, M180508627, M180508629, M180508630, M180508631, M180608632, M180608634, M180608635, M180608637, M180608638, M180608639, M180608642, M180608643, M180608644, M180608645, M180608646, M180608647, M180608648, M180608649, M180608651, M180608652, M180608653, M180608655, M180608656, M180608657, M180608658, M180608659, M180608660, M180608661, M180608662, M180608663, M180608667, M180608668, M180608669, M180608670, M180608671, M180608672, M180608673, M180608674, M180608677, M180608678, M180608679, M180608680, M180608681, M180608683, M180608684, M180608689, M180608690, M180608691, M180608692, M180608693, M180608694, M180608697, M180608698, M180608699, M180608700, M180608701, M180608702, M180608707, M180608708, M180608710, M180608711, M180608712, M180608713, M180608714, M180608717, M180608718, M180608719, M180608720, M180608721, M180608722, M180608723, M180608724, M180608725, M180608726, M180608727, M180608728, M180608730, M180608731, M180608732, M180608734, M180608735, M180608736, M180608737, M180608738, M180608739, M180608740, M180608743, M180608745, M180608746, M180608747, M180608748, M180608749, M180608750, M180608751, M180608752, M180608753, M180608754, M180608755, M180608756, M180608757, M180608758, M180608759, M180608760, M180608762, M180608763, M180608764, M180608765, M180608766, M180608768, M180608769, M180608770, M180608772, M180608773, M180608774, M180608775, M180608776, M180608777, M180608778, M180608779, M180608782, M180608783, M180608784, M180608785, M180608786, M180608787, M180608788, M180608789, M180608790, M180608792, M180608793, M180608794, M180608795, M180608796, M180608797, M180608798, M180608800, M180608802, M180608803, M180608804, M180608805, M180608807, M180608808, M180608809, M180608810, M180608811, M180608812, M180608813, M180608815, M180708816, M180708819, M180708820, M180708821, M180708822, M180708827, M180808831, M180808833, M180808834, M180808836, M180808843, M180808851, M180808852, M180808853, M180808854, M181008860, M181008861, M181008862, M181008863, M181008865, M181008866, M181008867, M181008868, M181008872, M181008873, M181008874, M181008875, M181008876, M181008877, M181008878, M181008882, M181008883, M181008884, M181008885, M181008887, M181008888, M181008889, M181008890, M181008891, M181008892, M181008893, M181008894, M181008895, M181008896, M181008897, M181008898, M181008899, M181108904, M181108905, M181108907, M181108908, M181108910, M181108912, M181108913, M181208914, M181208916, M181208917, M181208918, M181208919, M181208920, M181208921, M181208922, M181208923, M181208924, M190108925, M190108927, M190108929, M190108930, M190108931, M190108933, M190108934, M190108935, M190108936, M190108956, M190208964, M190208965, M190208966, M190208967, M190208968, M190208969, M190208970, M190208971, M190208972, M190208973, M190208974, M190208975, M190208987, M190208988, M190208989, M190208990, M190208992, M190208993, M190208994, M190208995, M190208996, M190300005, M190300006, M190300007, M190300008, M190300009, M190300010, M190300011, M190300014, M190300015, M190300016, M190300017, M190300018, M190300019, M190400005, M190400006, M190400007, M190400008, M190400010, M190500019, M190500042, M190500043, M190500052, M190600005, M190700001, M190700002, M190700003, M190700004, M190700005, M190800008, M190800009, M190800011, M190800012, M190800013, M190800014, M190800015, M190800016, M190800017, M190800018, M190800027, M190800028, M190800029, M190800030, M190800031, M190900001, M190900002, M190900005, M190900024, M190900025, M190900026, M190900028, M190900032, M190900034, M190900035, M191000001, M191000003, M191000004, M191000005, M191000006, M191000007, M191000008, M191000009, M191000010, M191000011, M191100008, M191100009, M191100010, M191100011, M191200001, M191200002, M191200003, M191200004, M191200005, M200100001, M200100003, M200100004, M200100005, M200100006, M200100007, M200200001, M200200002, M200200003, M200200004, M200200009, M200300001, M200300006, M200300009, M200400001, M200400002, M200400004, M200400005, M200400017, M200400018, M200400019, M200400023, M200400024, M200500002, M200500003, M200500004, M200500005, M200500006, M200500007, M200500008, M200500009, M200500010, M200600001, M200600002, M200600003, M200600004, M200600005, M200600006, M200600007, M200600008, M200600011, M200600012, M200600013, M200600015, M200600017, M200600020, M200600021, M200600022, M200600023, M200700001, M200700007, M200700015, M200700026, M200700027, M200700028, M200700029, M200700030, M200700031, M200700032, M200800001, M200800002, M200800003, M200800004, M200800005, M200800006, M200800007, M200800008, M200800009, M200800010, M200800011, M200800012, M200800013, M200800014, M200800015, M200800016, M200800017, M200900001, M200900002, M200900003, M200900004, M200900005, M200900006, M200900007, M200900008, M200900009, M200900010, M200900011, M200900012, M200900013, M200900014, M200900015, M200900016, M200900017, M200900018, M200900019, M200900020, M200900021, M201000003, M201000004, M201000005, M201000006, M201000007, M201000008, M201000009, M201000010, M201000011, M201000012, M201000013, M201000015, M201000018, M201000019, M201000020, M201000021, M201100001, M201100002, M201100003, M201100004, M201100005, M201100006, M201100007, M201100008, M201100059, M201100060, M201100062, M201100063, M201100064, M201100065, M201100066, M201100067, M201100068, M201200001, M201200002, M201200003, M201200005, M201200006, M210100001, M210100002, M210100003, M210100005, M210100006, M210100007, M210100008, M210100010, M210100011, M210100012, M210100013, M210100014, M210100015, M210100016, M210100017, M210100019, M210100020, M210100021, M210200001, M210200002, M210200003, M210200004, M210200005, M210200006, M210200007, M210200008, M210200009, M210200010, M210200012, M210200013, M210200014, M210200015, M210200016, M210200017, M210200018, M210200019, M210200025, M210200030, M210200031, M210200032, M210200033, M210200034, M210200035, M210300003, M210300004, M210300005, M210300006, M210300007, M210300009, M210300010, M210300011, M210300012, M210300013, M210300015, M210300016, M210300018, M210300019, M210300020, M210300022, M210300023, M210300024, M210300025, M210400007, M210400008, M210400009, M210400010, M210400011, M210400012, M210400013, M210400014, M210400015, M210400027, M210400028, M210400029, M210400030, M210400031, M210400032, M210400033, M210400034, M210400035, M210400036, M210400037, M210400038, M210400039, M210400040, M210400041, M210400042, M210400043, M210400044, M210400045, M210400046, M210400047, M210400048, M210400049, M210400050, M210500001, M210500003, M210500004, M210500005, M210500006, M210500007, M210500008, M210500009, M210500010, M210500011, M210500012, M210500013, M210500014, M210500015, M210500016, M210500017, M210500018, M210500019, M210500020, M210500021, M210500022, M210500023, M210500024, M210500025, M210500026, M210500027, M210500028, M210500029, M210500030, M210500031, M210500032, M210500033, M210500034, M210500036, M210500037, M210500039, M210500040, M210500041, M210800002, M210800003, M210800005, M210800006, M210800007, M210800008, M210800009, M210800010, M210800011, M210800012, M210800013, M210800014, M210800015, M210800016, M210800017, M210800019, M210800020, M210800021, M210800022, M210800023, M210800024, M210800025, M210800026, M210800027, M210800028, M210800029, M210800030, M210800031, M210800032, M210800033, M210900001, M210900003, M210900004, M210900005, M210900006, M210900007, M210900008, M210900010, M210900011, M210900012, M210900013, M210900014, M210900015, M210900016, M210900017, M210900018, M210900019, M210900020, M210900021, M210900022, M210900023, M210900024, M210900025, M211000001, M211000003, M211000004, M211000005, M211000007, M211000008, M211000009, M211000010, M211000011, M211000012, M211000013, M211000014, M211200001, M211200002, M211200003, M211200004, M211200005, M140600524, M140700798, M140700803, M140800826, M140800866, M140901181, M140901183, M141201855, M141201864, M150202510, M150202513, M150402908, M150402910, M150402933, M150503406, M150503408, M150503425, M150905423, M151005977, M151005980, M151005981, M151005982, M151005984, M151005985, M151005986, M151005987, M151005991, M151005992, M151005993, M151206552, M151206557, M160307208, M160307222, M160307225, M160307231, M160307235, M160707618, M160707619, M160707621, M160707622, M160707631, M160707632, M160707633, M160707636, M161208048, M161208051, M170108105, M170608276, M170608278, M170608279, M170608280, M170608281, M170608283, M170608284, M170608288, M170608291, M170608292, M170608299, M170608300, M170608304, M170608306, M170608307, M170708322, M170808330, M170808331, M170808332, M170808333, M170808334, M170808335, M170808348, M170808349, M170808350, M170808351, M171108405, M171108435, M180208513, M180308522, M180308524, M180308528, M180308530, M180308531, M180308532, M180308533, M180308534, M180408558, M180408559, M180408560, M180408561, M180408592, M180508594, M180508595, M180508613, M180508614, M180508615, M180508622, M180508623, M180508624, M180608640, M180608664, M180608665, M180608666, M180608675, M180608676, M180608685, M180608686, M180608704, M180608705, M180608706, M180608715, M180608716, M180708824, M180708825, M180708826, M180808828, M180808829, M180808838, M180808839, M180808842, M180808844, M180808846, M180808847, M180808848, M180808849, M180808850, M180908856, M180908858, M180908859, M181008869, M181008879, M181008880, M181008901, M181008902, M181008903, M190108937, M190108938, M190108939, M190108940, M190108941, M190108942, M190108943, M190108945, M190108947, M190108948, M190108949, M190108950, M190108951, M190108952, M190108953, M190108954, M190300001, M190300003, M190300004, M190400001, M190400003, M190400004, M190500001, M190500002, M190500003, M190500004, M190500005, M190500006, M190500007, M190500008, M190500009, M190500016, M190500017, M190500018, M190500021, M190500022, M190500044, M190500045, M190500046, M190500047, M190500048, M190500049, M190500050, M190500051, M190600001, M190600002, M190600006, M190600007, M190600008, M190600009, M190600010, M190600011, M190600012, M190600013, M190600014, M190600015, M190600017, M190600018, M190600019, M190600020, M190600021, M190600023, M190600026, M190600027, M190600028, M190600029, M190700006, M190700007, M190700008, M190700009, M190700010, M190700011, M190700012, M190700013, M190700014, M190700015, M190700016, M190700017, M190700018, M190700019, M190700020, M190700021, M190700022, M190700023, M190700024, M190700025, M190700026, M190700027, M190700029, M190700030, M190700031, M190700032, M190700033, M190700034, M190700035, M190700036, M190800002, M190800003, M190800004, M190800005, M190800006, M190800007, M190800019, M190800020, M190800021, M190800022, M190800023, M190800024, M190800025, M190800026, M190900039, M190900041, M200700002, M200700003, M200700004, M200700005, M200700006, M200700008, M200700009, M200700010, M200700011, M200700012, M200700013, M200700014, M200700016, M200700017, M200700018, M200700019, M200700020, M200700021, M200700022, M200700024, M200700025, M210200020, M210200021, M210200022, M210200023, M210200024, M210200026, M210200027, M210200028, M210200029, M210200038, M210200039, M210300001, M210300002, M210300014, M210300017, M210400001, M210400002, M210400003, M210400004, M210400006, M210400018, M210400019, M210400020, M210400021, M210400022, M210400023, M210400025, M210400026, M210500042, M210500043, M210600001, M210600002, M210600003, M210600004, M210600005, M210600006, M210600007, M210800035, M210800036, M210800037, M210800038, M210800039, M210800040, M210800041, M210800042, M210800043, M210800044, M210800045, M210800046, M211000023, M211000024, M211000025, M220600002, M220600003, M220600004, M220600005, M220600006, M220600008, M220600011, M220600034, M220600036, M220600037, M220600038, M220600039, M220600040, M220600041, M220600042, M220700189, M220700192
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CardioMEMS™ CM3000Cardiomems, Inc. · System, hemodynamic, implantable3 recalls65 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91606).
- Z-1223-2023 · Open, ClassifiedCardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and manage…
- Z-1225-2023 · Open, ClassifiedCardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and manage…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.