CardioMEMS HF System PA Sensor and Delivery System, Model… recallZ-1225-2023 · St. Jude Medical · initiated Feb 7, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from St. Jude Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.

FDA root cause category: Device Design

What the company is doing

Abbott notified consignees on about 02/07/2023 via hand-delivery of the letter within the US. The letter instructed consignees to do the following: 1. Prior to the implantation procedure, sensor data should be entered, and the clinician should proceed to the next screen to complete the Error #8 screening, prior to venipuncture. 2. If an Error #8 message occurs, the sensor should not be implanted. 3. Work with your Abbott Sales Representative to exchange the affected sensor. 4. Review Appendix B for labeling and supplemental information related to implantation and monitoring of CardioMEMS PA Sensors.

Recall details

Recall number
Z-1225-2023
Status
Open, Classified
Initiated by the company
Feb 7, 2023
Posted by the FDA
Mar 6, 2023
Recalling firm
St. Jude Medical
Quantity
317 units
Product code
MOM
FDA submissions
P100045
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.

Product description

CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

UDI/DI 05415067034724: M220200004, M220200005, M220200006, M220200008, M220200011, M220200014, M220200016, M220200017, M220200018, M220200019, M220200021, M220200022, M220200026, M220200028, M220200029, M220200031, M220200032, M220200034, M220200035, M220200036, M220200037, M220200039, M220200040, M220200041, M220200042, M220300002, M220300003, M220300004, M220300005, M220300007, M220300008, M220300009, M220300010, M220300011, M220300012, M220300013, M220300014, M220300016, M220300017, M220300018, M220300019, M220300021, M220300022, M220300023, M220300024, M220300025, M220300026, M220300027, M220300028, M220300029, M220300030, M220300031, M220300032, M220300035, M220300036, M220300040, M220300041, M220300042, M220300043, M220300044, M220300045, M220300046, M220300047, M220300048, M220300049, M220300050, M220300051, M220300052, M220300053, M220300054, M220300055, M220300057, M220300058, M220300059, M220300060, M220300061, M220300062, M220300063, M220300064, M220300065, M220300066, M220300067, M220300069, M220300070, M220300071, M220300073, M220300074, M220300075, M220300077, M220300078, M220300080, M220300081, M220300082, M220300083, M220300084, M220300085, M220300087, M220300088, M220300089, M220300091, M220300092, M220300093, M220300096, M220300097, M220300101, M220300102, M220300103, M220300104, M220300105, M220300107, M220300108, M220300109, M220300111, M220300112, M220300114, M220300115, M220300116, M220300118, M220300119, M220300120, M220300121, M220300122, M220300123, M220300125, M220300126, M220300127, M220300129, M220300131, M220300136, M220300138, M220300139, M220300140, M220300141, M220300142, M220300143, M220300144, M220300145, M220300146, M220300147, M220300150, M220300151, M220300152, M220300153, M220300154, M220300155, M220300156, M220300157, M220300158, M220300160, M220300161, M220300162, M220300163, M220300164, M220300166, M220300167, M220300168, M220300169, M220300170, M220300171, M220300172, M220300173, M220300174, M220300175, M220300176, M220300177, M220300178, M220300179, M220300180, M220300181, M220300182, M220300183, M220300184, M220300186, M220300187, M220300188, M220300189, M220300190, M220300191, M220300192, M220300193, M220300194, M220300195, M220300196, M220700025, M220700026, M220700027, M220700028, M220700031, M220700033, M220700034, M220700035, M220700036, M220700037, M220700041, M220700042, M220700043, M220700045, M220700051, M220700052, M220700054, M220700059, M220700061, M220700097, M220700098, M220700102, M220700104, M220700106, M220700107, M220700109, M220700110, M220700111, M220700112, M220700113, M220700114, M220700115, M220700116, M220700117, M220700118, M220700119, M220700120, M220700121, M220700122, M220700123, M220700124, M220700125, M220700126, M220700128, M220700129, M220700130, M220700131, M220700132, M220700133, M220700134, M220700135, M220700136, M220700137, M220700138, M220700139, M220700142, M220700143, M220700144, M220700145, M220700146, M220700147, M220900017, M220900018, M220900021, M220900023, M220900024, M220900025, M220900026, M220900030, M220900031, M220900032, M220900033, M220900034, M220900035, M220900037, M220900038, M220900040, M220900042, M220900044, M220900045, M220900046, M220900047, M220900048, M220900049, M220900050, M220900052, M220900063, M220900065, M220900066, M220900067, M220900068, M220900069, M220900070, M220900071, M220900072, M220900073, M220900075, M220900076, M220900078, M220900079, M220900080, M220900081, M220900086, M220900087, M220900088, M220900089, M220900090, M220900091, M220900093, M220900094, M220900095, M220900096, M220900097, M220900101, M220900104, M220900106, M220900109, M221000023, M221000030, M221000032, M221000034, M221000035, M221000038, M221000039, M221000041, M221000042, M221000044, M221000047, M221000048, M221000049, M221000050, M221000052, M221100002, M221000046