Tri-Staple 2.0 recallZ-1228-2023 · Covidien, LP · initiated Jan 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.
FDA root cause category: Process change control
What the company is doing
On January 25, 2023, the firm notified customers via Urgent Medical Device Recall letter. Customers were asked to take the following immediate actions: 1. Identify and quarantined all unused affected product. 2. Return unused affected product to Medtronic using the Customer Confirmation Form. 3. Forward the information to all who need to be aware within your organization, or to any organization where the affected product was transferred. There are no additional actions required for patients where a stapler in scope of this recall was used during a procedure. These patients should continue to be monitored in accordance with your medical facility s standard care protocols. If you have any questions regarding this communication, please contact your Medtronic Representative Customer Service at 800-962-9888, option 2.
Recall details
- Recall number
- Z-1228-2023
- Status
- Open, Classified
- Initiated by the company
- Jan 25, 2023
- Posted by the FDA
- Mar 6, 2023
- Recalling firm
- Covidien, LP
- Quantity
- 264 US; 5056 OUS
- Product code
- GDW
- FDA submissions
- K160176
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Arkansas, California, Florida, Georgia, Louisiana, Massachusetts, New York, North Carolina, Ohio, Washington and the countries of Australia, Austria, Belgium, Canada, Canary Islands, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, New Caledonia, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Product description
Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT
Code information (models, lots, serial numbers)
UDI-DI (GTIN): 20884521543598 Lots: N2D0002Y, N2D0004Y, N2D0195Y
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Tri-Staple 2.0 SIG60AXTCovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths942 reports
Product lines
Other devices under the same FDA submission
31 devices cleared under K160176 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- Signia SIGADAPTSTNDCovidien LP · STAPLE, IMPLANTABLE1 death1,431 reports
- Signia SIGADAPTXLCovidien LP · STAPLE, IMPLANTABLE452 reports
- Tri-Staple 2.0 SIG45AXTCovidien LP · STAPLE, IMPLANTABLE342 reports
- Tri-Staple 2.0 SIGTRS60AMTCovidien LP · STAPLE, IMPLANTABLE1 death309 reports
- Tri-Staple 2.0 SIGTRS60AXTCovidien LP · STAPLE, IMPLANTABLE292 reports
- Signia SIGADAPTSHORTCovidien LP · STAPLE, IMPLANTABLE284 reports
- Tri-Staple 2.0 SIG30CTAVMCovidien LP · STAPLE, IMPLANTABLE3 deaths236 reports
- Signia SIGPSHELLCovidien LP · STAPLE, IMPLANTABLE1 death227 reports
- Tri-Staple 2.0 SIG45CTAVMCovidien LP · STAPLE, IMPLANTABLE2 deaths198 reports
- Tri-Staple 2.0 SIG30AVMCovidien LP · STAPLE, IMPLANTABLE3 deaths137 reports
- Tri-Staple 2.0 SIGRADMTCovidien LP · STAPLE, IMPLANTABLE118 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page