HeartSine Samaritan PAD 350P recallZ-1217-2023 · HeartSine Technologies Ltd · initiated Feb 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from HeartSine Technologies Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Automated external defibrillators were potentially distributed with the incorrect language configuration, which can lead to a delay in therapy or no therapy.
FDA root cause category: Process control
What the company is doing
On 2/1/23, correction notices were mailed to customers who were asked to do the following: 1) Confirm that the device has incorrect language voice prompts by using the Device Language Check. 2) If the device is found to have incorrect language configuration, remove it form use. The recalling firm will provide a replacement device. 3) Inform the recalling firm of any of the subject devices have been distributed to other organizations. 4) Complete and return the acknowledgement form to [email protected] If you have questions or concerns contact Customer Support at +44 28 9093 9400, 9:00 A.M. to 5:00 P.M. (GMT), Monday Friday or by email at [email protected]. Language Check: 1) Place device face up on a flat surface and slide the PAD-Pak into the device until you hear a "double click". 2) Verify the green status indicator is blinking. 3) Press the On/Off button. Liston for, but do not follow the voice prompts to ensure no warning messages are played and that the device prompts are in the expected language. 4) Press On/Off button to turn off the device.
Recall details
- Recall number
- Z-1217-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 1, 2023
- Posted by the FDA
- Mar 3, 2023
- Recalling firm
- HeartSine Technologies Ltd
- Quantity
- 13
- Product code
- MKJ
- FDA submissions
- P160008
- Distribution
- US: NJ, PA, WI, NE, TX, PA, CA, AZ, AR OUS: Australia, Czechia, Canada, New Zealand
Product description
Heartsine Samaritan PAD: SAM 350P semi-automatic defibrillator, SAM 360P fully automatic defibrillator, SAM 500P semi-automatic defibrillator with CPR Advisor
Code information (models, lots, serial numbers)
REF/UDI-DI/Serial Numbers: SAM 350P/05060167120671/17D00006810, 19D00027547, 19D00030364, 20D00000454, 20D00000868, 20D00006232, 22D91125401, 16D01000729, 22D91145062; SAM 360P/5060167127120, 5060167128028/20E90006314, 21E91040881; SAM 500P/5060167126505, 5060167125843/17B00003113, 20B90000080
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HeartSine Samaritan PAD 350P SAM 350PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)3 recalls2 deaths83 reports
Product lines
Other devices under the same FDA submission
15 devices cleared under P160008 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Pad-Pak Pad-Pak-01Heartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall529 reports
- HeartSine Samaritan PAD 450P SAM450PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)2 recalls2 deaths23 reports
- HeartSine Samaritan PAD 350P SAM 350PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)2 recalls0 reports
- HeartSine Samaritan PAD 360P SAM 360PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall19 reports
- HeartSine Samaritan PAD 350P SAM 350PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 360P SAM360PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM450PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 350P SAM 350PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 360P SAM 360PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM450PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 360P SAM 360PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
- HeartSine Samaritan PAD 450P SAM 450PHeartsine Technologies Ltd · Automated external defibrillators (non-wearable)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page