Skip to content

Power-PRO 2 . Intended as an ambulance cot to transport a… recallZ-1256-2023 · Stryker Medical Division of Stryker Corporation · initiated Feb 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Medical Division of Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.

FDA root cause category: Under Investigation by firm

What the company is doing

Stryker issued Urgent Medical Device Correction letter dated 2/1/23. Letter states reason for recall, health risk and action to take: 1. Locate the units listed on the attached business reply form and remove them from service until repairs are complete. 2. Return the enclosed business reply form to confirm receipt of this notification by fax +1 269 488 8691 or email [email protected]. 3. Stryker will contact you to arrange for the modification/repair of your Power-PRO 2. 4. If you have loaned or sold or disposed of any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. Questions or concerns, contact Customer Service at +1 800 327 0770 Monday through Friday from 8:00 a.m. to 6:00 p.m. ET.

Recall details

Recall number
Z-1256-2023
Status
Open, Classified
Initiated by the company
Feb 1, 2023
Posted by the FDA
Mar 15, 2023
Recalling firm
Stryker Medical Division of Stryker Corporation
Quantity
4 units
Product code
FPO
Distribution
US Nationwide distribution in the states of NY, VA.

Product description

Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001

Code information (models, lots, serial numbers)

GTIN: 07613327559118 Serial Numbers: 2208002066, 2208002071, 2209001508, 2208001664

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page