Monaco recallZ-1233-2023 · Elekta Inc · initiated Feb 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Re-optimization, after adding contours without forced density outside the external structure, may result in inaccurate dose presentation.

FDA root cause category: Software design

What the company is doing

Elekta notified consignees of the issue on 02/28/2023 via email. Consignees were instructed to contact Elekta Care Support if assistance is needed to identify affected patients, post the recall information in a place accessible to all users until the action is complete, and complete and return the Important Field Safety Notice Acknowledgement form. Additionally, the issue is corrected in Monaco releases 5.40 and above. If you are running a 5.11.xx version, please contact Elekta Care Support to upgrade to a newer software version.

Recall details

Recall number
Z-1233-2023
Status
Open, Classified
Initiated by the company
Feb 28, 2023
Posted by the FDA
Mar 8, 2023
Recalling firm
Elekta Inc
Quantity
2,020 units
Product code
MUJ
FDA submissions
K151233
Distribution
Nationwide distribution in the US. International distribution to Albania, Algeria, Antigua and Barbuda, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Occupied, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Zimbabwe.

Product description

Monaco RTP System. Used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Code information (models, lots, serial numbers)

Software Build: 5.11.00, 5.11.01, 5.11.02, 5.11.03; UDI Numbers: (01)00858164002169(10)5.11.00, (01)00858164002169(10)5.11.01, (01)00858164002169(10)5.11.02, (01)00858164002169(10)5.11.03