BioPlex 2200 recallZ-1378-2023 · Bio-Rad Laboratories, Inc. · initiated Feb 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bio-Rad Laboratories, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8 96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1 98.8%). The reduction in specificity may lead to an increase in the number of false positives.There is no performance issue with the calibrator lot# 53727. They are included in the recall only because the calibrators are lot matched to the impacted reagent pack.

FDA root cause category: Under Investigation by firm

What the company is doing

On February 21, 2023, Bio-Rad Laboratories issued a "Urgent: Medical Device Recall" notification to affected consignees. In addition to informing consignees about the recall, Bio-Rad asked consignees to take the following actions: 1. We are requesting that customers discontinue the use of and dispose of according to local waste management procedures any unused BioPlex 2200 Anti-CCP Reagent Pack (Catalog No. 665-3250, UDI GTIN 00847865000857) Lot No. 301481 and BioPlex 2200 Anti-CCPCalibrator Set (Catalog No. 663-3200, UDI GTIN 00847865000819) Lot No. 53727. 2. If you have distributed or transferred this product to any other Bio-Rad customers, or to any other laboratories from your site, please notify those customers of this field action. 3. Please complete the attached Customer Response Form and return it to your local Bio-Rad Technical Support so they can assist you with obtaining replacement reagent packs and calibrator sets Bio-Rad is working to provide a new lot of BioPlex 2200 Anti-CCP Reagent Pack and matching calibrator sets as soon as possible. Please provide the number of replacement units required in the attached customer response form.

Recall details

Recall number
Z-1378-2023
Status
Open, Classified
Initiated by the company
Feb 21, 2023
Posted by the FDA
Apr 12, 2023
Recalling firm
Bio-Rad Laboratories, Inc.
Quantity
225 units
Product code
NHX
FDA submissions
K093954, K112810
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Czech Republic, Germany, Spain, France, United Kingdom, Israel, Italy, Norway, Sweden.

Product description

BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD

Code information (models, lots, serial numbers)

Lot # 53727; UDI GTIN: 00847865000819

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 91854).