Why it was recalled

Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure

FDA root cause category: Under Investigation by firm

What the company is doing

AngioDynamics issued Urgent Medical Device Recall Letter on March 2, 2023 via Federal Express. Letter states reason for recall health risk and action to take: IMMEDIATELY Stop using the product subject to recall. Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location. Segregate this product in a secure location for return to AngioDynamics, Inc. Forward a copy of this recall notification to all sites to which you have distributed affected product. 2. ComplFx Reply Verification Tracking Form: Attn: NK Probe 5-Pack Recall Coordinator Fax number 1-855-273-0519 3. Package and Return the Recalled Product. Package any product that is being returned in an appropriate shipping box. Write the RMA number on the RMA/Address label (provided on the Recall Verification Tracking Form) and affix the label to the outside of the shipping box. Seal the box and return to: AngioDynamics, Inc. 24 Native Drive Queensbury, NY 12804 Attn: NK Probe 5-Pack Recall Coordinator

Recall details

Recall number
Z-1381-2023
Status
Open, Classified
Initiated by the company
Mar 2, 2023
Posted by the FDA
Apr 12, 2023
Recalling firm
Angiodynamics, Inc.
Quantity
10 boxes( 5 probes/box)
Product code
OAB
FDA submissions
K183385
Distribution
US Nationwide Distribution: CA,CO, FL, IA, NY, OH

Product description

NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN: H787204001110

Code information (models, lots, serial numbers)

UDI-DI: 15051684029643 Lot Number: 5762110

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183385 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.