Hamilton-C6, Ref: 160021 recallZ-1429-2023 · Hamilton Medical AG · initiated Mar 14, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

FDA root cause category: Software design

What the company is doing

On 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to [email protected] 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Technical Support at 1-800-426-6331, option #2, or email [email protected]

Recall details

Recall number
Z-1429-2023
Status
Open, Classified
Initiated by the company
Mar 14, 2023
Posted by the FDA
Apr 25, 2023
Recalling firm
Hamilton Medical AG
Quantity
103
Product code
CBK
FDA submissions
K201658
Distribution
US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

Product description

HAMILTON-C6, REF: 160021

Code information (models, lots, serial numbers)

Software Version: 1.2.1, UDI-DI: 07630002808590