Skip to content

STA R Max fully automatic clinical instrument designed to… recallZ-1575-2023 · Diagnostica Stago, Inc. · initiated Mar 22, 2023

Class II Open, Classified All recalls posted in May 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased frequency of occurrence of a technical error reported by the analyzer (error 13).

FDA root cause category: Device Design

What the company is doing

Customers were contacted via telephone and an URGENT: MEDICAL DEVICE CORRECTION notification letter beginning 3/22/23. Actions: Stago downgraded your firmware to the previous firmware version as a precaution. Your downgrade has already occurred. Consequently, the risk of this bug only existed during the time period shown above (i.e. from new firmware installation date to the downgrade date). As patient results are interpreted in association with other biological tests, we leave the decision to review previous patient results based on the complete clinical context to your discretion. Our Customer Service team is available to help you carry out your impact analysis. For additional information, please contact your Stago Hotline at 1-800-725-0607.

Recall details

Recall number
Z-1575-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Mar 22, 2023
Posted by the FDA
May 10, 2023
Recalling firm
Diagnostica Stago, Inc.
Quantity
6 analyzers
Product code
JPA
FDA submissions
K151867
Distribution
US Nationwide distribution in the states of IL, MA, MN, NE, TX, and WI.

Product description

STA R Max fully automatic clinical instrument designed to perform tests on human plasmas, Reference Numbers 59013 and 59026

Code information (models, lots, serial numbers)

UDI-DI: 036074505901; Serial Numbers: CP86010393 DB68093186 CP87010814 CP86070669 CP86090743 CP86090744

Other devices under the same FDA submission

2 devices cleared under K151867 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.