Incisive CT recallZ-1585-2023 · Philips North America · initiated Mar 20, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
FDA root cause category: Under Investigation by firm
What the company is doing
Philips Medical issued Urgent Medical Device Correction Letter Multi-Function Switch Unload Pedal Foot Entrapment on 3/20/23 to US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on your system. " Instead of using the Unload Pedal, please use the Unload Patient function on the Gantry Control Panels, or use other Table In/Out, Up/Down functions on the Gantry Control Panels, CT Scan Control Box, or Hardware lnterventional Controls to fulfill the normal patient unloading function, as instructed in the IFU. " You may continue to use Load Pedal, and Free-Float Pedal of Foot Switch, as they are not affected by this Urgent Medical Device Correction Letter. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve this issue (refer to FCO72800794). If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-1585-2023
- Status
- Open, Classified
- Initiated by the company
- Mar 20, 2023
- Posted by the FDA
- May 12, 2023
- Recalling firm
- Philips North America
- Quantity
- 10 units US; 36 units OUS
- Product code
- JAK
- FDA submissions
- K212441, K180015
- Distribution
- Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN, TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand, United Kingdom, Utd. Arab. Emir.
Product description
Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72814
Code information (models, lots, serial numbers)
UDI-DI: (01)00884838105508(21) + Serial number US Serial Numbers: 550243, 34111, 34092, 34066, 34064, 34047, 34042, 34025, 34031, 34033 OUS Serial Numbers: System Serial Number 550263 550249 550239 550215 550211 550091 550188 550187 550186 550174 550172 552041 550164 34107 550140 34062 550099 550082 550071 550068 550059 34020 550038 550040 550021 550036 550018 550014 550011 550010 550009 550008 550007 550240 550126 550241
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, X-Ray, Tomography, Computed9 recalls20 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K212441, K180015 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92117).
- Z-1586-2023 · Open, ClassifiedPhilips Incisive CT -Computed Tomography X-Ray System Model Number: 72813
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page