Skip to content

Intelli-C, Left side suspended Tabletop, REF: 03400010 recallZ-1669-2023 · Nrt X-Ray A/S · initiated Apr 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nrt X-Ray A/S or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

FDA root cause category: Device Design

What the company is doing

Between the dates of April 27, 2023 - May 1, 2023, NRT X-Ray A/S issued a "urgent: Medical Device Correction" via E-Mail to affected consignees. Per the customer communication, NRT ask consignees to take the following actions: Step 1 Short term action The first step of the recall is a corrective action which requires the existing fixation bracket components to be removed. Service Providers of the Intelli-C systems should follow the instructions in page 4 through 7 of the communication and report back to NRT using the Customer Response Card on page 8. The process is coordinated through US dealer Alpha Imaging in Willoughby, Ohio. Step 2 Long term action The second step is a preventive action to replace the IDE drive shaft and relating components. A separate set of instructions for this action will be provided in May together with the required parts for the replacement. These instructions will include a separate acknowledgement form for this action. If you have questions regarding the contents of this letter, please contact your local Service provider or contact NRT X-RAY A/S directly using the following contact information: Email: [email protected] Phone: +45 2055 3309 Monday through Friday, (03:00)-11:00 Eastern Standard Time (9:00 AM to 5:00 PM, Central European Time)

Recall details

Recall number
Z-1669-2023
Status
Open, Classified
Initiated by the company
Apr 27, 2023
Posted by the FDA
May 31, 2023
Recalling firm
Nrt X-Ray A/S
Quantity
13 units
Product code
OWB
FDA submissions
K173631
Distribution
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.

Product description

Intelli-C, Left side suspended Tabletop, REF: 03400010

Code information (models, lots, serial numbers)

UDI-DI:05713464000121/Serial Numbers: 647009, 651002, 651006, 666009, 682003

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K173631 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92230).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page