Intelli-C, Left side suspended Tabletop, REF: 03400010 recallZ-1669-2023 · Nrt X-Ray A/S · initiated Apr 27, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nrt X-Ray A/S or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
FDA root cause category: Device Design
What the company is doing
Between the dates of April 27, 2023 - May 1, 2023, NRT X-Ray A/S issued a "urgent: Medical Device Correction" via E-Mail to affected consignees. Per the customer communication, NRT ask consignees to take the following actions: Step 1 Short term action The first step of the recall is a corrective action which requires the existing fixation bracket components to be removed. Service Providers of the Intelli-C systems should follow the instructions in page 4 through 7 of the communication and report back to NRT using the Customer Response Card on page 8. The process is coordinated through US dealer Alpha Imaging in Willoughby, Ohio. Step 2 Long term action The second step is a preventive action to replace the IDE drive shaft and relating components. A separate set of instructions for this action will be provided in May together with the required parts for the replacement. These instructions will include a separate acknowledgement form for this action. If you have questions regarding the contents of this letter, please contact your local Service provider or contact NRT X-RAY A/S directly using the following contact information: Email: [email protected] Phone: +45 2055 3309 Monday through Friday, (03:00)-11:00 Eastern Standard Time (9:00 AM to 5:00 PM, Central European Time)
Recall details
- Recall number
- Z-1669-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 27, 2023
- Posted by the FDA
- May 31, 2023
- Recalling firm
- Nrt X-Ray A/S
- Quantity
- 13 units
- Product code
- OWB
- FDA submissions
- K173631
- Distribution
- US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Product description
Intelli-C, Left side suspended Tabletop, REF: 03400010
Code information (models, lots, serial numbers)
UDI-DI:05713464000121/Serial Numbers: 647009, 651002, 651006, 666009, 682003
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Intelli-C Left Side Suspended Table 03400010Nrt - Nordisk Røntgen Teknik A/S · System, X-Ray, Fluoroscopic, Image-Intensified2 recalls1 report
Other devices under the same FDA submission
3 devices cleared under K173631 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Intelli-C Right Side Suspended Table 03400000Nrt - Nordisk Røntgen Teknik A/S · Interventional Fluoroscopic X-Ray System2 recalls0 reports
- Intelli-C Accessory 03282000Nrt - Nordisk Røntgen Teknik A/S · Interventional Fluoroscopic X-Ray System1 recall1 report
- Intelli-C Accessory 03283000Nrt - Nordisk Røntgen Teknik A/S · Interventional Fluoroscopic X-Ray System1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92230).
- Z-1668-2023 · Open, ClassifiedIntelli-C, Right side suspended Tabletop, REF: 03400000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page