Skip to content

Intelli-C Accessory 03282000Recalls & reported problems

Nrt - Nordisk Røntgen Teknik A/S · Footrest for the Intelli-C X-ray system to support the patient during table top angulation. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05713464000336), received between Jan 21, 2022 and Jan 21, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Slipped” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K173631); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 21, 2022Open, Classified

    Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.

    Z-0526-2022 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Device Slipped1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2022Device Slipped

    IT WAS REPORTED THAT THE FOOTREST SLID DOWN ALONG THE PATIENT TABLE WHILE IN USE WITH A PATIENT. THE PATIENT GRABBED THE PATIENT TABLE AND DID NOT FALL.

    Patient problems: Insufficient Information

    MAUDE report 13318352 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K173631. One submission often covers many models, so check the product and lot details in each record.

  • Z-1784-2024 · Apr 17, 2024 · Open, Classified
    Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Intelli-C X-ray system
  • Z-1669-2023 · Apr 27, 2023 · Open, Classified
    Intelli-C, Left side suspended Tabletop, REF: 03400010
  • Z-1668-2023 · Apr 27, 2023 · Open, Classified
    Intelli-C, Right side suspended Tabletop, REF: 03400000

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.