Intelli-C Left Side Suspended Table 03400010Recalls & reported problems
Nrt - Nordisk Røntgen Teknik A/S · The Intelli-C X-ray device is a multi-purpose - also sometimes known as a universal C-arm fluoroscopy system, designed for diagnostic medical imaging examinations, from barium studies to interventional and angiography studies. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 05713464000121), received between Feb 28, 2023 and Feb 28, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
2 other recalls cover devices cleared under the same FDA submission (K173631); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Apr 17, 2024Class IIOpen, Classified
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
Z-1784-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Apr 27, 2023Class IIOpen, Classified
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Z-1669-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 28, 2023Break; Break
DURING AN EXAMINATION OF A PATIENT, IT WAS FOUND THAT A DRIVE SHAFT IN THE IDE MOVEMENT OF THE INTELLI-C X-RAY DEVICE HAD BROKEN. THE IDE MOVEMENT THAT DRIVES THE DETECTOR INTO POSITION, MOVED UNTIL IT WAS STOPPED BY THE MOVEMENT END STOP. THE EVENT DID NOT CAUSE ANY INJURY TO PATIENT OR OPERATOR AS NO ONE WAS HIT BY THE UNINTENDED MOVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16452169 ↗
FDA clearance and classification
Cleared through 510(k) K173631 on Jul 11, 2018 (decision: Substantially Equivalent). Submitted by Nrt X-Ray A/S as “Intelli-C”. The FDA review took 229 days. FDA record ↗
The FDA regulates product code JAA (system, x-ray, fluoroscopic, image-intensified) under 21 CFR 892.1650 as a Class II device. 21 CFR 892.1650 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K173631. One submission often covers many models, so check the product and lot details in each record.
- Z-1668-2023 · Apr 27, 2023 · Open, ClassifiedIntelli-C, Right side suspended Tabletop, REF: 03400000
- Z-0526-2022 · Jan 21, 2022 · Open, Classified1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)
Common questions about the Intelli-C Left Side Suspended Table 03400010
Has the Intelli-C Left Side Suspended Table 03400010 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-1784-2024 (Class II), began on Apr 17, 2024: X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Intelli-C Left Side Suspended Table 03400010?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 28, 2023. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Intelli-C Left Side Suspended Table 03400010?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Intelli-C Left Side Suspended Table 03400010 FDA approved?
It was cleared through 510(k) K173631 on Jul 11, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Intelli-C Left Side Suspended Table 03400010, and is it still sold?
The FDA's device database lists Nrt - Nordisk Røntgen Teknik A/S as the company responsible for it (brand Intelli-C Left Side Suspended Table, model 03400010). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, X-Ray, Fluoroscopic, Image-Intensified devices
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified9 recalls7 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified7 recalls1 report
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · System, x-ray, fluoroscopic, image-intensified6 recalls4 reports
- INTERVENTIONAL ANGIOGRAPHY SYSTEM INFX-8000CCanon Inc. · System, x-ray, fluoroscopic, image-intensified5 recalls5 reports
- ProxiDiagnost N90 706110Philips Medical Systems DMC GmbH · System, X-Ray, Fluoroscopic, Image-Intensified5 recalls1 report
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- CombiDiagnost R90 709031Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls10 reports
- ProxiDiagnost N90 706100Philips Medical Systems DMC GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page