Incisive CT recallZ-1623-2023 · Philips Headquarters Cambridge · initiated Apr 12, 2023
Class II Open, Classified All recalls posted in May 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Headquarters Cambridge or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required
FDA root cause category: Software design
What the company is doing
Philips issued Urgent Medical Device Correction Letters on 4/12/23 US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Refer to the IFU of your Incisive CT system to ensure proper use of the Precise Position function (IFU Section 4.9). " If the patient orientation automatically selected by Precise Position (see Figure 2) is inconsistent with the preset protocol, please correct the settings as described below: o Manually select the correct orientation in the Select Patient Orientation area in the New Patient interface, or o Open the Select Patient Orientation list by clicking the patient orientation icon in the left upper corner of the series list and select the correct orientation. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 5. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a software upgrade to address the issue (reference FCO72800793). Additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-1623-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Apr 12, 2023
- Posted by the FDA
- May 23, 2023
- Recalling firm
- Philips Headquarters Cambridge
- Quantity
- 22 units
- Product code
- JAK
- FDA submissions
- K212441, K180015
- Distribution
- AR, CA, OK, TX Foreign: Country Argentina Australia Austria China Colombia Czech Republic Dominican Rep France Germany Guadeloupe Hong Kong India Japan Kuwait Latvia Lithuania Netherlands Norway Portugal Romania Spain Switzerland Taiwan Thailand United Kingdom Vietnam
Product description
Incisive CT- Whole-body computed tomography (CT) X-Ray System Model Number: 728143 (OUS)
Code information (models, lots, serial numbers)
UDI-DI: (01)00884838085015(21) Serial Numbers: 75592542 81629039 82435668 84375501 85146962 85469672 86188405 86188817 86820074 87367597 92174932 92622328 92637836 92680372 92823844 92870027 92889235 94581069 95050209 95055576 96094114 96462378 Serial Numbers: 75592542 81629039 82435668 84375501 85146962 85469672 86188405 86188817 86820074 87367597 92174932 92622328 92637836 92680372 92823844 92870027 92889235 94581069 95050209 95055576 96094114 96462378
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92238).
- Z-1622-2023 · Open, ClassifiedIncisive CT- whole-body computed tomography (CT) X-Ray System Model Number: 728144
- Z-1624-2023 · Open, ClassifiedIncisive CT Power- Whole-body computed tomography (CT) X-Ray System Model Number: 728148 (OUS)
- Z-1625-2023 · Open, ClassifiedIncisive CT Plus -Whole-body computed tomography (CT) X-Ray System Model Number: 728149 (OUS)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page