Elekta Medical Linear Accelerator recallZ-1665-2023 · Elekta, Inc. · initiated May 5, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A component within a limited number of Elekta Linacs has been realized to be non-compliant to IEC 60601

FDA root cause category: Under Investigation by firm

What the company is doing

Elekta notified customers, via email, on 05/05/2023. Customers were instructed that Elekta will contact each impacted customer to perform an onsite inspection of the power supply unit and replace as needed. Customers were asked to complete and return the Acknowledgement Form.

Recall details

Recall number
Z-1665-2023
Status
Open, Classified
Initiated by the company
May 5, 2023
Posted by the FDA
May 30, 2023
Recalling firm
Elekta, Inc.
Quantity
46 units
Product code
IYE
FDA submissions
K210500
Distribution
Worldwide distribution. US distribution to AR, IN, LA, OK, TN, VA, WA. International distribution to Belgium, Canada, Chile, China, France, Germany, India, Iraq, Italy, Jordan, Libyan Arab Jamahiriya, Mauritius, Mexico, Morocco, Poland, Russian Federation, Spain.

Product description

Elekta Medical Linear Accelerator (ELMA) included in the below platforms: a. Elekta Infinity; b. Versa HD; c. Elekta Synergy

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Varex Part Number 95510354H-01 (Varex serial numbers 354-01H_A0010004PC to 354-01H_A0010269PC). a. Elekta Infinity: UDI-DI Number: 05060191071543, 05060191071512 (Serial Numbers: 156073, 152223, 153681); b. Versa HD: UDI-DI Number: 05060191071574, 05060191071529 (Serial Numbers: 156941, 156989, 156724, 156998); c. Elekta Synergy: UDI-DI Number: 05060191071550, 05060191071505 (Serial Number: 152483) Additional serial numbers distributed OUS: 109331, 109333, 109335, 109336, 109339, 109341, 109342, 151879, 152194, 153038, 153299, 153410, 153970, 156841, 156849, 156966, 156968, 156969, 156970, 156971, 156973, 156977, 156978, 156979, 156980, 156981, 156982, 156983, 156984, 156985, 156986, 156987, 156988, 156990, 156992, 156995, 156996, 156997.