Visia AF MRI S VR SureScan recallZ-1807-2023 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated May 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 05/10/2023 by UPS 2-day delivery or email. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. The notice explained the problem and the risk and requested the following patient management recommendations: - Prophylactic device replacement is NOT recommended. - Program all HV therapy pathways B>AX in all therapy zones to minimize the risk for this issue. - Prioritize reprogramming patients who have both a history of HV therapy and Rx1 programmed AX>B. - For remaining patients with AX>B programming in any HV therapy sequence, schedule (with appropriate discretion) the next follow-up for in-clinic reprogramming to minimize potential for reduced- or no-energy HV therapies to occur. - Contact Medtronic Technical Services (1-800-929-4043) or your local representative if one of the following is observed as these may be an indication of either a device or lead-related issue: o Reduced- or no-energy HV therapy is displayed in Episode Text (regardless of programmed pathway) o A persistent drop of approximately 50% in RA, RV and LV pacing lead impedance measurements as this may be an indication of increased potential for a future reduced- or no-energy therapy. Update 10/10/2023: Beginning 09/27/2023, Medtronic issued communication via letter to consignees regarding the availability of the revised IFU and SmartSync Software. Consignees are asked to forward this notification to those who need to be aware within the organization and to any location which the devices have been distributed. Consignees are asked to complete and return a confirmation certificate (or equivalent record) confirming they have received the notification.
Recall details
- Recall number
- Z-1807-2023
- Status
- Open, Classified
- Initiated by the company
- May 10, 2023
- Posted by the FDA
- Jun 16, 2023
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
- Quantity
- 5622 units
- Product code
- LWS
- FDA submissions
- P980016S551
- Distribution
- US Nationwide - Worldwide Distribution
Product description
ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibrillators
Code information (models, lots, serial numbers)
Show the full code information (68k characters)
GTIN 00643169837812, Lot Serial numbers: CWJ604194S, CWJ604707S, CWJ604057S, CWJ602128S, CWJ604288S, CWJ602870S, CWJ601914S, CWJ603587S, CWJ603802S, CWJ601000S, CWJ601001S, CWJ601158S, CWJ601706S, CWJ601799S, CWJ601845S, CWJ604343S, CWJ604387S, CWJ601910S, CWJ603379S, CWJ603449S, CWJ603454S, CWJ603109S, CWJ603385S, CWJ603717S, CWJ604036S, CWJ604043S, CWJ604092S, CWJ604118S, CWJ604397S, CWJ603165S, CWJ603501S, CWJ603684S, CWJ603685S, CWJ603693S, CWJ603831S, CWJ604218S, CWJ604314S, CWJ604338S, CWJ601875S, CWJ601903S, CWJ603686S, CWJ603836S, CWJ603869S, CWJ603875S, CWJ604377S, CWJ601894S, CWJ601905S, CWJ601909S, CWJ603429S, CWJ603660S, CWJ603060S, CWJ603731S, CWJ603760S, CWJ604336S, CWJ604355S, CWJ603859S, CWJ603697S, CWJ603700S, CWJ603849S, CWJ604337S, CWJ603687S, CWJ603695S, CWJ601906S, CWJ603445S, CWJ603873S, CWJ604179S, CWJ603842S, CWJ604042S, CWJ604094S, CWJ604339S, CWJ604358S, CWJ604117S, CWJ601891S, CWJ602956S, CWJ603275S, CWJ604684S, CWJ603881S, CWJ603884S, CWJ603939S, CWJ603957S, CWJ604167S, CWJ604178S, CWJ600977S, CWJ601408S, CWJ601788S, CWJ602434S, CWJ602745S, CWJ602880S, CWJ603241S, CWJ603297S, CWJ604580S, CWJ602700S, CWJ603061S, CWJ603107S, CWJ600881S, CWJ601208S, CWJ601611S, CWJ602039S, CWJ602115S, CWJ602503S, CWJ603723S, CWJ600971S, CWJ600974S, CWJ601067S, CWJ602190S, CWJ601175S, CWJ602150S, CWJ601675S, CWJ601315S, CWJ601498S, CWJ601661S, CWJ601714S, CWJ603486S, CWJ603828S, CWJ601669S, CWJ601686S, CWJ601724S, CWJ601911S, CWJ602067S, CWJ602106S, CWJ602168S, CWJ602176S, CWJ602271S, CWJ602425S, CWJ602629S, CWJ602701S, CWJ602702S, CWJ602725S, CWJ602821S, CWJ602881S, CWJ602895S, CWJ602899S, CWJ602963S, CWJ602984S, CWJ603144S, CWJ603150S, CWJ603254S, CWJ603293S, CWJ603313S, CWJ603328S, CWJ603368S, CWJ603369S, CWJ603370S, CWJ603416S, CWJ603467S, CWJ603471S, CWJ603491S, CWJ603610S, CWJ603621S, CWJ603623S, CWJ603646S, CWJ603774S, CWJ603786S, CWJ603805S, CWJ603848S, CWJ603882S, CWJ603887S, CWJ603894S, CWJ604025S, CWJ604054S, CWJ604107S, CWJ604187S, CWJ604196S, CWJ601206S, CWJ601873S, CWJ602005S, CWJ602130S, CWJ602245S, CWJ602771S, CWJ602832S, CWJ602962S, CWJ603893S, CWJ601936S, CWJ602424S, CWJ604412S, CWJ604480S, CWJ602820S, CWJ602244S, CWJ602409S, CWJ602568S, CWJ602711S, CWJ601798S, CWJ603058S, CWJ601131S, CWJ601132S, CWJ601133S, CWJ602566S, CWJ604230S, CWJ604413S, CWJ604447S, CWJ602252S, CWJ602492S, CWJ602583S, CWJ601003S, CWJ601668S, CWJ602043S, CWJ602232S, CWJ603068S, CWJ600874S, CWJ601002S, CWJ601253S, CWJ601268S, CWJ601420S, CWJ601433S, CWJ601663S, CWJ601771S, CWJ601774S, CWJ601846S, CWJ601851S, CWJ601877S, CWJ601883S, CWJ602494S, CWJ602567S, CWJ602655S, CWJ603033S, CWJ603070S, CWJ603299S, CWJ603378S, CWJ603494S, CWJ603502S, CWJ603505S, CWJ603804S, CWJ604596S, CWJ604691S, CWJ604702S, CWJ603421S, CWJ603793S, CWJ603798S, CWJ604228S, CWJ604278S, CWJ604604S, CWJ603148S, CWJ603532S, CWJ602266S, CWJ602295S, CWJ600878S, CWJ601872S, CWJ603982S, CWJ601852S, CWJ602355S, CWJ602729S, CWJ602442S, CWJ603038S, CWJ603308S, CWJ603533S, CWJ603644S, CWJ603778S, CWJ603899S, CWJ604052S, CWJ604232S, CWJ604291S, CWJ601166S, CWJ603779S, CWJ604344S, CWJ604475S, CWJ603784S, CWJ604540S, CWJ602877S, CWJ602940S, CWJ600885S, CWJ601044S, CWJ601248S, CWJ601540S, CWJ601789S, CWJ602042S, CWJ602069S, CWJ602605S, CWJ602606S, CWJ602707S, CWJ602853S, CWJ602941S, CWJ603821S, CWJ604161S, CWJ603694S, CWJ603847S, CWJ604046S, CWJ604047S, CWJ604225S, CWJ604281S, CWJ604463S, CWJ600889S, CWJ602497S, CWJ603347S, CWJ603090S, CWJ603479S, CWJ603495S, CWJ601209S, CWJ601587S, CWJ601238S, CWJ601240S, CWJ602413S, CWJ603708S, CWJ603979S, CWJ603387S, CWJ601916S, CWJ601406S, CWJ602900S, CWJ602943S, CWJ602052S, CWJ601547S, CWJ601649S, CWJ602007S, CWJ603386S, CWJ601650S, CWJ602112S, CWJ603103S, CWJ601405S, CWJ603094S, CWJ604162S, CWJ601005S, CWJ601900S, CWJ601791S, CWJ602769S, CWJ603896S, CWJ604304S, CWJ604444S, CWJ601160S, CWJ601411S, CWJ601461S, CWJ601648S, CWJ601822S, CWJ602175S, CWJ602231S, CWJ602633S, CWJ602636S, CWJ602845S, CWJ602931S, CWJ603089S, CWJ603173S, CWJ603216S, CWJ603350S, CWJ602122S, CWJ602124S, CWJ600966S, CWJ600991S, CWJ601015S, CWJ601080S, CWJ601086S, CWJ601183S, CWJ601403S, CWJ601439S, CWJ601441S, CWJ601508S, CWJ601578S, CWJ601603S, CWJ601621S, CWJ601634S, CWJ601653S, CWJ601777S, CWJ601792S, CWJ601847S, CWJ601884S, CWJ602011S, CWJ602033S, CWJ602068S, CWJ602208S, CWJ602228S, CWJ602435S, CWJ602443S, CWJ602505S, CWJ602554S, CWJ602571S, CWJ602572S, CWJ602574S, CWJ602592S, CWJ602637S, CWJ602645S, CWJ602669S, CWJ602719S, CWJ602720S, CWJ602935S, CWJ602942S, CWJ602992S, CWJ603039S, CWJ603047S, CWJ603078S, CWJ603095S, CWJ603174S, CWJ603292S, CWJ603551S, CWJ603617S, CWJ603628S, CWJ603813S, CWJ603814S, CWJ603880S, CWJ603908S, CWJ603978S, CWJ604172S, CWJ604214S, CWJ604310S, CWJ604471S, CWJ600869S, CWJ600965S, CWJ601045S, CWJ601460S, CWJ601638S, CWJ601986S, CWJ602061S, CWJ602088S, CWJ602411S, CWJ602483S, CWJ602518S, CWJ602936S, CWJ603056S, CWJ603086S, CWJ603138S, CWJ603220S, CWJ603371S, CWJ603443S, CWJ603542S, CWJ602326S, CWJ603298S, CWJ602343S, CWJ600936S, CWJ601039S, CWJ601476S, CWJ601762S, CWJ603902S, CWJ604222S, CWJ600873S, CWJ601414S, CWJ601477S, CWJ602561S, CWJ603129S, CWJ603781S, CWJ604007S, CWJ601427S, CWJ602200S, CWJ602982S, CWJ603792S, CWJ601258S, CWJ601549S, CWJ602220S, CWJ602469S, CWJ604201S, CWJ604202S, CWJ601548S, CWJ602027S, CWJ602093S, CWJ602195S, CWJ604589S, CWJ602381S, CWJ602586S, CWJ602611S, CWJ601117S, CWJ601436S, CWJ601848S, CWJ604350S, CWJ604537S, CWJ603077S, CWJ601657S, CWJ602893S, CWJ600946S, CWJ603085S, CWJ603301S, CWJ601062S, CWJ603937S, CWJ603946S, CWJ603952S, CWJ603966S, CWJ604147S, CWJ604193S, CWJ604220S, CWJ604290S, CWJ604312S, CWJ604481S, CWJ602481S, CWJ602997S, CWJ604029S, CWJ604159S, CWJ604647S, CWJ603732S, CWJ603846S, CWJ604353S, CWJ604597S, CWJ604648S, CWJ603612S, CWJ602484S, CWJ603004S, CWJ604731S, CWJ603092S, CWJ603384S, CWJ601087S, CWJ601155S, CWJ601904S, CWJ601937S, CWJ602081S, CWJ602172S, CWJ602261S, CWJ602522S, CWJ602547S, CWJ602704S, CWJ602767S, CWJ603040S, CWJ603082S, CWJ603122S, CWJ603135S, CWJ603209S, CWJ603212S, CWJ603309S, CWJ603543S, CWJ603746S, CWJ602170S, CWJ601148S, CWJ602104S, CWJ602477S, CWJ601554S, CWJ602615S, CWJ603091S, CWJ603317S, CWJ603352S, CWJ603712S, CWJ601147S, CWJ601773S, CWJ602202S, CWJ602856S, CWJ601667S, CWJ602643S, CWJ603351S, CWJ603373S, CWJ603434S, CWJ603768S, CWJ603777S, CWJ604276S, CWJ604296S, CWJ604618S, CWJ604719S, CWJ604735S, CWJ604758S, CWJ601871S, CWJ601612S, CWJ603658S, CWJ603829S, CWJ603901S, CWJ604101S, CWJ604490S, CWJ601491S, CWJ603096S, CWJ602476S, CWJ602741S, CWJ601136S, CWJ601139S, CWJ601141S, CWJ601192S, CWJ601660S, CWJ601140S, CWJ601583S, CWJ601641S, CWJ601704S, CWJ601756S, CWJ601988S, CWJ602001S, CWJ601006S, CWJ601431S, CWJ601432S, CWJ601446S, CWJ601680S, CWJ601959S, CWJ602117S, CWJ602196S, CWJ602218S, CWJ602243S, CWJ602334S, CWJ602358S, CWJ602398S, CWJ602453S, CWJ602507S, CWJ602516S, CWJ602647S, CWJ602650S, CWJ602713S, CWJ602738S, CWJ602739S, CWJ602784S, CWJ602800S, CWJ602801S, CWJ602862S, CWJ602886S, CWJ602950S, CWJ603014S, CWJ603023S, CWJ603121S, CWJ603131S, CWJ603134S, CWJ603153S, CWJ603159S, CWJ603224S, CWJ603336S, CWJ603437S, CWJ603611S, CWJ603655S, CWJ603659S, CWJ603713S, CWJ603724S, CWJ603733S, CWJ603874S, CWJ603923S, CWJ603998S, CWJ604215S, CWJ604407S, CWJ604505S, CWJ604593S, CWJ604757S, CWJ601374S, CWJ601635S, CWJ601647S, CWJ601666S, CWJ601678S, CWJ601849S, CWJ603497S, CWJ602912S, CWJ603270S, CWJ603271S, CWJ603273S, CWJ604582S, CWJ604584S, CWJ601094S, CWJ601104S, CWJ602608S, CWJ602868S, CWJ603151S, CWJ604530S, CWJ602312S, CWJ603426S, CWJ602345S, CWJ603665S, CWJ604045S, CWJ601533S, CWJ604031S, CWJ602640S, CWJ602485S, CWJ604000S, CWJ601237S, CWJ601281S, CWJ603958S, CWJ602766S, CWJ603059S, CWJ603062S, CWJ603064S, CWJ603752S, CWJ602630S, CWJ602794S, CWJ602932S, CWJ602991S, CWJ602994S, CWJ603257S, CWJ603263S, CWJ603279S, CWJ603285S, CWJ603349S, CWJ603354S, CWJ603959S, CWJ601282S, CWJ604714S, CWJ602002S, CWJ603177S, CWJ604727S, CWJ601809S, CWJ602810S, CWJ602920S, CWJ603797S, CWJ601811S, CWJ602785S, CWJ603680S, CWJ603922S, CWJ600901S, CWJ601230S, CWJ601231S, CWJ601363S, CWJ601410S, CWJ601430S, CWJ601450S, CWJ601571S, CWJ601580S, CWJ601582S, CWJ601779S, CWJ601831S, CWJ601865S, CWJ602026S, CWJ602135S, CWJ602161S, CWJ602246S, CWJ602249S, CWJ602308S, CWJ602364S, CWJ602399S, CWJ602470S, CWJ602472S, CWJ602773S, CWJ602786S, CWJ602846S, CWJ602847S, CWJ602916S, CWJ602976S, CWJ603007S, CWJ603268S, CWJ603269S, CWJ603638S, CWJ603661S, CWJ603663S, CWJ603670S, CWJ603709S, CWJ603710S, CWJ601181S, CWJ601142S, CWJ603168S, CWJ603669S, CWJ604049S, CWJ604408S, CWJ602158S, CWJ602363S, CWJ602347S, CWJ602348S, CWJ602204S, CWJ602541S, CWJ602727S, CWJ603483S, CWJ604680S, CWJ601018S, CWJ600992S, CWJ601149S, CWJ601501S, CWJ601601S, CWJ602055S, CWJ602120S, CWJ602344S, CWJ602402S, CWJ602625S, CWJ602664S, CWJ602793S, CWJ602946S, CWJ602953S, CWJ603001S, CWJ603423S, CWJ603664S, CWJ603926S, CWJ604174S, CWJ604406S, CWJ601532S, CWJ601536S, CWJ601679S, CWJ601772S, CWJ602125S, CWJ602146S, CWJ602153S, CWJ602255S, CWJ602369S, CWJ602478S, CWJ602607S, CWJ603227S, CWJ603346S, CWJ603399S, CWJ600875S, CWJ600919S, CWJ600928S, CWJ600940S, CWJ600958S, CWJ600994S, CWJ601011S, CWJ601057S, CWJ601061S, CWJ601120S, CWJ601143S, CWJ601187S, CWJ601188S, CWJ601197S, CWJ601222S, CWJ601290S, CWJ601292S, CWJ601335S, CWJ601392S, CWJ601412S, CWJ601416S, CWJ601671S, CWJ601690S, CWJ601733S, CWJ601748S, CWJ602171S, CWJ602224S, CWJ602225S, CWJ602248S, CWJ602250S, CWJ602251S, CWJ602259S, CWJ602260S, CWJ602299S, CWJ602311S, CWJ602318S, CWJ602337S, CWJ602342S, CWJ602370S, CWJ602391S, CWJ602501S, CWJ602504S, CWJ602550S, CWJ602562S, CWJ602564S, CWJ602838S, CWJ602852S, CWJ602865S, CWJ602882S, CWJ602965S, CWJ602972S, CWJ602980S, CWJ602987S, CWJ603035S, CWJ603037S, CWJ603128S, CWJ603198S, CWJ603325S, CWJ603366S, CWJ603592S, CWJ603620S, CWJ603773S, CWJ603803S, CWJ603826S, CWJ602084S, CWJ602103S, CWJ603722S, CWJ601793S, CWJ601954S, CWJ602174S, CWJ601215S, CWJ601759S, CWJ600960S, CWJ601507S, CWJ604534S, CWJ604573S, CWJ604216S, CWJ604410S, CWJ604681S, CWJ604567S, CWJ600882S, CWJ600938S, CWJ602339S, CWJ602559S, CWJ602885S, CWJ603367S, CWJ604219S, CWJ604576S, CWJ602959S, CWJ603650S, CWJ603834S, CWJ603879S, CWJ602706S, CWJ602737S, CWJ602872S, CWJ603073S, CWJ600945S, CWJ601879S, CWJ602593S, CWJ601833S, CWJ601013S, CWJ601261S, CWJ603383S, CWJ602083S, CWJ602145S, CWJ604634S, CWJ602921S, CWJ603806S, CWJ603898S, CWJ604411S, CWJ601182S, CWJ601616S, CWJ601757S, CWJ601765S, CWJ602057S, CWJ602302S, CWJ602515S, CWJ603118S, CWJ603143S, CWJ603236S, CWJ603291S, CWJ604287S, CWJ604532S, CWJ604578S, CWJ603435S, CWJ603757S, CWJ604504S, CWJ603463S, CWJ603639S, CWJ600895S, CWJ600896S, CWJ600897S, CWJ601399S, CWJ601400S, CWJ601424S, CWJ601567S, CWJ601628S, CWJ601629S, CWJ601630S, CWJ601896S, CWJ601897S, CWJ601994S, CWJ601996S, CWJ601999S, CWJ602090S, CWJ602095S, CWJ602097S, CWJ602098S, CWJ602126S, CWJ602129S, CWJ602178S, CWJ602180S, CWJ602206S, CWJ602389S, CWJ602390S, CWJ602722S, CWJ602723S, CWJ603050S, CWJ603051S, CWJ603054S, CWJ603076S, CWJ603157S, CWJ603158S, CWJ603160S, CWJ603164S, CWJ603166S, CWJ603222S, CWJ603223S, CWJ603225S, CWJ603510S, CWJ603511S, CWJ603529S, CWJ603530S, CWJ603531S, CWJ603795S, CWJ604039S, CWJ604040S, CWJ604297S, CWJ604479S, CWJ604491S, CWJ604492S, CWJ604657S, CWJ600892S, CWJ600898S, CWJ600942S, CWJ600943S, CWJ600981S, CWJ600982S, CWJ601034S, CWJ601035S, CWJ601037S, CWJ601064S, CWJ601065S, CWJ601066S, CWJ601200S, CWJ601201S, CWJ601202S, CWJ601751S, CWJ601752S, CWJ601753S, CWJ601754S, CWJ601919S, CWJ601921S, CWJ601926S, CWJ601927S, CWJ601929S, CWJ601933S, CWJ601945S, CWJ602017S, CWJ602181S, CWJ602184S, CWJ602185S, CWJ602187S, CWJ602423S, CWJ602762S, CWJ602763S, CWJ602764S, CWJ603043S, CWJ603044S, CWJ603045S, CWJ603139S, CWJ603141S, CWJ603142S, CWJ603375S, CWJ603376S, CWJ603377S, CWJ603440S, CWJ603441S, CWJ603637S, CWJ603640S, CWJ603641S, CWJ603913S, CWJ603921S, CWJ603994S, CWJ604210S, CWJ604211S, CWJ604212S, CWJ604349S, CWJ604393S, CWJ604394S, CWJ604486S, CWJ604487S, CWJ601004S, CWJ601058S, CWJ603145S, CWJ602693S, CWJ602848S, CWJ602849S, CWJ602851S, CWJ602969S, CWJ602971S, CWJ603116S, CWJ603117S, CWJ603242S, CWJ603243S, CWJ603245S, CWJ603290S, CWJ603412S, CWJ603799S, CWJ603905S, CWJ604177S, CWJ604602S, CWJ600891S, CWJ600949S, CWJ600950S, CWJ600957S, CWJ601180S, CWJ601319S, CWJ601369S, CWJ601370S, CWJ601544S, CWJ601545S, CWJ601938S, CWJ602031S, CWJ602333S, CWJ602386S, CWJ602591S, CWJ603137S, CWJ603187S, CWJ603230S, CWJ603294S, CWJ603677S, CWJ603749S, CWJ603936S, CWJ604309S, CWJ604605S, CWJ603074S, CWJ603414S, CWJ603995S, CWJ603951S, CWJ602099S, CWJ603021S, CWJ604694S, CWJ604441S, CWJ602022S, CWJ602094S, CWJ602267S, CWJ602276S, CWJ602277S, CWJ602290S, CWJ602300S, CWJ602307S, CWJ602315S, CWJ602317S, CWJ602394S, CWJ602580S, CWJ602624S, CWJ602626S, CWJ602631S, CWJ602744S, CWJ603049S, CWJ603052S, CWJ603053S, CWJ603057S, CWJ603083S, CWJ603239S, CWJ603240S, CWJ603244S, CWJ603249S, CWJ603372S, CWJ603472S, CWJ603474S, CWJ603509S, CWJ603520S, CWJ603521S, CWJ603523S, CWJ603651S, CWJ603652S, CWJ603653S, CWJ603654S, CWJ603656S, CWJ603657S, CWJ603858S, CWJ603956S, CWJ604122S, CWJ604123S, CWJ604124S, CWJ604146S, CWJ604148S, CWJ604619S, CWJ604622S, CWJ604623S, CWJ604626S, CWJ604627S, CWJ604630S, CWJ604632S, CWJ604633S, CWJ604635S, CWJ604637S, CWJ604638S, CWJ604713S, CWJ604722S, CWJ604724S, CWJ601295S, CWJ601296S, CWJ601299S, CWJ601300S, CWJ601301S, CWJ601302S, CWJ601303S, CWJ601340S, CWJ601343S, CWJ601347S, CWJ601349S, CWJ601350S, CWJ601353S, CWJ601354S, CWJ601355S, CWJ601358S, CWJ601359S, CWJ601646S, CWJ601651S, CWJ601652S, CWJ601659S, CWJ601975S, CWJ602138S, CWJ602144S, CWJ602459S, CWJ602461S, CWJ602462S, CWJ604059S, CWJ604060S, CWJ600918S, CWJ601145S, CWJ601270S, CWJ601312S, CWJ601449S, CWJ601457S, CWJ601717S, CWJ602127S, CWJ602201S, CWJ602335S, CWJ602552S, CWJ602863S, CWJ602961S, CWJ603624S, CWJ603737S, CWJ602627S, CWJ601338S, CWJ601379S, CWJ601494S, CWJ601689S, CWJ602673S, CWJ602724S, CWJ604710S, CWJ603314S, CWJ601052S, CWJ603863S, CWJ604696S, CWJ602379S, CWJ601524S, CWJ602137S, CWJ600988S, CWJ601563S, CWJ601713S, CWJ602177S, CWJ602197S, CWJ602199S, CWJ602268S, CWJ602310S, CWJ602822S, CWJ603140S, CWJ603725S, CWJ603800S, CWJ603832S, CWJ603911S, CWJ603938S, CWJ603943S, CWJ604001S, CWJ604176S, CWJ604213S, CWJ604585S, CWJ604610S, CWJ604612S, CWJ604613S, CWJ601083S, CWJ601244S, CWJ601688S, CWJ604184S, CWJ604629S, CWJ602427S, CWJ603589S, CWJ603890S, CWJ603907S, CWJ603726S, CWJ603912S, CWJ604051S, CWJ602835S, CWJ603755S, CWJ603614S, CWJ601346S, CWJ601328S, CWJ602526S, CWJ603920S, CWJ604717S, CWJ604718S, CWJ604711S, CWJ601119S, CWJ601565S, CWJ603081S, CWJ603844S, CWJ600972S, CWJ601553S, CWJ602901S, CWJ603112S, CWJ603133S, CWJ603935S, CWJ603961S, CWJ601396S, CWJ601790S, CWJ602327S, CWJ602375S, CWJ602418S, CWJ602471S, CWJ602573S, CWJ602728S, CWJ602857S, CWJ603759S, CWJ603776S, CWJ604053S, CWJ604175S, CWJ604292S, CWJ604611S, CWJ601151S, CWJ602050S, CWJ601447S, CWJ603498S, CWJ603499S, CWJ603817S, CWJ603850S, CWJ604103S, CWJ604289S, CWJ604306S, CWJ604531S, CWJ604616S, CWJ601130S, CWJ601304S, CWJ602226S, CWJ602396S, CWJ603218S, CWJ603932S, CWJ601196S, CWJ601331S, CWJ601615S, CWJ602046S, CWJ602548S, CWJ602641S, CWJ602705S, CWJ602805S, CWJ602806S, CWJ602815S, CWJ603567S, CWJ604152S, CWJ604153S, CWJ604379S, CWJ602400S, CWJ602401S, CWJ601022S, CWJ601144S, CWJ601146S, CWJ601176S, CWJ601226S, CWJ601742S, CWJ601857S, CWJ601862S, CWJ602078S, CWJ602080S, CWJ602498S, CWJ602751S, CWJ602807S, CWJ602809S, CWJ602914S, CWJ603108S, CWJ603493S, CWJ604099S, CWJ604120S, CWJ600962S, CWJ601278S, CWJ601500S, CWJ601676S, CWJ602205S, CWJ602253S, CWJ602948S, CWJ604010S, CWJ604507S, CWJ602675S, CWJ603042S, CWJ601376S, CWJ602966S, CWJ602473S, CWJ603609S, CWJ601105S, CWJ601106S, CWJ601134S, CWJ601259S, CWJ601297S, CWJ601323S, CWJ601336S, CWJ601371S, CWJ601397S, CWJ601401S, CWJ601418S, CWJ601448S, CWJ601528S, CWJ601529S, CWJ601530S, CWJ601605S, CWJ601606S, CWJ601613S, CWJ601625S, CWJ601703S, CWJ601775S, CWJ601941S, CWJ601964S, CWJ601982S, CWJ601984S, CWJ601985S, CWJ601987S, CWJ602032S, CWJ602109S, CWJ602351S, CWJ602372S, CWJ602419S, CWJ602451S, CWJ602517S, CWJ602545S, CWJ602834S, CWJ602836S, CWJ602897S, CWJ602905S, CWJ602906S, CWJ603197S, CWJ603217S, CWJ603461S, CWJ603470S, CWJ603991S, CWJ603992S, CWJ604013S, CWJ604028S, CWJ604340S, CWJ604423S, CWJ604439S, CWJ604440S, CWJ604457S, CWJ604482S, CWJ604489S, CWJ604579S, CWJ601135S, CWJ601326S, CWJ601398S, CWJ601521S, CWJ601586S, CWJ601734S, CWJ602495S, CWJ602620S, CWJ602918S, CWJ603590S, CWJ601234S, CWJ601317S, CWJ601943S, CWJ603226S, CWJ601159S, CWJ601895S, CWJ603980S, CWJ604342S, CWJ602560S, CWJ603093S, CWJ603345S, CWJ603433S, CWJ603451S, CWJ604108S, CWJ601047S, CWJ601288S, CWJ602100S, CWJ602102S, CWJ602105S, CWJ602585S, CWJ603748S, CWJ604656S, CWJ602540S, CWJ602542S, CWJ603231S, CWJ603475S, CWJ603835S, CWJ601442S, CWJ602062S, CWJ602499S, CWJ603181S, CWJ602894S, CWJ603213S, CWJ601207S, CWJ601502S, CWJ602134S, CWJ602340S, CWJ604229S, CWJ603891S, CWJ604098S, CWJ601440S, CWJ601254S, CWJ601318S, CWJ601551S, CWJ603155S, CWJ603041S, CWJ604100S, CWJ601046S, CWJ601614S, CWJ601684S, CWJ601918S, CWJ602194S, CWJ600984S, CWJ602393S, CWJ601010S, CWJ601469S, CWJ604280S, CWJ601808S, CWJ602556S, CWJ603815S, CWJ604155S, CWJ601564S, CWJ600884S, CWJ603775S, CWJ603327S, CWJ602426S, CWJ600920S, CWJ600921S, CWJ600922S, CWJ600930S, CWJ601620S, CWJ601691S, CWJ601711S, CWJ601749S, CWJ601755S, CWJ601758S, CWJ601867S, CWJ601878S, CWJ601993S, CWJ602025S, CWJ602086S, CWJ602192S, CWJ602193S, CWJ602234S, CWJ602420S, CWJ602422S, CWJ602577S, CWJ602721S, CWJ602999S, CWJ603088S, CWJ603304S, CWJ603380S, CWJ604688S, CWJ601375S, CWJ604011S, CWJ601886S, CWJ601977S, CWJ602656S, CWJ602743S, CWJ602747S, CWJ603580S, CWJ604348S, CWJ603744S, CWJ604284S, CWJ601352S, CWJ601454S, CWJ601531S, CWJ601538S, CWJ601730S, CWJ602066S, CWJ603046S, CWJ603296S, CWJ604642S, CWJ602589S, CWJ602746S, CWJ600987S, CWJ601763S, CWJ601471S, CWJ602214S, CWJ602385S, CWJ603321S, CWJ603323S, CWJ603837S, CWJ603845S, CWJ604217S, CWJ604293S, CWJ604418S, CWJ601825S, CWJ602811S, CWJ601913S, CWJ603104S, CWJ601239S, CWJ601243S, CWJ601330S, CWJ601463S, CWJ601466S, CWJ601255S, CWJ601535S, CWJ601955S, CWJ602464S, CWJ603492S, CWJ604171S, CWJ602107S, CWJ601112S, CWJ601626S, CWJ602691S, CWJ602716S, CWJ602717S, CWJ603072S, CWJ604529S, CWJ603071S, CWJ600944S, CWJ601356S, CWJ601876S, CWJ601930S, CWJ601539S, CWJ602406S, CWJ602298S, CWJ603941S, CWJ604300S, CWJ603119S, CWJ601289S, CWJ604116S, CWJ601622S, CWJ601429S, CWJ601576S, CWJ604621S, CWJ601550S, CWJ601179S, CWJ601664S, CWJ602272S, CWJ600956S, CWJ603960S, CWJ601546S, CWJ603883S, CWJ604198S, CWJ604617S, CWJ601121S, CWJ600999S, CWJ603428S, CWJ604227S, CWJ603318S, CWJ601082S, CWJ600865S, CWJ601803S, CWJ601113S, CWJ601562S, CWJ602038S, CWJ602247S, CWJ604197S, CWJ601053S, CWJ602048S, CWJ600870S, CWJ601165S, CWJ601184S, CWJ601542S, CWJ601806S, CWJ600886S, CWJ600888S, CWJ601191S, CWJ601885S, CWJ603673S, CWJ603707S, CWJ603720S, CWJ601386S, CWJ601387S, CWJ601409S, CWJ601639S, CWJ601838S, CWJ602049S, CWJ602523S, CWJ602524S, CWJ602594S, CWJ603320S, CWJ603679S, CWJ603681S, CWJ603830S, CWJ603971S, CWJ604090S, CWJ604205S, CWJ604207S, CWJ604209S, CWJ604285S, CWJ604345S, CWJ604346S, CWJ604354S, CWJ604460S, CWJ604462S, CWJ604470S, CWJ604499S, CWJ604500S, CWJ604503S, CWJ601901S, CWJ602314S, CWJ600973S, CWJ602779S, CWJ602780S, CWJ602925S, CWJ603178S, CWJ603338S, CWJ603766S, CWJ602883S, CWJ602989S, CWJ604282S, CWJ601348S, CWJ601351S, CWJ602789S, CWJ602790S, CWJ603807S, CWJ601267S, CWJ601493S, CWJ602154S, CWJ602878S, CWJ602890S, CWJ603084S, CWJ603381S, CWJ601161S, CWJ601162S, CWJ601902S, CWJ600961S, CWJ600980S, CWJ601800S, CWJ601801S, CWJ602210S, CWJ602286S, CWJ602294S, CWJ602618S, CWJ602619S, CWJ603341S, CWJ604050S, CWJ603538S, CWJ601750S, CWJ602765S, CWJ601972S, CWJ601743S, CWJ602336S, CWJ602879S, CWJ604048S, CWJ604104S, CWJ602303S, CWJ602628S, CWJ603130S, CWJ603162S, CWJ603996S, CWJ604009S, CWJ603825S, CWJ604055S, CWJ600989S, CWJ601380S, CWJ601712S, CWJ601715S, CWJ601725S, CWJ601726S, CWJ601727S, CWJ601729S, CWJ601732S, CWJ601737S, CWJ601740S, CWJ601741S, CWJ602638S, CWJ602642S, CWJ602659S, CWJ602660S, CWJ603149S, CWJ603161S, CWJ603182S, CWJ603219S, CWJ603335S, CWJ603455S, CWJ603468S, CWJ603662S, CWJ603705S, CWJ603750S, CWJ603763S, CWJ603877S, CWJ604150S, CWJ604587S, CWJ604650S, CWJ604693S, CWJ604715S, CWJ604736S, CWJ601698S, CWJ601864S, CWJ602014S, CWJ603963S, CWJ604425S, CWJ603171S, CWJ602797S, CWJ602798S, CWJ603003S, CWJ603544S, CWJ603545S, CWJ603553S, CWJ603603S, CWJ603627S, CWJ603631S, CWJ603981S, CWJ604608S, CWJ601150S, CWJ601152S, CWJ601645S, CWJ601837S, CWJ602077S, CWJ602262S, CWJ602412S, CWJ602479S, CWJ602555S, CWJ602600S, CWJ601434S, CWJ602954S, CWJ603857S, CWJ601633S, CWJ601840S, CWJ601931S, CWJ602421S, CWJ602840S, CWJ602842S, CWJ601195S, CWJ601404S, CWJ604764S, CWJ600887S, CWJ601157S, CWJ601218S, CWJ601293S, CWJ601709S, CWJ602221S, CWJ602553S, CWJ603055S, CWJ603606S, CWJ603889S, CWJ603942S, CWJ603974S, CWJ604102S, CWJ604484S, CWJ604543S, CWJ603063S, CWJ601279S, CWJ601495S, CWJ602440S, CWJ602803S, CWJ602933S, CWJ603458S, CWJ603459S, CWJ603838S, CWJ603967S, CWJ603968S, CWJ604002S, CWJ604003S, CWJ604158S, CWJ604208S, CWJ604603S, CWJ604609S, CWJ604726S, CWJ604754S, CWJ601316S, CWJ601324S, CWJ601719S, CWJ602736S, CWJ604088S, CWJ604188S, CWJ604299S, CWJ603300S, CWJ604021S, CWJ604459S, CWJ601198S, CWJ600715S, CWJ601437S, CWJ600872S, CWJ600903S, CWJ601059S, CWJ601060S, CWJ601262S, CWJ601263S, CWJ601393S, CWJ601407S, CWJ601674S, CWJ601708S, CWJ601760S, CWJ601992S, CWJ602229S, CWJ602380S, CWJ602404S, CWJ602844S, CWJ602967S, CWJ603176S, CWJ603344S, CWJ603944S, CWJ604163S, CWJ601893S, CWJ602024S, CWJ602280S, CWJ602617S, CWJ603132S, CWJ601718S, CWJ603478S, CWJ603481S, CWJ604472S, CWJ600953S, CWJ600990S, CWJ601075S, CWJ601076S, CWJ601079S, CWJ601114S, CWJ601115S, CWJ601116S, CWJ601122S, CWJ601153S, CWJ601154S, CWJ601167S, CWJ601168S, CWJ601212S, CWJ601214S, CWJ601275S, CWJ601276S, CWJ601305S, CWJ601307S, CWJ601364S, CWJ601368S, CWJ601464S, CWJ601465S, CWJ601467S, CWJ601468S, CWJ601475S, CWJ601510S, CWJ601511S, CWJ601526S, CWJ601556S, CWJ601558S, CWJ601559S, CWJ601560S, CWJ601588S, CWJ601593S, CWJ601596S, CWJ601599S, CWJ601600S, CWJ601658S, CWJ601672S, CWJ601701S, CWJ601702S, CWJ601705S, CWJ601783S, CWJ601785S, CWJ601786S, CWJ601787S, CWJ601854S, CWJ601856S, CWJ601858S, CWJ601859S, CWJ601874S, CWJ601892S, CWJ601966S, CWJ601967S, CWJ601971S, CWJ602156S, CWJ602373S, CWJ602374S, CWJ602376S, CWJ602377S, CWJ602383S, CWJ602415S, CWJ602428S, CWJ602508S, CWJ602513S, CWJ602565S, CWJ602616S, CWJ602732S, CWJ602733S, CWJ602734S, CWJ602975S, CWJ602977S, CWJ602978S, CWJ603099S, CWJ603289S, CWJ603507S, CWJ603548S, CWJ603550S, CWJ603635S, CWJ603675S, CWJ603782S, CWJ603854S, CWJ603856S, CWJ603945S, CWJ604093S, CWJ604096S, CWJ604221S, CWJ604644S, CWJ604658S, CWJ600954S, CWJ601173S, CWJ601264S, CWJ601284S, CWJ601472S, CWJ601474S, CWJ601673S, CWJ601807S, CWJ601818S, CWJ601861S, CWJ601968S, CWJ601969S, CWJ601970S, CWJ602215S, CWJ602281S, CWJ602284S, CWJ602416S, CWJ602510S, CWJ602576S, CWJ602839S, CWJ602841S, CWJ602843S, CWJ602888S, CWJ602898S, CWJ602924S, CWJ603102S, CWJ603205S, CWJ603206S, CWJ603262S, CWJ603392S, CWJ603394S, CWJ603605S, CWJ603636S, CWJ603783S, CWJ603827S, CWJ604313S, CWJ604416S, CWJ604417S, CWJ604454S, CWJ601109S, CWJ601110S, CWJ601185S, CWJ601193S, CWJ601194S, CWJ601552S, CWJ601819S, CWJ601820S, CWJ601915S, CWJ602065S, CWJ602148S, CWJ602149S, CWJ602188S, CWJ602189S, CWJ602263S, CWJ602321S, CWJ602322S, CWJ602431S, CWJ602487S, CWJ602488S, CWJ602489S, CWJ602491S, CWJ603048S, CWJ603340S, CWJ603668S, CWJ603751S, CWJ604199S, CWJ604200S, CWJ604429S, CWJ604601S, CWJ604685S, CWJ601169S, CWJ601170S, CWJ601211S, CWJ601213S, CWJ601265S, CWJ601280S, CWJ602159S, CWJ603100S, CWJ603506S, CWJ600883S, CWJ600902S, CWJ601012S, CWJ601056S, CWJ601174S, CWJ601294S, CWJ601334S, CWJ601423S, CWJ601425S, CWJ601462S, CWJ601710S, CWJ601747S, CWJ601952S, CWJ602432S, CWJ602604S, CWJ602960S, CWJ602973S, CWJ603167S, CWJ603365S, CWJ603585S, CWJ603622S, CWJ603643S, CWJ603649S, CWJ603788S, CWJ603791S, CWJ603843S, CWJ603851S, CWJ603969S, CWJ604024S, CWJ604203S, CWJ604223S, CWJ604226S, CWJ604308S, CWJ604404S, CWJ604466S, CWJ604467S, CWJ603101S, CWJ601171S, CWJ601210S, CWJ601365S, CWJ601557S, CWJ601685S, CWJ601860S, CWJ601965S, CWJ602509S, CWJ602511S, CWJ603303S, CWJ603382S, CWJ603536S, CWJ603540S, CWJ603546S, CWJ603780S, CWJ603853S, CWJ604315S, CWJ602087S, CWJ602352S, CWJ602816S, CWJ602817S, CWJ604765S, CWJ603558S, CWJ603997S, CWJ601728S, CWJ603820S, CWJ602668S, CWJ601665S, CWJ602622S, CWJ602648S, CWJ602913S, CWJ602930S, CWJ602981S, CWJ604063S, CWJ604422S, CWJ601908S, CWJ603787S, CWJ602465S, CWJ602468S, CWJ601260S, CWJ601769S, CWJ601722S, CWJ601889S, CWJ602211S, CWJ602212S, CWJ604461S, CWJ604607S, CWJ601016S, CWJ601677S, CWJ602191S, CWJ603554S, CWJ601720S, CWJ602240S, CWJ602241S, CWJ602242S, CWJ602854S, CWJ603288S, CWJ600893S, CWJ601594S, CWJ601597S, CWJ601770S, CWJ602654S, CWJ602799S, CWJ602979S, CWJ603126S, CWJ603229S, CWJ601048S, CWJ601078S, CWJ601123S, CWJ601124S, CWJ601127S, CWJ601991S, CWJ600876S, CWJ600955S, CWJ601008S, CWJ601496S, CWJ601623S, CWJ601624S, CWJ601643S, CWJ601829S, CWJ602045S, CWJ602661S, CWJ602755S, CWJ602915S, CWJ603456S, CWJ603469S, CWJ603595S, CWJ603789S, CWJ601216S, CWJ601277S, CWJ601670S, CWJ602378S, CWJ602382S, CWJ602455S, CWJ602480S, CWJ603583S, CWJ604533S, CWJ600913S, CWJ601074S, CWJ601111S, CWJ601232S, CWJ601245S, CWJ601834S, CWJ602937S, CWJ602938S, CWJ603029S, CWJ601137S, CWJ601235S, CWJ602096S, CWJ602887S, CWJ603188S, CWJ603496S, CWJ603500S, CWJ603527S, CWJ603626S, CWJ603692S, CWJ603904S, CWJ604624S, CWJ601890S, CWJ602092S, CWJ602512S, CWJ602514S, CWJ601451S, CWJ600952S, CWJ601946S, CWJ602519S, CWJ602535S, CWJ602579S, CWJ603002S, CWJ603080S, CWJ603667S, CWJ603735S, CWJ603833S, CWJ604012S, CWJ602551S, CWJ601021S, CWJ601128S, CWJ601189S, CWJ601247S, CWJ601249S, CWJ601250S, CWJ601252S, CWJ601453S, CWJ601525S, CWJ601568S, CWJ601592S, CWJ601644S, CWJ601805S, CWJ601821S, CWJ601827S, CWJ601932S, CWJ601944S, CWJ602254S, CWJ602708S, CWJ602742S, CWJ603017S, CWJ603024S, CWJ603069S, CWJ603147S, CWJ603211S, CWJ603348S, CWJ603430S, CWJ603754S, CWJ603772S, CWJ604044S, CWJ601251S, CWJ601377S, CWJ601912S, CWJ601956S, CWJ601960S, CWJ602013S, CWJ602230S, CWJ602482S, CWJ602712S, CWJ602735S, CWJ602988S, CWJ602998S, CWJ603032S, CWJ603232S, CWJ603274S, CWJ603330S, CWJ603576S, CWJ603604S, CWJ603678S, CWJ603690S, CWJ603790S, CWJ604414S, CWJ601700S, CWJ603823S, CWJ602111S, CWJ604279S, CWJ604451S, CWJ604453S, CWJ601866S, CWJ601107S, CWJ601339S, CWJ601850S, CWJ601853S, CWJ601855S, CWJ602053S, CWJ602082S, CWJ602119S, CWJ602203S, CWJ602207S, CWJ602506S, CWJ602570S, CWJ602670S, CWJ602730S, CWJ602826S, CWJ602889S, CWJ602909S, CWJ602929S, CWJ602934S, CWJ602974S, CWJ603016S, CWJ603034S, CWJ603036S, CWJ603105S, CWJ603183S, CWJ603184S, CWJ603221S, CWJ603307S, CWJ603312S, CWJ603425S, CWJ603431S, CWJ603476S, CWJ603477S, CWJ603716S, CWJ603745S, CWJ603865S, CWJ604041S, CWJ604464S, CWJ604581S, CWJ604697S, CWJ604701S, CWJ600997S, CWJ601178S, CWJ601236S, CWJ602582S, CWJ602639S, CWJ603026S, CWJ603027S, CWJ603189S, CWJ603208S, CWJ603324S, CWJ601723S, CWJ600923S, CWJ600926S, CWJ600927S, CWJ600969S, CWJ601068S, CWJ601069S, CWJ601219S, CWJ601455S, CWJ601456S, CWJ601488S, CWJ601489S, CWJ601681S, CWJ601721S, CWJ602008S, CWJ602009S, CWJ602213S, CWJ602441S, CWJ602602S, CWJ602603S, CWJ602788S, CWJ602891S, CWJ602926S, CWJ602995S, CWJ603237S, CWJ603322S, CWJ603343S, CWJ603503S, CWJ603556S, CWJ603557S, CWJ604195S, CWJ604231S, CWJ604283S, CWJ604615S, CWJ604631S, CWJ603918S, CWJ603228S, CWJ601089S, CWJ601422S, CWJ602121S, CWJ602665S, CWJ602855S, CWJ603207S, CWJ603210S, CWJ601088S, CWJ601499S, CWJ601504S, CWJ601627S, CWJ601636S, CWJ601841S, CWJ601928S, CWJ602657S, CWJ602892S, CWJ603419S, CWJ604149S, CWJ601372S, CWJ601373S, CWJ601444S, CWJ601518S, CWJ601591S, CWJ601608S, CWJ602123S, CWJ602387S, CWJ602662S, CWJ603860S, CWJ603861S, CWJ600906S, CWJ601118S, CWJ601246S, CWJ601257S, CWJ601497S, CWJ601766S, CWJ601767S, CWJ601835S, CWJ601917S, CWJ601973S, CWJ602018S, CWJ602029S, CWJ602227S, CWJ602532S, CWJ602613S, CWJ602824S, CWJ602859S, CWJ602871S, CWJ602874S, CWJ603018S, CWJ603019S, CWJ603175S, CWJ603287S, CWJ603315S, CWJ603316S, CWJ603422S, CWJ603721S, CWJ603925S, CWJ603947S, CWJ603950S, CWJ604706S, CWJ601233S, CWJ601329S, CWJ601577S, CWJ601781S, CWJ602635S, CWJ603233S, CWJ603234S, CWJ603528S, CWJ603691S, CWJ603738S, CWJ603989S, CWJ602417S, CWJ603886S, CWJ604027S, CWJ602091S, CWJ601241S, CWJ602520S, CWJ603885S, CWJ601523S, CWJ601541S, CWJ601836S, CWJ602131S, CWJ603025S, CWJ603413S, CWJ603439S, CWJ603450S, CWJ603482S, CWJ601490S, CWJ601692S, CWJ603360S, CWJ603897S, CWJ603771S, CWJ604409S, CWJ604485S, CWJ602947S, CWJ603452S, CWJ603484S, CWJ603739S, CWJ603740S, CWJ600985S, CWJ602436S, CWJ600871S, CWJ601090S, CWJ601366S, CWJ601389S, CWJ601459S, CWJ601637S, CWJ601979S, CWJ602072S, CWJ602116S, CWJ602183S, CWJ602646S, CWJ602709S, CWJ602850S, CWJ603028S, CWJ603030S, CWJ603251S, CWJ604592S, CWJ601452S, CWJ601361S, CWJ601503S, CWJ601844S, CWJ601962S, CWJ602034S, CWJ602388S, CWJ602496S, CWJ602529S, CWJ602710S, CWJ602731S, CWJ603087S, CWJ600866S, CWJ601953S, CWJ602663S, CWJ601092S, CWJ602703S, CWJ603011S, CWJ601283S, CWJ601543S, CWJ601598S, CWJ601699S, CWJ601925S, CWJ601949S, CWJ602179S, CWJ602186S, CWJ602521S, CWJ601739S, CWJ602726S, CWJ602993S, CWJ604588S, CWJ600986S, CWJ601081S, CWJ600964S, CWJ601085S, CWJ603136S, CWJ604037S, CWJ604302S, CWJ604380S, CWJ604583S, CWJ602996S, CWJ601487S, CWJ601839S, CWJ601976S, CWJ602166S, CWJ602182S, CWJ602649S, CWJ602652S, CWJ602658S, CWJ602833S, CWJ603466S, CWJ601093S, CWJ601479S, CWJ602621S, CWJ603286S, CWJ603305S, CWJ603584S, CWJ603647S, CWJ603756S, CWJ603764S, CWJ603824S, CWJ603917S, CWJ604402S, CWJ604449S, CWJ604450S, CWJ601313S, CWJ601707S, CWJ601978S, CWJ602059S, CWJ602147S, CWJ602152S, CWJ602354S, CWJ602356S, CWJ602500S, CWJ602578S, CWJ602610S, CWJ602623S, CWJ602651S, CWJ602697S, CWJ602698S, CWJ602748S, CWJ602750S, CWJ602804S, CWJ602812S, CWJ602823S, CWJ603009S, CWJ603013S, CWJ603169S, CWJ603238S, CWJ603266S, CWJ603326S, CWJ603342S, CWJ603374S, CWJ603464S, CWJ603508S, CWJ603714S, CWJ603743S, CWJ603765S, CWJ603822S, CWJ603895S, CWJ603900S, CWJ603949S, CWJ603970S, CWJ604034S, CWJ604065S, CWJ604424S, CWJ604455S, CWJ604456S, CWJ604497S, CWJ600714S, CWJ601974S, CWJ604095S, CWJ603295S, CWJ601402S, CWJ604035S, CWJ601744S, CWJ601687S, CWJ604390S, CWJ604600S, CWJ604655S, CWJ600948S, CWJ601602S, CWJ602588S, CWJ602590S, CWJ603098S, CWJ601395S, CWJ603929S, CWJ604016S, CWJ604744S, CWJ604748S, CWJ602671S, CWJ602672S, CWJ602674S, CWJ602679S, CWJ602684S, CWJ602688S, CWJ602689S, CWJ602690S, CWJ603927S, CWJ603928S, CWJ603931S, CWJ603972S, CWJ604004S, CWJ604023S, CWJ604398S, CWJ604405S, CWJ604431S, CWJ604443S, CWJ604448S, CWJ604452S, CWJ604498S, CWJ602667S, CWJ604445S, CWJ602209S, CWJ601746S, CWJ602020S, CWJ602113S, CWJ603753S, CWJ601007S, CWJ601566S, CWJ602075S, CWJ602085S, CWJ602309S, CWJ602884S, CWJ602296S, CWJ602270S, CWJ600894S, CWJ601017S, CWJ601509S, CWJ603005S, CWJ601199S, CWJ601337S, CWJ601360S, CWJ601417S, CWJ601438S, CWJ602474S, CWJ602653S, CWJ602819S, CWJ603006S, CWJ603008S, CWJ603015S, CWJ603075S, CWJ603113S, CWJ603120S, CWJ603356S, CWJ603485S, CWJ603535S, CWJ603666S, CWJ603852S, CWJ601826S, CWJ604392S, CWJ601519S, CWJ601816S, CWJ601778S, CWJ601907S, CWJ603180S, CWJ603608S, CWJ603762S, CWJ604170S, CWJ604347S, CWJ604427S, CWJ603319S, CWJ601632S, CWJ601505S, CWJ601517S, CWJ601810S, CWJ602609S, CWJ603276S, CWJ601014S, CWJ602285S, CWJ603246S, CWJ603306S, CWJ604311S, CWJ604305S, CWJ603962S, CWJ604168S, CWJ604362S, CWJ602408S, CWJ602632S, CWJ602770S, CWJ602923S, CWJ603097S, CWJ604501S, CWJ600890S, CWJ602584S, CWJ603602S, CWJ603955S, CWJ604154S, CWJ601640S, CWJ603339S, CWJ604298S, CWJ604363S, CWJ604488S, CWJ604508S, CWJ604204S, CWJ604294S, CWJ604721S, CWJ602265S, CWJ602575S, CWJ601084S, CWJ603302S, CWJ601333S, CWJ601804S, CWJ603215S, CWJ604015S, CWJ601177S, CWJ602922S, CWJ602858S, CWJ600975S, CWJ604536S, CWJ604690S, CWJ603487S, CWJ604018S, CWJ604740S, CWJ603411S, CWJ604751S, CWJ601072S, CWJ601516S, CWJ602320S, CWJ603561S, CWJ600939S, CWJ601020S, CWJ601656S, CWJ601823S, CWJ602142S, CWJ603515S, CWJ603518S, CWJ603696S, CWJ601286S, CWJ601381S, CWJ602073S, CWJ602676S, CWJ603560S, CWJ600929S, CWJ601025S, CWJ601027S, CWJ601028S, CWJ601163S, CWJ601227S, CWJ601327S, CWJ601484S, CWJ601486S, CWJ601512S, CWJ601963S, CWJ602305S, CWJ603358S, CWJ603359S, CWJ603524S, CWJ603526S, CWJ603719S, CWJ601229S, CWJ602016S, CWJ602368S, CWJ603332S, CWJ600925S, CWJ603247S, CWJ601527S, CWJ602047S, CWJ601391S, CWJ601513S, CWJ601595S, CWJ601948S, CWJ602044S, CWJ602118S, CWJ602331S, CWJ602346S, CWJ602445S, CWJ602446S, CWJ602456S, CWJ602460S, CWJ602546S, CWJ602678S, CWJ602685S, CWJ603522S, CWJ603570S, CWJ603632S, CWJ603715S, CWJ602304S, CWJ602306S, CWJ602543S, CWJ600910S, CWJ601070S, CWJ600713S, CWJ600905S, CWJ600909S, CWJ601043S, CWJ601097S, CWJ601172S, CWJ601204S, CWJ601225S, CWJ601228S, CWJ601310S, CWJ601378S, CWJ601388S, CWJ601481S, CWJ601483S, CWJ601574S, CWJ601590S, CWJ601695S, CWJ601764S, CWJ601824S, CWJ601935S, CWJ602329S, CWJ602429S, CWJ602598S, CWJ602680S, CWJ602758S, CWJ602876S, CWJ603114S, CWJ603125S, CWJ603203S, CWJ603353S, CWJ603403S, CWJ601382S, CWJ603525S, CWJ601036S, CWJ601042S, CWJ601223S, CWJ601345S, CWJ601384S, CWJ601390S, CWJ601478S, CWJ601934S, CWJ602063S, CWJ602283S, CWJ603124S, CWJ603196S, CWJ603402S, CWJ603876S, CWJ601205S, CWJ600924S, CWJ601031S, CWJ601095S, CWJ601096S, CWJ601325S, CWJ601344S, CWJ601589S, CWJ602275S, CWJ602904S, CWJ603170S, CWJ600931S, CWJ601843S, CWJ602070S, CWJ603598S, CWJ601575S, CWJ602861S, CWJ603398S, CWJ603698S, CWJ601026S, CWJ601203S, CWJ601482S, CWJ601654S, CWJ601983S, CWJ602139S, CWJ602362S, CWJ602596S, CWJ602683S, CWJ602753S, CWJ602754S, CWJ602825S, CWJ603250S, CWJ603389S, CWJ603559S, CWJ600907S, CWJ603172S, CWJ603261S, CWJ600915S, CWJ601473S, CWJ601480S, CWJ601573S, CWJ602599S, CWJ603200S, CWJ603252S, CWJ603400S, CWJ603401S, CWJ600904S, CWJ603152S, CWJ602558S, CWJ602677S, CWJ602749S, CWJ602866S, CWJ603115S, CWJ603388S, CWJ601287S, CWJ601782S, CWJ602367S, CWJ602467S, CWJ602873S, CWJ603204S, CWJ603248S, CWJ603596S, CWJ603761S, CWJ601311S, CWJ601341S, CWJ603597S, CWJ600916S, CWJ602783S, CWJ600911S, CWJ600933S, CWJ601009S, CWJ601164S, CWJ601224S, CWJ601485S, CWJ601951S, CWJ602136S, CWJ602278S, CWJ602444S, CWJ602449S, CWJ602644S, CWJ602682S, CWJ602772S, CWJ602781S, CWJ602782S, CWJ602864S, CWJ603283S, CWJ603337S, CWJ603357S, CWJ603409S, CWJ603410S, CWJ603630S, CWJ603734S, CWJ602527S, CWJ600908S, CWJ602279S, CWJ602557S, CWJ603186S, CWJ603577S, CWJ601100S, CWJ601101S, CWJ601102S, CWJ601221S, CWJ602274S, CWJ602597S, CWJ603334S, CWJ601990S, CWJ603396S, CWJ603397S, CWJ603404S, CWJ603593S, CWJ603629S, CWJ604700S, CWJ601385S, CWJ602019S, CWJ602534S, CWJ602595S, CWJ603202S, CWJ603362S, CWJ603395S, CWJ603568S, CWJ603578S, CWJ600941S, CWJ601019S, CWJ601040S, CWJ601342S, CWJ601922S, CWJ603516S, CWJ603563S, CWJ603564S, CWJ601073S, CWJ602371S, CWJ602528S, CWJ603742S, CWJ603581S, CWJ601506S, CWJ602687S, CWJ603201S, CWJ601023S, CWJ601272S, CWJ601514S, CWJ601515S, CWJ602054S, CWJ602133S, CWJ602911S, CWJ603718S, CWJ604698S, CWJ604709S, CWJ602466S, CWJ603333S, CWJ601032S, CWJ601655S, CWJ601924S, CWJ602222S, CWJ602282S, CWJ602361S, CWJ602365S, CWJ602430S, CWJ603284S, CWJ603355S, CWJ603566S, CWJ603648S, CWJ603702S, CWJ603818S, CWJ602869S, CWJ601041S, CWJ602056S, CWJ602458S, CWJ603163S, CWJ603361S, CWJ601322S, CWJ603613S, CWJ604704S, CWJ601103S, CWJ602533S, CWJ602601S, CWJ602681S, CWJ604695S, CWJ601030S, CWJ602525S, CWJ602903S, CWJ603110S, CWJ602313S, CWJ604033S, CWJ602802S, CWJ603711S, CWJ600867S, CWJ604739S, CWJ604743S, CWJ604746S, CWJ601522S, CWJ600947S, CWJ601320S, CWJ601332S, CWJ601443S, CWJ601445S, CWJ601555S, CWJ601898S, CWJ601899S, CWJ601995S, CWJ602089S, CWJ602256S, CWJ602319S, CWJ602357S, CWJ602359S, CWJ602403S, CWJ602407S, CWJ602612S, CWJ602614S, CWJ602944S, CWJ602945S, CWJ603311S, CWJ604544S, CWJ601569S, CWJ601570S, CWJ601735S, CWJ601736S, CWJ601738S, CWJ602768S, CWJ602538S, CWJ603280S, CWJ603427S, CWJ604169S, CWJ604014S, CWJ601413S, CWJ601619S, CWJ601920S, CWJ601923S, CWJ600917S, CWJ601682S, CWJ602012S, CWJ602064S, CWJ602223S, CWJ602752S, CWJ602867S, CWJ603156S, CWJ603265S, CWJ603504S, CWJ603565S, CWJ603600S, CWJ602536S, CWJ600912S, CWJ602829S, CWJ602830S, CWJ601869S, CWJ603701S, CWJ601458S, CWJ601428S, CWJ601534S, CWJ601716S, CWJ603417S, CWJ603418S, CWJ603442S, CWJ603460S, CWJ601256S, CWJ601394S, CWJ602015S, CWJ601939S, CWJ601940S, CWJ601942S, CWJ601947S, CWJ601957S, CWJ602235S, CWJ602236S, CWJ602239S, CWJ602257S, CWJ602258S, CWJ602437S, CWJ602438S, CWJ602439S, CWJ602447S, CWJ602448S, CWJ602774S, CWJ602775S, CWJ602776S, CWJ602777S, CWJ602778S, CWJ602917S, CWJ602919S, CWJ603190S, CWJ603191S, CWJ603192S, CWJ603193S, CWJ603194S, CWJ603512S, CWJ603513S, CWJ603514S, CWJ603809S, CWJ603810S, CWJ603811S, CWJ603812S, CWJ603816S, CWJ603870S, CWJ603871S, CWJ603872S, CWJ604056S, CWJ604058S, CWJ604125S, CWJ604430S, CWJ604434S, CWJ602010S, CWJ604750S, CWJ604741S, CWJ601274S, CWJ601298S, CWJ601584S, CWJ601604S, CWJ601242S, CWJ601306S, CWJ601357S, CWJ601383S, CWJ601776S, CWJ601842S, CWJ602000S, CWJ602155S, CWJ602233S, CWJ602297S, CWJ602759S, CWJ602760S, CWJ602761S, CWJ602983S, CWJ603179S, CWJ603988S, CWJ604569S, CWJ600868S, CWJ603146S, CWJ603683S, CWJ603796S, CWJ604224S, CWJ604468S, CWJ604546S, CWJ604568S, CWJ600914S, CWJ601683S, CWJ601731S, CWJ600877S, CWJ600879S, CWJ600932S, CWJ601024S, CWJ601033S, CWJ601049S, CWJ601050S, CWJ601051S, CWJ601063S, CWJ601108S, CWJ601156S, CWJ601186S, CWJ601269S, CWJ601419S, CWJ601579S, CWJ601581S, CWJ601768S, CWJ601812S, CWJ601813S, CWJ601814S, CWJ602108S, CWJ602110S, CWJ602353S, CWJ602581S, CWJ602714S, CWJ602715S, CWJ602718S, CWJ602949S, CWJ602951S, CWJ602985S, CWJ602986S, CWJ602813S, CWJ602814S, CWJ602818S, CWJ603671S, CWJ603672S, CWJ603801S, CWJ603888S, CWJ603934S, CWJ604017S, CWJ604020S, CWJ604614S, CWJ601055S, CWJ602539S, CWJ600959S, CWJ600968S, CWJ600970S, CWJ600976S, CWJ600979S, CWJ600983S, CWJ600995S, CWJ600996S, CWJ600998S, CWJ601271S, CWJ601415S, CWJ601561S, CWJ601572S, CWJ601693S, CWJ601697S, CWJ601794S, CWJ601795S, CWJ601796S, CWJ601797S, CWJ601958S, CWJ601961S, CWJ602058S, CWJ602060S, CWJ602164S, CWJ602165S, CWJ602167S, CWJ602217S, CWJ602219S, CWJ602287S, CWJ602288S, CWJ602291S, CWJ602292S, CWJ602293S, CWJ602791S, CWJ602792S, CWJ602795S, CWJ602957S, CWJ602958S, CWJ603065S, CWJ603066S, CWJ603067S, CWJ603541S, CWJ603547S, CWJ603549S, CWJ603552S, CWJ603555S, CWJ604110S, CWJ604113S, CWJ604115S, CWJ604190S, CWJ604191S, CWJ604192S, CWJ604206S, CWJ604473S, CWJ604495S, CWJ604496S, CWJ604566S, CWJ602537S, CWJ600880S, CWJ601125S, CWJ602587S, CWJ603000S, CWJ604747S, CWJ601029S, CWJ603111S, CWJ603569S, CWJ601308S, CWJ601309S, CWJ601321S, CWJ601989S, CWJ602037S, CWJ602141S, CWJ602332S, CWJ602544S, CWJ603517S, CWJ603562S, CWJ604699S, CWJ604483S, CWJ604738S, CWJ601832S, CWJ602392S, CWJ603954S, CWJ604185S, CWJ602939S, CWJ603572S, CWJ603574S, CWJ604606S, CWJ600935S, CWJ604005S, CWJ604653S, CWJ602875S, CWJ602970S, CWJ603986S, CWJ600963S, CWJ600967S, CWJ600993S, CWJ601091S, CWJ601426S, CWJ601435S, CWJ601585S, CWJ601662S, CWJ601761S, CWJ601780S, CWJ601784S, CWJ601950S, CWJ601980S, CWJ602040S, CWJ602198S, CWJ602264S, CWJ602349S, CWJ602350S, CWJ602530S, CWJ602907S, CWJ602908S, CWJ602955S, CWJ603010S, CWJ603020S, CWJ603031S, CWJ603453S, CWJ604723S, CWJ604725S, CWJ603987S, CWJ603255S, CWJ603256S, CWJ603258S, CWJ603259S, CWJ603260S, CWJ603267S, CWJ603272S, CWJ603277S, CWJ603278S, CWJ603281S, CWJ603615S, CWJ603616S, CWJ604006S, CWJ604026S, CWJ604061S, CWJ604064S, CWJ604066S, CWJ604119S, CWJ604121S, CWJ604166S, CWJ604233S, CWJ604400S, CWJ604401S, CWJ604419S, CWJ604428S, CWJ604640S, CWJ604641S, CWJ604652S, CWJ604654S, CWJ604760S, CWJ604761S, CWJ604762S, CWJ604763S, CWJ604766S, CWJ604767S, CWJ604768S, CWJ604234S, CWJ604255S, CWJ604476S, CWJ604477S, CWJ604478S, CWJ603488S, CWJ603489S, CWJ603490S, CWJ600899S, CWJ600900S, CWJ601362S, CWJ601815S, CWJ601696S, CWJ602366S, CWJ602450S, CWJ603123S, CWJ603747S, CWJ604703S, CWJ604182S, CWJ603571S, CWJ603976S, CWJ604114S, CWJ604742S, CWJ602902S, CWJ604689S, CWJ603364S, CWJ603391S, CWJ603699S, CWJ603729S, CWJ604032S, CWJ604356S, CWJ603703S, CWJ601421S, CWJ601888S, CWJ602952S, CWJ603682S, CWJ604682S, CWJ601367S, CWJ601617S, CWJ601631S, CWJ601817S, CWJ601887S, CWJ602414S, CWJ603480S, CWJ603983S, CWJ604183S, CWJ604019S, CWJ604180S, CWJ601882S, CWJ602301S, CWJ602269S, CWJ602360S, CWJ602569S, CWJ604038S, CWJ604745S, CWJ604164S, CWJ603940S, CWJ603948S, CWJ603964S, CWJ603965S, CWJ604474S, CWJ604733S, CWJ603985S, CWJ604165S, CWJ603903S, CWJ603127S, CWJ604109S, CWJ603990S, CWJ604386S, CWJ604591S; GTIN 00643169837829, Lot Serial Numbers: CWJ200825H, CWJ200998H, CWJ201373H, CWJ201418H, CWJ200632H, CWJ200665H, CWJ200764H, CWJ200913H, CWJ200944H, CWJ200956H, CWJ201075H, CWJ201211H, CWJ201241H, CWJ201034H, CWJ201341H, CWJ200984H, CWJ201093H, CWJ201114H, CWJ201136H, CWJ201141H, CWJ201162H, CWJ201246H, CWJ201257H, CWJ200712H, CWJ201285H, CWJ200476H, CWJ200539H, CWJ201045H, CWJ201256H, CWJ201484H, CWJ200449H, CWJ200451H, CWJ200502H, CWJ200519H, CWJ200526H, CWJ200527H, CWJ200563H, CWJ200582H, CWJ200606H, CWJ201270H, CWJ201314H, CWJ201322H, CWJ201332H, CWJ201368H, CWJ201376H, CWJ201399H, CWJ200726H, CWJ201458H, CWJ201542H, CWJ200458H, CWJ200644H, CWJ200720H, CWJ200734H, CWJ200793H, CWJ200813H, CWJ200824H, CWJ200886H, CWJ200894H, CWJ200979H, CWJ200981H, CWJ201209H, CWJ201330H, CWJ201346H, CWJ201367H, CWJ201516H, CWJ201517H, CWJ200630H, CWJ200661H, CWJ201022H, CWJ200421H, CWJ200676H, CWJ200746H, CWJ201040H, CWJ201049H, CWJ201187H, CWJ201292H, CWJ201344H, CWJ200512H, CWJ200547H, CWJ200681H, CWJ201033H, CWJ201074H, CWJ201237H, CWJ201377H, CWJ200513H, CWJ200947H, CWJ201305H, CWJ200500H, CWJ201120H, CWJ201394H, CWJ201401H, CWJ201420H, CWJ201537H, CWJ201538H, CWJ200903H, CWJ201170H, CWJ200679H, CWJ201252H, CWJ201472H, CWJ200445H, CWJ200472H, CWJ200496H, CWJ200562H, CWJ200573H, CWJ200594H, CWJ200601H, CWJ200633H, CWJ200637H, CWJ200674H, CWJ200822H, CWJ200883H, CWJ200934H, CWJ200960H, CWJ201069H, CWJ201173H, CWJ201217H, CWJ201219H, CWJ201405H, CWJ201451H, CWJ201454H, CWJ201495H, CWJ201567H, CWJ200469H, CWJ200487H, CWJ200592H, CWJ200604H, CWJ200994H, CWJ201060H, CWJ201083H, CWJ201087H, CWJ201167H, CWJ201203H, CWJ201234H, CWJ201312H, CWJ201440H, CWJ201462H, CWJ201483H, CWJ201535H, CWJ200560H, CWJ200423H, CWJ200492H, CWJ200530H, CWJ200534H, CWJ200687H, CWJ200702H, CWJ200743H, CWJ200762H, CWJ200871H, CWJ200999H, CWJ201107H, CWJ201117H, CWJ201118H, CWJ201179H, CWJ201254H, CWJ201349H, CWJ201410H, CWJ201519H, CWJ201520H, CWJ201541H, CWJ200533H, CWJ200564H, CWJ200602H, CWJ200778H, CWJ200803H, CWJ200807H, CWJ201030H, CWJ201130H, CWJ201392H, CWJ201501H, CWJ200411H, CWJ200520H, CWJ200902H, CWJ200911H, CWJ201143H, CWJ200430H, CWJ200628H, CWJ200663H, CWJ200777H, CWJ201099H, CWJ200639H, CWJ200857H, CWJ200881H, CWJ200919H, CWJ200950H, CWJ201002H, CWJ201061H, CWJ201089H, CWJ201200H, CWJ201202H, CWJ201313H, CWJ201327H, CWJ201478H, CWJ201533H, CWJ200797H, CWJ201031H, CWJ200927H, CWJ200436H, CWJ200447H, CWJ200503H, CWJ200612H, CWJ200969H, CWJ201043H, CWJ201085H, CWJ201160H, CWJ201164H, CWJ201165H, CWJ201417H, CWJ201435H, CWJ201450H, CWJ200611H, CWJ200872H, CWJ200988H, CWJ201029H, CWJ201135H, CWJ201156H, CWJ201066H, CWJ200335H, CWJ200667H, CWJ200722H, CWJ200761H, CWJ200794H, CWJ200819H, CWJ200848H, CWJ200867H, CWJ200930H, CWJ201044H, CWJ201133H, CWJ201199H, CWJ201208H, CWJ201308H, CWJ201342H, CWJ201425H, CWJ200477H, CWJ200846H, CWJ200971H, CWJ201059H, CWJ200660H, CWJ200912H, CWJ201331H, CWJ200996H, CWJ200538H, CWJ200908H, CWJ200962H, CWJ201032H, CWJ201123H, CWJ201356H, CWJ200716H, CWJ200861H, CWJ200689H, CWJ200978H, CWJ201469H, CWJ201474H, CWJ201065H, CWJ201539H, CWJ200584H, CWJ201115H, CWJ200455H, CWJ200495H, CWJ200775H, CWJ201013H, CWJ201101H, CWJ201132H, CWJ201269H, CWJ201407H, CWJ201508H, CWJ201548H, CWJ200931H, CWJ200771H, CWJ201152H, CWJ201212H, CWJ201249H, CWJ200485H, CWJ200610H, CWJ200614H, CWJ201046H, CWJ201242H, CWJ201372H, CWJ201487H, CWJ201509H, CWJ200418H, CWJ200465H, CWJ200532H, CWJ200561H, CWJ200613H, CWJ200642H, CWJ200651H, CWJ200754H, CWJ200918H, CWJ201279H, CWJ201328H, CWJ201158H, CWJ200404H, CWJ200424H, CWJ200425H, CWJ200452H, CWJ200453H, CWJ200568H, CWJ200579H, CWJ200640H, CWJ200684H, CWJ200710H, CWJ200714H, CWJ200757H, CWJ200783H, CWJ200784H, CWJ200785H, CWJ200805H, CWJ200809H, CWJ200858H, CWJ200924H, CWJ200933H, CWJ200980H, CWJ201036H, CWJ201057H, CWJ201070H, CWJ201129H, CWJ201177H, CWJ201276H, CWJ201280H, CWJ201293H, CWJ201296H, CWJ201315H, CWJ201421H, CWJ201427H, CWJ201464H, CWJ201485H, CWJ201486H, CWJ201529H, CWJ201534H, CWJ201536H, CWJ201562H, CWJ200414H, CWJ200443H, CWJ200488H, CWJ200549H, CWJ200553H, CWJ200624H, CWJ200652H, CWJ200723H, CWJ200733H, CWJ200735H, CWJ200736H, CWJ201053H, CWJ201336H, CWJ201343H, CWJ201514H, CWJ201525H, CWJ201261H, CWJ200525H, CWJ201229H, CWJ200463H, CWJ200690H, CWJ200338H, CWJ200552H, CWJ200987H, CWJ201094H, CWJ201195H, CWJ201220H, CWJ201398H, CWJ201479H, CWJ201457H, CWJ200426H, CWJ200816H, CWJ200938H, CWJ201272H, CWJ201554H, CWJ200741H, CWJ200989H, CWJ201088H, CWJ200541H, CWJ200965H, CWJ201172H, CWJ200529H, CWJ200570H, CWJ200780H, CWJ200756H, CWJ200402H, CWJ200415H, CWJ200508H, CWJ200511H, CWJ200540H, CWJ200574H, CWJ200591H, CWJ200607H, CWJ200625H, CWJ200643H, CWJ200678H, CWJ200693H, CWJ200724H, CWJ200737H, CWJ200770H, CWJ200967H, CWJ201005H, CWJ201008H, CWJ201009H, CWJ201019H, CWJ201102H, CWJ201138H, CWJ201175H, CWJ201194H, CWJ201294H, CWJ201340H, CWJ201387H, CWJ201428H, CWJ201442H, CWJ201480H, CWJ201491H, CWJ200446H, CWJ200484H, CWJ200866H, CWJ201146H, CWJ201163H, CWJ201545H, CWJ200589H, CWJ201317H, CWJ200407H, CWJ200413H, CWJ200645H, CWJ200877H, CWJ201182H, CWJ201225H, CWJ201362H, CWJ201564H, CWJ201063H, CWJ200577H, CWJ200600H, CWJ200975H, CWJ201001H, CWJ201419H, CWJ200901H, CWJ200817H, CWJ200801H, CWJ200688H, CWJ201540H, CWJ200462H, CWJ200585H, CWJ200820H, CWJ201149H, CWJ201174H, CWJ201369H, CWJ201513H, CWJ200460H, CWJ200595H, CWJ200731H, CWJ201233H, CWJ201268H, CWJ200483H, CWJ200392H, CWJ201084H, CWJ201147H, CWJ201470H, CWJ200873H, CWJ200963H, CWJ200995H, CWJ201037H, CWJ201112H, CWJ201122H, CWJ201124H, CWJ201159H, CWJ201216H, CWJ201361H, CWJ201378H, CWJ201408H, CWJ201453H, CWJ201468H, CWJ201477H, CWJ201543H, CWJ200522H, CWJ201039H, CWJ201151H, CWJ200405H, CWJ200429H, CWJ201012H, CWJ201263H, CWJ201267H, CWJ200471H, CWJ200543H, CWJ200544H, CWJ200566H, CWJ200580H, CWJ200636H, CWJ200707H, CWJ200828H, CWJ200831H, CWJ200895H, CWJ200900H, CWJ200974H, CWJ201185H, CWJ201207H, CWJ201357H, CWJ201504H, CWJ201505H, CWJ201568H, CWJ200862H, CWJ201355H, CWJ200390H, CWJ200554H, CWJ200834H, CWJ200839H, CWJ200951H, CWJ201073H, CWJ201360H, CWJ201431H, CWJ201496H, CWJ201062H, CWJ201354H, CWJ201550H, CWJ201244H, CWJ200337H, CWJ200433H, CWJ200475H, CWJ200649H, CWJ200701H, CWJ201335H, CWJ201353H, CWJ201358H, CWJ201215H, CWJ201047H, CWJ201348H, CWJ200422H, CWJ201071H, CWJ200618H, CWJ200972H, CWJ201086H, CWJ201098H, CWJ201105H, CWJ201134H, CWJ201148H, CWJ201214H, CWJ201281H, CWJ201383H, CWJ201430H, CWJ201443H, CWJ201444H, CWJ200394H, CWJ200740H, CWJ200666H, CWJ200431H, CWJ200481H, CWJ200482H, CWJ200719H, CWJ200936H, CWJ200949H, CWJ201078H, CWJ201145H, CWJ201301H, CWJ201320H, CWJ201334H, CWJ201363H, CWJ201441H, CWJ201446H, CWJ201460H, CWJ201511H, CWJ201556H, CWJ201017H, CWJ201050H, CWJ201236H, CWJ201499H, CWJ200659H, CWJ200773H, CWJ200410H, CWJ200569H, CWJ200572H, CWJ200576H, CWJ200608H, CWJ200699H, CWJ200729H, CWJ200798H, CWJ200821H, CWJ200823H, CWJ200833H, CWJ200843H, CWJ201068H, CWJ201082H, CWJ201110H, CWJ201153H, CWJ201154H, CWJ201161H, CWJ201456H, CWJ201481H, CWJ201498H, CWJ201510H, CWJ201531H, CWJ200427H, CWJ200838H, CWJ201096H, CWJ201197H, CWJ201250H, CWJ201121H, CWJ201155H, CWJ200408H, CWJ200655H, CWJ200656H, CWJ200657H, CWJ200767H, CWJ200768H, CWJ200964H, CWJ201180H, CWJ201181H, CWJ201333H, CWJ201411H, CWJ201437H, CWJ201463H, CWJ201555H, CWJ201561H, CWJ201228H, CWJ201329H, CWJ201471H, CWJ200685H, CWJ200700H, CWJ201273H, CWJ201389H, CWJ201397H, CWJ201473H, CWJ201494H, CWJ200622H, CWJ200730H, CWJ200763H, CWJ200395H, CWJ200401H, CWJ200464H, CWJ200494H, CWJ200509H, CWJ200876H, CWJ200977H, CWJ201240H, CWJ201299H, CWJ201321H, CWJ201326H, CWJ201422H, CWJ200537H, CWJ200957H, CWJ200400H, CWJ200439H, CWJ200473H, CWJ200567H, CWJ200691H, CWJ200829H, CWJ200849H, CWJ200910H, CWJ200922H, CWJ201092H, CWJ201128H, CWJ201198H, CWJ201461H, CWJ201184H, CWJ200507H, CWJ200677H, CWJ200885H, CWJ200898H, CWJ201003H, CWJ201035H, CWJ201196H, CWJ200870H, CWJ200982H, CWJ201038H, CWJ201109H, CWJ200459H, CWJ200504H, CWJ200704H, CWJ200593H, CWJ200887H, CWJ200890H, CWJ201055H, CWJ201264H, CWJ201414H, CWJ201423H, CWJ200814H, CWJ200815H, CWJ201051H, CWJ201277H, CWJ201306H, CWJ201307H, CWJ200795H, CWJ201274H, CWJ201490H, CWJ200435H, CWJ201076H, CWJ201337H, CWJ201396H, CWJ200556H, CWJ200440H, CWJ200550H, CWJ200738H, CWJ200765H, CWJ200830H, CWJ201258H, CWJ201386H, CWJ200990H, CWJ201020H, CWJ200524H, CWJ200728H, CWJ200836H, CWJ200976H, CWJ201201H, CWJ201403H, CWJ200683H, CWJ200810H, CWJ200811H, CWJ200888H, CWJ201010H, CWJ201119H, CWJ201150H, CWJ201191H, CWJ201231H, CWJ201248H, CWJ201265H, CWJ201271H, CWJ201288H, CWJ201374H, CWJ201518H, CWJ201552H, CWJ200658H, CWJ201027H, CWJ200620H, CWJ201565H, CWJ200391H, CWJ200406H, CWJ200456H, CWJ201104H, CWJ201188H, CWJ201227H, CWJ201500H, CWJ200491H, CWJ200812H, CWJ201028H, CWJ201080H, CWJ201178H, CWJ201283H, CWJ200799H, CWJ200939H, CWJ201238H, CWJ201006H, CWJ200409H, CWJ200441H, CWJ200497H, CWJ200498H, CWJ200499H, CWJ200638H, CWJ200669H, CWJ200671H, CWJ200673H, CWJ200718H, CWJ200769H, CWJ200791H, CWJ200840H, CWJ200863H, CWJ201183H, CWJ201232H, CWJ201310H, CWJ201058H, CWJ200941H, CWJ201404H, CWJ200749H, CWJ200751H, CWJ200752H, CWJ200590H, CWJ201278H, CWJ201311H, CWJ201546H, CWJ200586H, CWJ200398H, CWJ200772H, CWJ201385H, CWJ201488H, CWJ201551H, CWJ200621H, CWJ200781H, CWJ200787H, CWJ200788H, CWJ200991H, CWJ201024H, CWJ201077H, CWJ201316H, CWJ201339H, CWJ201259H, CWJ200709H, CWJ201226H, CWJ200725H, CWJ201166H, CWJ201210H, CWJ200444H, CWJ200461H, CWJ200517H, CWJ200521H, CWJ200535H, CWJ200548H, CWJ200597H, CWJ200599H, CWJ200708H, CWJ200717H, CWJ200802H, CWJ200925H, CWJ200942H, CWJ201095H, CWJ201186H, CWJ201260H, CWJ201303H, CWJ201304H, CWJ201309H, CWJ201338H, CWJ201371H, CWJ201382H, CWJ201400H, CWJ201412H, CWJ201447H, CWJ201448H, CWJ201467H, CWJ201502H, CWJ201507H, CWJ201522H, CWJ201547H, CWJ201559H, CWJ201563H, CWJ201251H, CWJ200973H, CWJ200641H, CWJ200869H, CWJ201016H, CWJ201168H, CWJ201544H, CWJ201406H, CWJ201452H, CWJ201489H, CWJ201482H, CWJ201056H, CWJ201395H, CWJ200992H, CWJ200555H, CWJ200932H, CWJ201097H, CWJ201131H, CWJ201318H, CWJ201379H, CWJ200428H, CWJ201290H, CWJ200904H, CWJ201090H, CWJ200774H, CWJ200786H, CWJ200850H, CWJ200928H, CWJ200953H, CWJ201052H, CWJ201380H, CWJ201436H, CWJ201014H, CWJ200835H, CWJ200493H, CWJ200528H, CWJ200531H, CWJ200623H, CWJ200635H, CWJ200672H, CWJ200711H, CWJ201018H, CWJ200523H, CWJ200970H, CWJ200742H, CWJ200909H, CWJ200437H, CWJ200616H, CWJ200695H, CWJ200748H, CWJ200800H, CWJ200917H, CWJ200929H, CWJ200948H, CWJ201091H, CWJ201213H, CWJ201350H, CWJ201351H, CWJ201465H, CWJ200851H, CWJ200986H, CWJ200442H, CWJ201391H, CWJ201064H, CWJ201193H, CWJ201116H, CWJ201381H, CWJ201206H, CWJ201127H, CWJ200438H, CWJ200474H, CWJ200505H, CWJ200565H, CWJ200670H, CWJ200755H, CWJ200759H, CWJ200779H, CWJ200853H, CWJ200923H, CWJ200946H, CWJ200954H, CWJ200959H, CWJ201021H, CWJ201111H, CWJ201113H, CWJ201176H, CWJ201190H, CWJ201432H, CWJ201433H, CWJ201438H, CWJ201476H, CWJ201492H, CWJ201503H; GTIN 00763000060107, Lot Serial Numbers: CWJ604565S, CWJ605053S, CWJ605035S, CWJ605052S, CWJ605252S, CWJ604317S, CWJ604667S, CWJ604080S, CWJ605534S, CWJ605535S, CWJ605558S, CWJ605573S, CWJ605579S, CWJ606047S, CWJ606055S, CWJ606068S, CWJ603916S, CWJ604241S, CWJ604243S, CWJ604263S, CWJ604264S, CWJ604272S, CWJ604273S, CWJ604318S, CWJ604368S, CWJ605036S, CWJ605037S, CWJ605050S, CWJ605060S, CWJ605061S, CWJ605066S, CWJ605239S, CWJ605244S, CWJ605254S, CWJ605408S, CWJ605410S, CWJ605414S, CWJ606171S, CWJ606064S, CWJ606168S, CWJ603406S, CWJ604070S, CWJ604072S, CWJ604238S, CWJ604319S, CWJ604322S, CWJ604325S, CWJ604360S, CWJ604365S, CWJ604374S, CWJ604674S, CWJ605033S, CWJ605034S, CWJ605058S, CWJ605062S, CWJ605065S, CWJ605243S, CWJ604137S, CWJ604261S, CWJ604323S, CWJ604265S, CWJ604268S, CWJ604327S, CWJ604329S, CWJ604359S, CWJ605251S, CWJ605547S, CWJ605555S, CWJ605576S, CWJ604328S, CWJ604666S, CWJ605045S, CWJ605238S, CWJ605248S, CWJ604366S, CWJ605551S, CWJ605580S, CWJ604246S, CWJ604247S, CWJ604248S, CWJ604254S, CWJ604375S, CWJ604513S, CWJ604554S, CWJ604659S, CWJ604665S, CWJ604670S, CWJ605396S, CWJ605404S, CWJ605241S, CWJ605068S, CWJ604251S, CWJ604075S, CWJ604661S, CWJ604068S, CWJ604082S, CWJ604252S, CWJ604316S, CWJ604364S, CWJ604510S, CWJ604678S, CWJ604138S, CWJ604663S, CWJ604249S, CWJ604367S, CWJ606044S, CWJ606061S, CWJ604078S, CWJ605543S, CWJ604269S, CWJ605557S, CWJ603914S, CWJ604256S, CWJ604361S, CWJ604509S, CWJ604547S, CWJ605413S, CWJ604079S, CWJ605556S, CWJ603405S, CWJ604135S, CWJ604512S, CWJ605063S, CWJ604372S, CWJ604330S, CWJ606052S, CWJ604266S, CWJ604084S, CWJ604333S, CWJ604384S, CWJ605039S, CWJ605249S, CWJ605412S, CWJ605577S, CWJ606054S, CWJ604250S, CWJ604136S, CWJ604083S, CWJ604563S, CWJ605548S, CWJ604258S, CWJ604320S, CWJ604519S, CWJ604560S, CWJ605031S, CWJ605032S, CWJ605054S, CWJ605055S, CWJ604141S, CWJ604143S, CWJ604382S, CWJ605399S, CWJ604073S, CWJ604128S, CWJ604253S, CWJ604071S, CWJ604129S, CWJ604245S, CWJ604086S, CWJ604673S, CWJ604236S, CWJ605057S, CWJ604383S, CWJ603407S, CWJ603915S, CWJ604259S, CWJ604373S, CWJ604522S, CWJ605571S, CWJ605549S, CWJ605574S, CWJ605567S, CWJ605566S, CWJ606066S, CWJ605561S, CWJ604332S, CWJ604664S, CWJ604260S, CWJ604074S, CWJ606048S, CWJ604559S, CWJ604067S, CWJ605536S, CWJ604334S, CWJ605400S, CWJ605405S, CWJ605545S, CWJ604237S, CWJ604511S, CWJ605049S, CWJ605240S, CWJ605583S, CWJ604126S, CWJ605582S, CWJ604257S, CWJ604552S, CWJ604555S, CWJ604676S, CWJ605256S, CWJ605397S, CWJ605398S, CWJ605564S, CWJ605565S, CWJ604557S, CWJ604672S, CWJ605550S, CWJ604145S, CWJ604562S, CWJ605041S, CWJ605042S, CWJ605043S, CWJ605401S, CWJ605569S, CWJ606167S, CWJ604069S, CWJ604558S, CWJ604660S, CWJ604385S, CWJ604235S, CWJ605237S, CWJ604144S, CWJ604335S, CWJ605538S, CWJ604295S, CWJ604553S, CWJ604240S, CWJ604140S, CWJ604378S, CWJ604370S, CWJ604662S, CWJ605059S, CWJ604271S, CWJ604520S, CWJ604675S, CWJ606045S, CWJ606062S, CWJ605040S, CWJ606056S, CWJ606057S, CWJ604545S, CWJ604525S; GTIN 00763000206369, Lot Serial Numbers: CWJ605981S, CWJ606001S, CWJ606339S, CWJ605970S, CWJ605394S, CWJ605729S, CWJ605856S, CWJ605485S, CWJ605493S, CWJ606120S, CWJ605921S, CWJ606249S, CWJ605511S, CWJ605635S, CWJ606348S, CWJ606203S, CWJ605874S, CWJ605833S, CWJ605679S, CWJ606221S, CWJ606240S, CWJ606243S, CWJ605901S, CWJ605428S, CWJ605385S, CWJ605478S, CWJ605483S, CWJ605484S, CWJ605813S, CWJ605837S, CWJ604788S, CWJ605225S, CWJ605630S, CWJ605333S, CWJ605424S, CWJ604816S, CWJ605426S, CWJ605629S, CWJ604787S, CWJ604817S, CWJ605854S, CWJ605860S, CWJ605337S, CWJ605883S, CWJ605151S, CWJ606136S, CWJ606208S, CWJ605625S, CWJ605626S, CWJ605853S, CWJ605513S, CWJ605332S, CWJ604833S, CWJ604930S, CWJ604961S, CWJ605267S, CWJ605364S, CWJ605417S, CWJ605496S, CWJ605812S, CWJ605286S, CWJ604949S, CWJ605166S, CWJ604967S, CWJ605158S, CWJ604848S, CWJ605222S, CWJ605430S, CWJ605500S, CWJ605591S, CWJ605023S, CWJ605914S, CWJ604835S, CWJ604862S, CWJ605205S, CWJ605235S, CWJ604818S, CWJ606148S, CWJ605220S, CWJ605264S, CWJ605180S, CWJ605273S, CWJ605437S, CWJ605605S, CWJ604950S, CWJ605795S, CWJ605796S, CWJ605383S, CWJ605810S, CWJ605486S, CWJ605593S, CWJ605594S, CWJ605645S, CWJ605687S, CWJ605748S, CWJ605802S, CWJ605828S, CWJ605864S, CWJ605898S, CWJ605622S, CWJ605196S, CWJ605510S, CWJ605636S, CWJ605025S, CWJ605073S, CWJ605117S, CWJ605185S, CWJ605452S, CWJ605800S, CWJ605899S, CWJ605197S, CWJ605705S, CWJ606129S, CWJ604921S, CWJ605213S, CWJ604789S, CWJ605608S, CWJ605126S, CWJ604793S, CWJ604872S, CWJ604993S, CWJ605295S, CWJ605371S, CWJ605479S, CWJ605638S, CWJ605690S, CWJ605739S, CWJ605797S, CWJ606128S, CWJ605750S, CWJ605751S, CWJ605752S, CWJ605857S, CWJ604922S, CWJ604925S, CWJ604927S, CWJ605124S, CWJ605152S, CWJ605271S, CWJ605272S, CWJ605515S, CWJ605516S, CWJ605623S, CWJ605624S, CWJ605633S, CWJ605909S, CWJ606122S, CWJ606198S, CWJ604825S, CWJ604827S, CWJ604934S, CWJ604935S, CWJ604936S, CWJ604956S, CWJ604958S, CWJ604959S, CWJ604960S, CWJ605304S, CWJ605307S, CWJ605319S, CWJ604901S, CWJ604946S, CWJ605270S, CWJ605354S, CWJ605451S, CWJ605595S, CWJ606123S, CWJ606153S, CWJ604943S, CWJ604845S, CWJ605146S, CWJ605269S, CWJ605897S, CWJ605011S, CWJ605020S, CWJ604829S, CWJ604832S, CWJ604953S, CWJ605621S, CWJ605896S, CWJ604843S, CWJ604814S, CWJ605206S, CWJ605207S, CWJ605208S, CWJ605209S, CWJ605322S, CWJ605324S, CWJ605325S, CWJ605327S, CWJ605328S, CWJ605423S, CWJ605585S, CWJ605586S, CWJ605588S, CWJ605596S, CWJ605600S, CWJ605805S, CWJ605806S, CWJ605807S, CWJ605808S, CWJ605916S, CWJ605917S, CWJ605918S, CWJ605919S, CWJ605920S, CWJ606251S, CWJ606252S, CWJ606253S, CWJ605109S, CWJ605347S, CWJ605349S, CWJ606101S, CWJ606146S, CWJ604892S, CWJ605492S, CWJ605199S, CWJ604986S, CWJ604776S, CWJ604933S, CWJ604962S, CWJ604972S, CWJ605024S, CWJ605096S, CWJ605121S, CWJ605431S, CWJ605601S, CWJ605650S, CWJ605862S, CWJ605875S, CWJ605306S, CWJ605504S, CWJ604836S, CWJ605005S, CWJ605499S, CWJ605501S, CWJ605503S, CWJ605505S, CWJ605512S, CWJ606116S, CWJ606117S, CWJ604945S, CWJ604812S, CWJ605699S, CWJ605348S, CWJ605184S, CWJ605619S, CWJ604942S, CWJ605380S, CWJ606383S, CWJ604929S, CWJ604951S, CWJ604954S, CWJ605187S, CWJ605461S, CWJ605160S, CWJ605105S, CWJ604932S, CWJ605865S, CWJ604919S, CWJ605095S, CWJ605118S, CWJ605119S, CWJ605162S, CWJ605163S, CWJ605164S, CWJ605168S, CWJ605233S, CWJ605234S, CWJ605263S, CWJ605301S, CWJ605438S, CWJ605442S, CWJ605482S, CWJ605507S, CWJ605508S, CWJ605602S, CWJ605603S, CWJ605604S, CWJ605749S, CWJ606215S, CWJ606224S, CWJ606225S, CWJ606378S, CWJ606379S, CWJ605420S, CWJ605498S, CWJ605616S, CWJ605326S, CWJ604964S, CWJ605266S, CWJ605323S, CWJ605816S, CWJ605872S, CWJ605873S, CWJ605352S, CWJ605422S, CWJ605287S, CWJ605356S, CWJ605357S, CWJ605358S, CWJ605519S, CWJ605520S, CWJ605597S, CWJ605217S, CWJ605359S, CWJ605517S, CWJ605598S, CWJ605691S, CWJ605617S, CWJ605219S, CWJ604895S, CWJ605148S, CWJ605201S, CWJ604837S, CWJ605259S, CWJ605261S, CWJ605275S, CWJ605288S, CWJ605309S, CWJ605418S, CWJ605421S, CWJ605432S, CWJ605433S, CWJ605434S, CWJ605435S, CWJ605480S, CWJ605487S, CWJ605488S, CWJ605489S, CWJ605490S, CWJ605491S, CWJ605644S, CWJ605675S, CWJ605871S, CWJ605274S, CWJ604920S, CWJ605153S, CWJ605127S, CWJ605130S, CWJ605643S, CWJ604912S, CWJ604944S, CWJ605027S, CWJ605069S, CWJ605181S, CWJ605260S, CWJ604905S, CWJ604907S, CWJ604915S, CWJ605258S, CWJ604931S, CWJ605026S, CWJ605178S, CWJ605198S, CWJ605200S, CWJ604839S, CWJ605131S, CWJ605228S, CWJ605204S, CWJ605618S, CWJ605839S, CWJ605840S, CWJ605842S, CWJ606009S, CWJ606010S, CWJ606011S, CWJ606018S, CWJ606020S, CWJ606071S, CWJ606072S, CWJ606079S, CWJ606081S, CWJ606082S, CWJ606083S, CWJ606089S, CWJ606112S, CWJ606113S, CWJ606114S, CWJ606356S, CWJ606357S, CWJ606358S, CWJ606359S, CWJ606360S, CWJ606361S, CWJ606362S, CWJ606363S, CWJ606364S, CWJ606365S, CWJ606366S, CWJ606367S, CWJ606393S, CWJ606394S, CWJ605425S, CWJ606347S, CWJ605214S, CWJ605159S, CWJ605905S, CWJ604939S, CWJ605202S, CWJ604785S, CWJ604794S, CWJ604801S, CWJ604842S, CWJ605167S, CWJ605218S, CWJ605316S, CWJ605648S, CWJ604809S, CWJ604963S, CWJ604989S, CWJ605087S, CWJ605183S, CWJ605277S, CWJ605350S, CWJ605351S, CWJ605481S, CWJ605686S, CWJ604808S, CWJ605122S, CWJ605753S, CWJ605809S, CWJ604437S, CWJ604781S, CWJ604805S, CWJ605110S, CWJ605111S, CWJ605227S, CWJ605276S, CWJ605697S, CWJ605906S, CWJ606259S, CWJ604830S, CWJ604821S, CWJ605818S, CWJ605203S, CWJ605108S, CWJ605506S, CWJ605018S, CWJ604438S, CWJ604998S, CWJ605186S, CWJ605888S, CWJ605682S, CWJ604970S, CWJ604977S, CWJ604988S, CWJ604969S, CWJ604971S, CWJ605019S, CWJ605224S, CWJ605236S, CWJ605315S, CWJ605360S, CWJ605374S, CWJ605379S, CWJ605475S, CWJ605634S, CWJ605667S, CWJ605678S, CWJ605733S, CWJ605734S, CWJ605849S, CWJ605855S, CWJ605859S, CWJ604896S, CWJ605177S, CWJ605210S, CWJ605215S, CWJ605216S, CWJ605231S, CWJ605232S, CWJ605279S, CWJ605280S, CWJ605283S, CWJ605429S, CWJ606206S, CWJ606207S, CWJ605949S, CWJ605155S, CWJ604866S, CWJ605393S, CWJ605175S, CWJ604820S, CWJ604861S, CWJ606248S, CWJ604890S, CWJ604894S, CWJ605879S, CWJ605176S, CWJ604904S, CWJ605660S, CWJ605707S, CWJ605708S, CWJ605709S, CWJ605710S, CWJ605727S, CWJ605728S, CWJ605825S, CWJ605987S, CWJ605996S, CWJ606130S, CWJ606143S, CWJ606258S, CWJ606287S, CWJ606288S, CWJ606309S, CWJ606369S, CWJ606373S, CWJ606374S, CWJ604913S, CWJ606144S, CWJ606231S, CWJ604889S, CWJ605824S, CWJ605182S, CWJ604782S, CWJ604856S, CWJ604857S, CWJ604876S, CWJ604926S, CWJ605278S, CWJ605590S, CWJ605637S, CWJ605979S, CWJ604774S, CWJ604882S, CWJ604886S, CWJ604855S, CWJ605190S, CWJ605195S, CWJ604800S, CWJ604806S, CWJ605194S, CWJ604779S, CWJ604924S, CWJ604928S, CWJ604850S, CWJ604859S, CWJ604804S, CWJ604870S, CWJ604897S, CWJ604878S, CWJ604772S, CWJ604860S, CWJ604871S, CWJ605976S, CWJ604879S, CWJ604911S, CWJ604796S, CWJ605704S, CWJ606385S, CWJ606386S, CWJ605799S, CWJ605904S, CWJ604908S, CWJ604973S, CWJ604974S, CWJ604898S, CWJ604899S, CWJ605925S, CWJ606110S, CWJ606150S, CWJ605495S, CWJ606154S, CWJ605518S, CWJ604770S, CWJ604771S, CWJ604773S, CWJ604775S, CWJ604777S, CWJ604807S, CWJ605133S, CWJ605134S, CWJ605135S, CWJ605142S, CWJ605143S, CWJ605144S, CWJ605145S, CWJ605339S, CWJ605341S, CWJ605342S, CWJ605343S, CWJ605345S, CWJ605366S, CWJ605367S, CWJ605370S, CWJ605373S, CWJ605376S, CWJ605377S, CWJ605378S, CWJ605381S, CWJ605382S, CWJ605384S, CWJ605786S, CWJ605787S, CWJ605788S, CWJ605789S, CWJ605790S, CWJ605290S, CWJ605292S, CWJ605293S, CWJ605294S, CWJ605296S, CWJ605298S, CWJ605299S, CWJ605329S, CWJ605331S, CWJ605335S, CWJ605336S, CWJ605338S, CWJ605340S, CWJ605387S, CWJ605388S, CWJ605389S, CWJ605391S, CWJ605737S, CWJ605740S, CWJ605741S, CWJ605742S, CWJ605743S, CWJ605744S, CWJ605745S, CWJ605763S, CWJ605765S, CWJ605766S, CWJ605769S, CWJ605774S, CWJ605775S, CWJ604841S, CWJ605657S, CWJ604834S, CWJ604937S, CWJ605157S, CWJ605193S, CWJ605226S, CWJ605281S, CWJ605289S, CWJ605415S, CWJ605439S, CWJ605798S, CWJ605903S, CWJ605076S, CWJ605103S, CWJ605179S, CWJ605211S, CWJ605212S, CWJ605284S, CWJ605467S, CWJ605468S, CWJ605497S, CWJ605641S, CWJ605646S, CWJ605681S, CWJ605814S, CWJ605815S, CWJ605831S, CWJ606398S, CWJ606399S, CWJ606320S, CWJ605819S, CWJ605822S, CWJ605823S, CWJ605726S, CWJ604849S, CWJ604797S, CWJ604893S, CWJ604824S, CWJ605792S, CWJ605985S, CWJ605456S, CWJ605843S, CWJ605386S, CWJ604891S, CWJ604917S, CWJ605935S, CWJ605936S, CWJ605939S, CWJ605940S, CWJ605941S, CWJ605171S, CWJ604918S, CWJ604888S, CWJ604981S, CWJ605997S, CWJ604968S, CWJ605028S, CWJ605321S, CWJ605344S, CWJ605375S, CWJ604811S, CWJ604938S, CWJ605680S, CWJ604847S, CWJ604903S, CWJ606109S, CWJ605663S, CWJ605455S, CWJ605986S, CWJ605170S, CWJ604826S, CWJ604828S, CWJ604887S, CWJ604941S, CWJ604952S, CWJ604955S, CWJ604957S, CWJ604965S, CWJ605012S, CWJ605013S, CWJ605014S, CWJ605015S, CWJ605017S, CWJ605021S, CWJ605022S, CWJ605938S, CWJ605632S, CWJ605353S, CWJ604786S, CWJ604881S, CWJ604883S, CWJ604884S, CWJ604885S, CWJ605776S, CWJ605777S, CWJ605778S, CWJ605779S, CWJ605780S, CWJ605781S, CWJ605782S, CWJ605783S, CWJ605784S, CWJ605930S, CWJ605931S, CWJ605932S, CWJ605933S, CWJ605934S, CWJ605942S, CWJ605943S, CWJ605944S, CWJ606012S, CWJ606115S, CWJ604923S, CWJ605688S, CWJ605998S, CWJ604979S, CWJ604985S, CWJ605016S, CWJ604909S, CWJ604910S, CWJ604914S, CWJ605372S, CWJ605453S, CWJ605454S, CWJ605647S, CWJ605653S, CWJ605926S, CWJ605994S, CWJ606111S, CWJ605793S, CWJ605937S, CWJ605652S, CWJ605999S, CWJ605791S, CWJ604790S, CWJ604791S, CWJ604851S, CWJ604863S, CWJ604865S, CWJ604873S, CWJ604874S, CWJ604877S, CWJ604975S, CWJ604976S, CWJ604978S, CWJ604980S, CWJ604984S, CWJ604987S, CWJ604990S, CWJ604991S, CWJ604992S, CWJ604994S, CWJ604999S, CWJ605004S, CWJ605008S, CWJ605009S, CWJ605010S, CWJ605070S, CWJ605071S, CWJ605072S, CWJ605074S, CWJ605077S, CWJ605078S, CWJ605079S, CWJ605080S, CWJ605081S, CWJ605082S, CWJ605083S, CWJ605084S, CWJ605086S, CWJ605088S, CWJ605089S, CWJ605090S, CWJ605091S, CWJ605092S, CWJ605093S, CWJ605094S, CWJ605097S, CWJ605098S, CWJ605099S, CWJ605100S, CWJ605106S, CWJ605112S, CWJ605113S, CWJ605114S, CWJ605115S, CWJ605116S, CWJ605123S, CWJ605125S, CWJ605128S, CWJ605129S, CWJ605132S, CWJ605136S, CWJ605137S, CWJ605138S, CWJ605139S, CWJ605141S, CWJ605147S, CWJ605150S, CWJ605172S, CWJ605173S, CWJ605174S, CWJ605302S, CWJ605310S, CWJ605312S, CWJ605313S, CWJ605314S, CWJ605318S, CWJ605363S, CWJ605365S, CWJ605368S, CWJ605369S, CWJ605392S, CWJ605395S, CWJ605441S, CWJ605443S, CWJ605444S, CWJ605445S, CWJ605446S, CWJ605447S, CWJ605448S, CWJ605449S, CWJ605450S, CWJ605457S, CWJ605458S, CWJ605459S, CWJ605460S, CWJ605462S, CWJ605463S, CWJ605464S, CWJ605465S, CWJ605466S, CWJ605469S, CWJ605470S, CWJ605471S, CWJ605472S, CWJ605473S, CWJ605474S, CWJ605509S, CWJ605523S, CWJ605524S, CWJ605525S, CWJ605526S, CWJ605527S, CWJ605528S, CWJ605529S, CWJ605530S, CWJ605531S, CWJ605533S, CWJ605607S, CWJ605609S, CWJ605610S, CWJ605611S, CWJ605612S, CWJ605613S, CWJ605614S, CWJ605654S, CWJ605656S, CWJ605658S, CWJ605659S, CWJ605662S, CWJ605664S, CWJ605665S, CWJ605666S, CWJ605668S, CWJ605669S, CWJ605670S, CWJ605672S, CWJ605674S, CWJ605695S, CWJ605700S, CWJ605701S, CWJ605702S, CWJ605703S, CWJ605714S, CWJ605715S, CWJ605716S, CWJ605717S, CWJ605718S, CWJ605719S, CWJ605720S, CWJ605731S, CWJ605732S, CWJ605756S, CWJ605758S, CWJ605759S, CWJ605760S, CWJ605761S, CWJ605762S, CWJ605767S, CWJ605768S, CWJ605770S, CWJ605771S, CWJ605772S, CWJ605773S, CWJ605820S, CWJ605969S, CWJ605973S, CWJ606004S, CWJ606005S, CWJ606006S, CWJ606008S, CWJ606022S, CWJ606024S, CWJ606025S, CWJ606027S, CWJ606028S, CWJ606029S, CWJ606030S, CWJ605722S, CWJ605989S, CWJ605661S, CWJ605984S, CWJ605992S, CWJ606316S, CWJ606321S, CWJ605640S, CWJ605651S, CWJ605725S, CWJ605894S, CWJ605221S, CWJ605229S, CWJ605257S, CWJ605265S, CWJ605334S, CWJ605419S, CWJ605476S, CWJ605477S, CWJ605522S, CWJ605532S, CWJ605587S, CWJ605592S, CWJ605606S, CWJ605620S, CWJ605631S, CWJ605639S, CWJ605642S, CWJ605671S, CWJ605676S, CWJ605677S, CWJ605683S, CWJ605684S, CWJ605685S, CWJ605696S, CWJ605706S, CWJ605711S, CWJ605712S, CWJ605713S, CWJ605721S, CWJ605723S, CWJ605735S, CWJ605736S, CWJ605747S, CWJ605754S, CWJ605821S, CWJ605832S, CWJ605835S, CWJ605851S, CWJ605861S, CWJ605867S, CWJ605868S, CWJ605869S, CWJ605870S, CWJ605877S, CWJ605881S, CWJ605882S, CWJ605884S, CWJ605887S, CWJ605892S, CWJ605893S, CWJ605900S, CWJ605915S, CWJ605927S, CWJ605928S, CWJ605968S, CWJ605972S, CWJ605975S, CWJ605977S, CWJ605978S, CWJ605980S, CWJ605983S, CWJ605990S, CWJ605995S, CWJ606002S, CWJ606003S, CWJ606021S, CWJ606075S, CWJ606076S, CWJ606084S, CWJ606104S, CWJ606118S, CWJ606119S, CWJ606124S, CWJ606125S, CWJ606126S, CWJ606127S, CWJ606134S, CWJ606135S, CWJ606137S, CWJ606138S, CWJ606145S, CWJ606149S, CWJ606152S, CWJ606155S, CWJ606197S, CWJ606199S, CWJ606200S, CWJ606201S, CWJ606202S, CWJ606210S, CWJ606220S, CWJ606222S, CWJ606226S, CWJ606232S, CWJ606236S, CWJ606238S, CWJ606239S, CWJ606247S, CWJ606254S, CWJ606256S, CWJ606260S, CWJ606265S, CWJ606266S, CWJ606273S, CWJ606286S, CWJ606291S, CWJ606292S, CWJ606305S, CWJ606306S, CWJ606338S, CWJ606343S, CWJ606349S, CWJ606352S, CWJ606353S, CWJ606354S, CWJ606368S, CWJ606370S, CWJ606375S, CWJ606381S, CWJ606402S, CWJ606403S, CWJ606410S, CWJ606411S, CWJ606293S, CWJ606317S, CWJ606319S, CWJ606404S, CWJ606405S, CWJ606408S, CWJ606409S, CWJ605230S, CWJ605303S, CWJ605320S, CWJ605627S, CWJ605886S, CWJ605959S, CWJ606301S
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Visa AF MRI™ S VR SureScan™ DVFC3D1Medtronic, Inc. · Pulse generator, permanent, implantable1 recall1 death22 reports
- Visia AF MRI™ S VR SureScan™ DVFC3D1Medtronic, Inc. · Pulse generator, permanent, implantable1 recall6 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92328).
- Z-1704-2023 · Open, ClassifiedICD-DR DDBB1D1 EVERA XT IS1/DF1 US, Model Number DDBB1D1; Implantable Cardioverter Defibrillators
- Z-1705-2023 · Open, ClassifiedICD-DR DDBB1D4 EVERA XT IS1/DF4 US, Model Number DDBB1D4; Implantable Cardioverter Defibrillators
- Z-1709-2023 · Open, ClassifiedICD-DR DDBC3D1 EVERA S IS1/DF1 GLOB, Model Number DDBC3D1; Implantable Cardioverter Defibrillators
- Z-1710-2023 · Open, ClassifiedICD-DR DDBC3D4 EVERA S IS1/DF4 GLOB, Model Number DDBC3D4; Implantable Cardioverter Defibrillators
- Z-1711-2023 · Open, ClassifiedICD-DR DDMB1D1 EVERA MRI XT US DF1, Model Number DDMB1D1; Implantable Cardioverter Defibrillators
- Z-1712-2023 · Open, ClassifiedICD-DR DDMB1D4 EVERA MRI XT IS-1/DF4 US, Model Number DDMB1D4; Implantable Cardioverter Defibrillators
- Z-1715-2023 · Open, ClassifiedICD-DR DDMC3D1 EVERA MRI S OUS/US DF1, Model Number DDMC3D1; Implantable Cardioverter Defibrillators
- Z-1716-2023 · Open, ClassifiedICD-DR DDMC3D4 EVERA MRI S IS-1/DF4 GLOB, Model Number DDMC3D4; Implantable Cardioverter Defibrillators
- Z-1717-2023 · Open, ClassifiedICD-DR DDMD3D1 PRIMO MRI, Model Number DDMD3D1; Implantable Cardioverter Defibrillators
- Z-1718-2023 · Open, ClassifiedICD-DR DDMD3D4 PRIMO MRI, Model Number DDMD3D4; Implantable Cardioverter Defibrillators
- Z-1721-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF1, Model Number DDPA2D1; Implantable Cardioverter Defibrillators
- Z-1723-2023 · Open, ClassifiedICD COBALT XT DR MRI IS1 DF4, Model Number DDPA2D4; Implantable Cardioverter Defibrillators
- Z-1725-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF1, Model Number DDPB3D1; Implantable Cardioverter Defibrillators
- Z-1726-2023 · Open, ClassifiedICD COBALT DR MRI IS1 DF4, Model Number DDPB3D4; Implantable Cardioverter Defibrillators
- Z-1727-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF1, Model Number DDPC3D1; Implantable Cardioverter Defibrillators
- Z-1728-2023 · Open, ClassifiedICD CROME DR MRI IS1 DF4, Model Number DDPC3D4; Implantable Cardioverter Defibrillators
- Z-1729-2023 · Open, ClassifiedCRT-D DTBA1D1 VIVA XT IS1/DF1 US, Model Number DTBA1D1; Implantable Cardioverter Defibrillators
- Z-1731-2023 · Open, ClassifiedCRT-D DTBA1D4 VIVA XT IS1/DF4 US, Model Number DTBA1D4; Implantable Cardioverter Defibrillators
- Z-1732-2023 · Open, ClassifiedCRT-D DTBA1Q1 VIVA QUAD XT IS4/DF1 US, Model Number DTBA1Q1; Implantable Cardioverter Defibrillators
- Z-1733-2023 · Open, ClassifiedCRT-D DTBA1QQ VIVA QUAD XT IS4/DF4 US, Model Number DTBA1QQ; Implantable Cardioverter Defibrillators
- Z-1741-2023 · Open, ClassifiedCRT-D DTBB1D1 VIVA S IS1/DF1 US, Model Number DTBB1D1; Implantable Cardioverter Defibrillators
- Z-1742-2023 · Open, ClassifiedCRT-D DTBB1D4 VIVA S IS1/DF4 US, Model Number DTBB1D4; Implantable Cardioverter Defibrillators
- Z-1743-2023 · Open, ClassifiedCRT-D DTBB1Q1 VIVA QUAD S IS4/DF1 US, Model Number DTBB1Q1; Implantable Cardioverter Defibrillators
- Z-1744-2023 · Open, ClassifiedCRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
- Z-1753-2023 · Open, ClassifiedCRTD DTMA1D1 CLARIA MRI US DF1, Model Number DTMA1D1; Implantable Cardioverter Defibrillators
- Z-1754-2023 · Open, ClassifiedCRTD DTMA1D4 CLARIA MRI US DF4, Model Number DTMA1D4; Implantable Cardioverter Defibrillators
- Z-1755-2023 · Open, ClassifiedCRTD DTMA1Q1 CLARIA MRI QUAD US DF1, Model Number DTMA1Q1; Implantable Cardioverter Defibrillators
- Z-1756-2023 · Open, ClassifiedCRTD DTMA1QQ CLARIA MRI QUAD US DF4, Model Number DTMA1QQ; Implantable Cardioverter Defibrillators
- Z-1761-2023 · Open, ClassifiedCRTD DTMB1D1 AMPLIA MRI US DF1, Model Number DTMB1D1; Implantable Cardioverter Defibrillators
- Z-1762-2023 · Open, ClassifiedCRTD DTMB1D4 AMPLIA MRI US DF4, Model Number DTMB1D4; Implantable Cardioverter Defibrillators
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗