Always-On Tip Tracked Instruments (SPiN Drive instruments)… recallZ-2268-2023 · Olympus Corporation of the Americas · initiated May 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
FDA root cause category: Component design/selection
What the company is doing
Olympus issued "URGENT - Medical Device Field Corrective Action" on 5/15/23. Letter states reason for recall, health risk and action to take: Olympus implemented a new internal component and has observed a reduction in complaints for the above-mentioned issue. However, to minimize the duration of potential procedure delay, Olympus is revising the Instruction for Use to add the below precaution statement: It is required to have additional Always-On Tip Tracked instruments available for all procedures as tracking may be lost when: Connecting the instrument to the SPiN Thoracic Navigation System Inserting and withdrawing the instrument from the working channel of the bronchoscope during procedure Navigating and sampling in tortuous patient anatomy Action steps to be taken by the end user: Our records indicate that you have purchased one or more of the affected products. Olympus requires you to take the following action: 1. Notify and ensure your users are knowledgeable about this additional precaution requiring additional instruments be available for all procedures. 2. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. a) Go to https://olympusamerica.com/recall. b) Enter the recall number 0425 c) Complete the form as instructed. 3. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints associated with the use of this device, including patient injuries and adverse events, to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1 to report complaints). Contact (647) 999-3203 or at [email protected] for any additional information or support concerning this matter.
Recall details
- Recall number
- Z-2268-2023
- Status
- Open, Classified
- Initiated by the company
- May 15, 2023
- Posted by the FDA
- Jul 28, 2023
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 1908.2 units
- Product code
- JAK
- FDA submissions
- K122106, K170023
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: CN, DE, HK, IT, KR, SG, TH & TW.
Product description
Always-On Tip Tracked Instruments (SPiN Drive instruments) 22ga SPiN Flex Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5410
Code information (models, lots, serial numbers)
UDI-DI: 00815686020637 Lot Number: 06332230201 and below
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Always-On Tip Tracked 22ga SPiN Flex Needle, 12mm L, 1.8mm OD INS-5410Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls20 reports
Other devices under the same FDA submission
30 devices cleared under K122106, K170023 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Always-On Tip Tracked Triple Needle Brush, 12mm L, 1.8mm OD INS-5300Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls16 reports
- Always-On Tip Tracked Brush, 15mm L, 1.8mm OD INS-0352Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls15 reports
- Always-On Tip Tracked Forceps, 1.8mm OD, Serrated Cup INS-0372Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls7 reports
- Always-On Tip Tracked 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD INS-0392Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls7 reports
- Always-On Tip Tracked 19ga ANSO Histology Needle, 12mm L, 1.8mm OD INS-0382Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls6 reports
- SPiN Xtend 2.0mm OD, 21ga INS-5450Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed2 recalls4 reports
- ig4 System, 120V, 60Hz SYS-3000Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall1 death11 reports
- SPiNPlanning Laptop Workstation SYS-0185Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall4 reports
- SPiN Access Catheter, 90 Degree For Use With Olympus Scopes INS-5900Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall3 reports
- vPad2 Patient Tracker INS-0050Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall2 reports
- SPiN Access Catheter, 180 Degrees For Use With Olympus Scopes INS-5905Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall2 reports
- ig4 System V2, 120V, 60Hz SYS-2000Veran Medical Technologies Inc. · System, X-Ray, Tomography, Computed1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 92383).
- Z-2262-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anatomical structures where computed tomograph…
- Z-2263-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8 mm OD, Oval Cup-Intended for use in clinical interventions and for anatomical structures where computed tomo…
- Z-2264-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical interventions and for anatomical structures where computed t…
- Z-2265-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) 19ga ANSO Histology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures whe…
- Z-2266-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures wher…
- Z-2267-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where comp…
- Z-2269-2023 · Open, ClassifiedAlways-On Tip Tracked Instruments (SPiN Drive instruments) SPiN Xtend 2.0mm OD, 21ga Intended for use in clinical interventions and for anatomical structures where computed tomogr…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page