Skip to content

vPad2 Patient Tracker INS-0050Recalls & reported problems

Veran Medical Technologies Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815686020408), received between Oct 11, 2020 and Apr 25, 2021 — classified as 2 injuries.

1 FDA recall record name this device's identifier, 1 still open.

12 other recalls cover devices cleared under the same FDA submission (K122106); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 10, 2023Class IIOpen, Classified

    Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

    Z-2664-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Bradycardia1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DR. WAS UNABLE TO SUCCESSFULLY COMPLETE REGISTRATION. THE CASE DID NOT CONTINUE PAST THIS POINT. THEY WERE UNABLE TO COMPLETE REGISTRATION SO THEY ABORTED THE PROCEDURE ENTIRELY. THIS WAS A PERC CASE, SO THEY DID NOT COMPLETE THE CASE WITHOUT VERAN. THE PATIENT WAS ALREADY UNDER ANESTHESIA. THEY WERE WOKEN UP AND A FOLLOW-UP PROCEDURE WAS SCHEDULED. TECHNICAL SUPPORT WAS CONTACTED BEFORE THE PROCEDURE WAS ABORTED AND THEY WERE UNABLE TO FIX THE ISSUE. AFTER THE PROCEDURE WAS ABORTED, TECHNICAL SUPPORT FOLLOWED UP AND IT WAS DETERMINED THAT THE ISSUE WAS MOST LIKELY ISSUES WITH THE CT SCAN. IT WAS PRONE AND THE ORIENTATION WAS INCORRECT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11721780 ↗
  • InjuryOct 11, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    WHILE ANESTHESIA WAS PUTTING THE PATIENT UNDER SHE HAD BRACHYCARDIA AND STARTED CRASHING, COMPRESSIONS WERE STARTED AND THEY CALLED FOR A CRASH CART. VERAN REP STEPPED OUT OF THE ROOM. ONLY VPADS HAD BEEN PLACED FOR CT, THE PADS WERE NOT PLUGGED INTO THE CART YET, TOWER STILL OFF TO THE SIDE IN THE ROOM. THE CRASH CART WAS NOT UTILIZED BUT THE PATIENT WAS INTUBATED AND ADMITTED TO THE ICU. STAFF INFORMED VERAN REP THAT THE PATIENT IS AWAKE AND TALKING IN THE ICU. PACEMAKER WAS PLACED THE FOLLOWING MONDAY, (B)(6) 2020.

    Patient problems: Bradycardia

    MAUDE report 10660143 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122106. One submission often covers many models, so check the product and lot details in each record.

  • Z-2662-2023 · Aug 10, 2023 · Open, Classified
    Veran : Endobronchial Models: INS-5500 View Optical Probe INS-5450 SPiN Xtend 21ga, 1.8 OD, 5/Bx INS-5440 View Peripheral Catheter, 3.8mm OD, 2.5mm WC INS-5411 0.016
  • Z-2659-2023 · Aug 10, 2023 · Open, Classified
    Veran:SPiN Thoracic Navigation System Models: SYS-4230 SPiN Thoracic Navigation System - 230V 50Hz SYS-4000 SPiN Thoracic Navigation System SYS-3230 ig4 System, 230V, 50H
  • Z-2663-2023 · Aug 10, 2023 · Open, Classified
    Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "
  • Z-2661-2023 · Aug 10, 2023 · Open, Classified
    Veran : Percutaneous Models: INS-5620 SPiN Perc Biopsy Needle Guide Kit - 20cm, 5/Bx (includes INS-5028)" INS-5611 SPiN Perc Localization Needle - 2cm, 5/Bx INS-5610 SPiN
  • Z-2266-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) 21ga ANSO Cytology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures wher
  • Z-2263-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8 mm OD, Oval Cup-Intended for use in clinical interventions and for anatomical structures where computed tomo
  • Z-2269-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) SPiN Xtend 2.0mm OD, 21ga Intended for use in clinical interventions and for anatomical structures where computed tomogr
  • Z-2268-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) 22ga SPiN Flex Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where co
  • Z-2265-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) 19ga ANSO Histology Needle, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures whe
  • Z-2262-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) Brush, 15mm L, 1.8mm OD Intended for use in clinical interventions and for anatomical structures where computed tomograph
  • Z-2267-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where comp
  • Z-2264-2023 · May 15, 2023 · Open, Classified
    Always-On Tip Tracked Instruments (SPiN Drive instruments) Forceps, 1.8mm OD, Serrated Cup-Intended for use in clinical interventions and for anatomical structures where computed t

Other System, X-Ray, Tomography, Computed devices

All System, X-Ray, Tomography, Computed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page