REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter… recallZ-2224-2023 · Datascope Corp. · initiated Jun 23, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.
FDA root cause category: Under Investigation by firm
What the company is doing
Firm notified customers on June 23, 2023 via letter. Customers were made aware of the potential for the device to break during use and the risk of using the device. Customers were given alternate solutions to continue using the IAB Catheter with alternative sheath/dilator combinations. Customers may also return their affected product for a credit. The firm will continue to ship product to customers who choose to continue receiving it. However, product currently being shipped is potentially affected by the same issue. Datascope/Getinge has identified the issue with the introducer dilator and is working with the supplier towards a resolution. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Customer Service at 1-888-943-8872, options 2, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Recall details
- Recall number
- Z-2224-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 23, 2023
- Posted by the FDA
- Jul 21, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 321,609 total kits
- Product code
- DYB
- FDA submissions
- K002365
- Distribution
- Domestic distribution nationwide. International distribution worldwide.
Product description
REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10
Code information (models, lots, serial numbers)
Product Description UDI REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. 10607567106656 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. 10607567106694 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. 10607567107943
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Reinforced Introducer Set for Sensation Plus & Mega 8Fr 0684-00-0403-10Datascope Corp. · Introducer, Catheter2 recalls1 report
- Reinforced Introducer Set for Linear 7.5Fr & Mega 7.5Fr 0684-00-0403-05Datascope Corp. · Introducer, Catheter2 recalls0 reports
- Reinforced Introducer Set for Sensation 7Fr 0684-00-0403-06Datascope Corp. · Introducer, Catheter2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92527).
- Z-2216-2023 · Open, ClassifiedYAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description …
- Z-2217-2023 · Open, ClassifiedSENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2218-2023 · Open, ClassifiedTRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2219-2023 · Open, ClassifiedMEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2220-2023 · Open, ClassifiedMEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2221-2023 · Open, ClassifiedSENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2222-2023 · Open, ClassifiedSENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
- Z-2223-2023 · Open, ClassifiedLINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.