VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS… recallZ-2247-2023 · Johnson & Johnson Surgical Vision, Inc. · initiated May 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Johnson & Johnson Surgical Vision, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

FDA root cause category: Process control

What the company is doing

On 5/30/23, recall notices were mailed to customers who were asked to do the following: 1) Discontinue use and return affected devices to the recalling firm. 2) Share this notice with anyone within your organization that needs to be informed and to any organization where the potentially affected products have been transferred. 3) Complete and return the customer reply form to [email protected] In addition, distributors were asked to distribute the recall notice to their customers. Questions, adverse events, and complaints can be directed to the firm at 1-877-266-4543 Option 2.

Recall details

Recall number
Z-2247-2023
Status
Open, Classified
Initiated by the company
May 30, 2023
Posted by the FDA
Jul 28, 2023
Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Quantity
148,941
Product code
HQC
FDA submissions
K203060
Distribution
Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Korea¿(South), Kuwait, Lebanon, Malaysia, Morocco, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela.

Product description

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All lots within their expiration date as of 5/29/2023. Part Number/UDI-DI/Lot: VRT-AI/05050474700901/60298931, 60302489, 60302939, 60302941, 60302953, 60302954, 60304170, 60304175, 60304176, 60305660, 60305661, 60305662, 60305665, 60306934, 60306935, 60306936, 60306940, 60306942, 60308193, 60309850, 60309844, 60309845, 60309853, 60314642, 60314676, 60314677, 60316112, 60351692, 60372490, 60372491, 60374175, 60374945, 60376930, 60378840, 60378842, 60378844, 60400795, 60401984, 60413115, 60413116, 60413117, 60414191, 60415077, 60415078, 60416304, 60420003, 60421952, 60425389, 60426314, 60426315, 60428217, 60428216, 60429442, 60433653, 60433654, 60433655, 60435921, 60435922, 60435923, 60437988, 60437989, 60440239, 60440240, 60442450, 60442451, 60442452, 60442453, 60442454, 60444535, 60446059, 60448847, 60448856, 60448857, 60376931, 60379976, 60429443. VRT-AF/05050474700895/60304172, 60304173, 60304174, 60305657, 60305658, 60305659, 60306930, 60306931, 60306932, 60306933, 60308189, 60308190, 60309848, 60309847, 60314640, 60314641, 60314674, 60314675, 60378839, 60420004, 60446054, 60446055