Skip to content

VERITAS ADVANCED FLUIDICS PACK VRT-AFRecalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · VERITAS ADVANCED FLUIDICS PACK · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. May 30, 2023Open, Classified

    Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

    Z-2247-2023 · Root cause: Process control · Recall details · FDA record ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page