TRACHEAL INTUBATION FIBERSCOPE recallZ-2341-2023 · Olympus Corporation of the Americas · initiated May 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible
FDA root cause category: Labeling Change Control
What the company is doing
Olympus issued two Urgent: Medical Device Corrective Action letters: (1) VISERA TRACHEAL INTUBATION VIDEOSCOPE OLYMPUS LF TYPE V (LF-V) TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-P (LF-P) (2)ENDOSCOPE REPROCESSOR Affected Models: OER-ProTM, OER-MiniTM and OER-EliteTM. Letters states reason for recall, health risk and action to take: Our records indicate that your facility has purchased one or more LF-V and LF-P scopes. (1)Olympus requests you to take the following actions: 1. Ensure all personnel are completely knowledgeable and thoroughly trained on the changes to the reprocessing instructions described in this letter. 2. If you are using OER for high level disinfection of your LF-V and LF-P, please transition to another compatible method of reprocessing as soon as possible. 3. Locate and replace any on-hand copies of the LF-V and LF-P Instructions for Use with the updated version. Obtain a copy of the updated IFU by accessing our Olympus Connect website: How to access the Instructions for Use a) Log in to www.OlympusConnect.com b) Select Product Support , then Instruction Manuals c) Navigate to the document entitled LF-V Instruction Manual and LF-P Instruction Manual If you require additional information about the updated Instruction for use, or need assistance with accessing Olympus Connect, please contact our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). The TAC department can assist you in answering questions on the new Instructions Manual and new steps. 4. If a Scope ID tag (MAJ-1545) is used with your LF-V and LF-P, please dispose of your MAJ-1545 since the LF-V and LF-P cannot be used with OERs. 5. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email ([email protected]) or Fax (866-842-0470). For any questions about the acknowledgement form, please call the Sedgwick team at 888-912-7350. (2)Our records indicate that your facility has purchased one or more of the afore mentioned endoscope
Recall details
- Recall number
- Z-2341-2023
- Status
- Open, Classified
- Initiated by the company
- May 26, 2023
- Posted by the FDA
- Aug 4, 2023
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 686 units
- Product code
- EOQ
- FDA submissions
- K221631
- Distribution
- Nationwide
Product description
Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other ancillary equipment for management of Endotracheal Tube
Code information (models, lots, serial numbers)
UDI-DI: 04953170051098 All serial numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-POlympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)2 recalls3 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92633).
- Z-2340-2023 · Open, ClassifiedVISERA Tracheal Intubation Videoscope OLYMPUS LF Type V- For airway management which includes endoscopic observation to access airway anatomy, endotracheal/endobronchial intubation…
- Z-2342-2023 · Open, ClassifiedOlympus Endoscope Reprocessor OER-Pro-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.
- Z-2343-2023 · Open, ClassifiedOlympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.
- Z-2344-2023 · Open, ClassifiedOlympus Endoscope Reprocessor OER-Elite-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page