Skip to content

ENDOSCOPE REPROCESSOR recallZ-2342-2023 · Olympus Corporation of the Americas · initiated May 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible

FDA root cause category: Labeling Change Control

What the company is doing

Olympus issued two Urgent: Medical Device Corrective Action letters: (1) VISERA TRACHEAL INTUBATION VIDEOSCOPE OLYMPUS LF TYPE V (LF-V) TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-P (LF-P) (2)ENDOSCOPE REPROCESSOR Affected Models: OER-ProTM, OER-MiniTM and OER-EliteTM. Letters states reason for recall, health risk and action to take: Our records indicate that your facility has purchased one or more LF-V and LF-P scopes. (1)Olympus requests you to take the following actions: 1. Ensure all personnel are completely knowledgeable and thoroughly trained on the changes to the reprocessing instructions described in this letter. 2. If you are using OER for high level disinfection of your LF-V and LF-P, please transition to another compatible method of reprocessing as soon as possible. 3. Locate and replace any on-hand copies of the LF-V and LF-P Instructions for Use with the updated version. Obtain a copy of the updated IFU by accessing our Olympus Connect website: How to access the Instructions for Use a) Log in to www.OlympusConnect.com b) Select Product Support , then Instruction Manuals c) Navigate to the document entitled LF-V Instruction Manual and LF-P Instruction Manual If you require additional information about the updated Instruction for use, or need assistance with accessing Olympus Connect, please contact our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). The TAC department can assist you in answering questions on the new Instructions Manual and new steps. 4. If a Scope ID tag (MAJ-1545) is used with your LF-V and LF-P, please dispose of your MAJ-1545 since the LF-V and LF-P cannot be used with OERs. 5. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email ([email protected]) or Fax (866-842-0470). For any questions about the acknowledgement form, please call the Sedgwick team at 888-912-7350. (2)Our records indicate that your facility has purchased one or more of the afore mentioned endoscope

Recall details

Recall number
Z-2342-2023
Status
Open, Classified
Initiated by the company
May 26, 2023
Posted by the FDA
Aug 4, 2023
Recalling firm
Olympus Corporation of the Americas
Quantity
4244 units
Product code
FEB
FDA submissions
K103264
Distribution
Nationwide

Product description

Olympus Endoscope Reprocessor OER-Pro-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.

Code information (models, lots, serial numbers)

UDI-DI: 04953170258589 All serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92633).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page