Constellation recallZ-2406-2023 · Alcon Research, LTD. · initiated Jun 28, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alcon Research, LTD. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
FDA root cause category: Process control
What the company is doing
Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at [email protected]. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at [email protected] or by phone at 1-800-757-9780, Option 5.
Recall details
- Recall number
- Z-2406-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 28, 2023
- Posted by the FDA
- Aug 16, 2023
- Recalling firm
- Alcon Research, LTD.
- Quantity
- 10,580 units
- Product code
- HQC
- FDA submissions
- K141065
- Distribution
- Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Product description
Constellation Procedure Pak
Code information (models, lots, serial numbers)
Catalog No. 8065750958; Product Code 100007216; UDI-DI: 380657509584; Lots 14XM2J, 14Y1H6, 150HYJ, 150HYK, 151R72, 152CH0, 15309X, 156HWN, 15A13R, 15A13T, 15CF2X, 15CF2Y, 15CF30, 15CJ47, 15CJ48, 15CWAM, 15CWAN, 15D649.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Constellation 8065750958Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall8 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92698).
- Z-2407-2023 · Open, ClassifiedCenturion FMS Pak
- Z-2408-2023 · Open, ClassifiedInfinity FMS Pak
- Z-2409-2023 · Open, ClassifiedLegion FMS Pak
- Z-2410-2023 · Open, ClassifiedPhaco Standalone Tips
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page