Constellation 8065750958Recalls & reported problems
Alcon Laboratories, Inc. · 20 GA Fragmentation Pak · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380657509584), received between Aug 13, 2021 and Feb 10, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 28, 2023Open, Classified
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
Z-2406-2023 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Break7
- Adverse Event Without Identified Device or Use Problem1
- Failure to Cut1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
- Capsular Bag Tear1
- Retinal Detachment1
- Vitreous Hemorrhage1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2026Break
A CUSTOMER REPORTED THAT THE FRAGMENTATION TIP FRACTURED AND SHIFTED IN THE EYE OF THE PATIENT DURING SURGERY. THE PROCEDURE TYPE AND OTHER SURGERY DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24301737 ↗InjuryDec 10, 2025Break; Break
A CUSTOMER REPORTED THAT THE METAL SHAFT OF THE FRAGMATOME CRACKED HALF WAY AND BROKE AWAY WHILE IN THE RIGHT EYE OF A FEMALE ELDERLY PATIENT DURING VITRECTOMY AND LENSECTOMY SURGERY. WHEN THE TIP BROKE OFF, IT WAS REQUIRED TO ENLARGE THE SCLERAL INCISION AND CREATE A CORNEAL INCISION AND A COUPLE OF AREAS THAT NEEDED TO BE LASERED WHERE THE TIP HAD LANDED. PATIENT HAD VITREOUS HEMORRHAGE AND HAD TO SUTURE CORNEAL INCISION ALONG WITH LASER AND A GAS BUBBLE. THEY HAD TO OPEN A NEW FRAGMATOME TIP AND REPLACED IT TO TAKE OUT REST OF LENS AND TIP HAD TO BE REMOVED FROM THE EYE. THE CURRENT PATIENT CONDITION WAS UNKNOWN.
Patient problems: Vitreous Hemorrhage
MAUDE report 23764907 ↗MalfunctionOct 16, 2024Break; Break
A PHYSICIAN REPORTED THAT THE FRAGMENTATION TIP GOT BROKEN AND FELL INTO THE RETINA DURING SURGERY. SURGEON MANAGED AND REMOVED BROKEN TIP FROM RETINA. IT WAS UNKNOWN IF THE SURGERY WAS COMPLETED. THERE WAS NO INFORMATION ABOUT PROCEDURE TYPE AND PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 20459749 ↗MalfunctionNov 23, 2023Break; Break
A NURSE REPORTED THAT THE FRAGMENTATION TIP WAS BROKE INSIDE PATIENT'S EYE DURING THE PROCEDURE. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. BROKEN TIP REMOVED FROM THE PATIENT¿S EYE DURING THE SAME PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18197381 ↗InjurySep 29, 2023Failure to Cut; Adverse Event Without Identified Device or Use Problem; Failure to Cut; Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED THAT FRAG WAS NOT ABLE TO BREAK UP THE CATARACT EFFICIENTLY DURING SURGERY. THE PATIENT EXPERIENCED THE EVENT NUCLEUS DROPPING POSTERIORLY. THE CURRENT OUTCOME OF THE PATIENT WAS UNKNOWN. THIS IS FIRST OF FOUR REPORTS FOR THIS ISSUE AND REPRESENTS THE OPHTHALMIC PROBE.
Patient problems: Capsular Bag Tear
MAUDE report 17841663 ↗MalfunctionNov 1, 2021Break; Break
THE SAMPLE WAS NOT AVAILABLE FOR TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12729408 ↗MalfunctionOct 25, 2021Break; Break
A PHYSICIAN REPORTED THE FRAGMENTATION TIPS NEEDLE BROKE AND SCATTERED IN THE PATIENT'S EYE DURING A SURGERY. HE TRIED TO ASPIRATE BUT COULD NOT GET THEM ALL. THERE WAS NO PATIENT HARM REPORTED. THE SAMPLE IS NOT AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12693641 ↗InjuryAug 13, 2021Break; Break
A NURSE REPORTED A FRAGMENTATION NEEDLE BROKE AND FELL TO THE BACK OF THE PATIENT¿S EYE DURING A PROCEDURE AND WHEN IT WAS RETRIEVED, IT CAUSED A RETINAL DETACHMENT.
Patient problems: Retinal Detachment
MAUDE report 12321172 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page