CENTURION, INTREPID recallZ-2407-2023 · Alcon Research, LTD. · initiated Jun 28, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alcon Research, LTD. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.
FDA root cause category: Process control
What the company is doing
Customers were sent a Voluntary Medical Device Correction Notification by mail and dated 06/29/2023. Customers are asked to follow the risk mitigation precautions detailed in the notice to avoid the instance of overtightening the Phaco Tip and generating plastic particles. Additionally, customers are asked to forward the recall notification to all departments or organizations that use recalled devices. Customers are asked to complete and return the provided Response Form by fax at 817-302-4337 or email at [email protected]. Customers who have experienced any adverse events or quality related issues are to contact alcon via email at [email protected] or by phone at 1-800-757-9780, Option 5.
Recall details
- Recall number
- Z-2407-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 28, 2023
- Posted by the FDA
- Aug 16, 2023
- Recalling firm
- Alcon Research, LTD.
- Quantity
- 87,839 units
- Product code
- HQC
- FDA submissions
- K121555
- Distribution
- Worldwide distribution - US Nationwide and the countries of Aland Islands, Albania, Algeria, Andorran, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bermuda, Botswana, Brazil, Brunei Daruss., Bulgaria, Burma, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, French Guayana, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Isle of Man, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Fed., Saint Martin, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, USA, USVI, Utd. Arab Emir., Uzbekistan, Vietnam, and Yemen.
Product description
Centurion FMS Pak
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
Catalog No. 8065000133, 8065000134, 8065000138, 8065000139, 8065752192, 8065752193, 8065752200, 8065752201, 8065752202, 8065752203, 8065752216, 8065752217, 8065752245, 8065752917, 8065752918, 8065753134, 8065753135, 8065753136, 8065753159; Product Code 100241014, 100241015, 100241019, 100241020, 100055107, 100055110, 100009955, 100009963, 100009964, 100009965, 100009960, 100009961, 100010766, 100056645, 100055945, 100207171, 100207170, 100208838, 100208338; UDI-DI: 380650001337, 380650001344, 380650001382, 380650001399, 380657521920, 380657521937, 380657522002, 380657522019, 380657522026, 380657522033, 380657522163, 380657522170, 380657522453, 380657529179, 380657529186, 380657531349, 380657531356, 380657531363, 380657531592; Lots 14WW1F, 14WW1H, 14WW1J, 14X6CN, 14X6CP, 14X6CR, 14X6CT, 14X6CU, 14X6CV, 14XPE9, 14XPEA, 14XPEC, 14XPED, 14XPEE, 14XPEF, 14XPEH, 14Y8C2, 14Y8C3, 14Y8C4, 14Y8C5, 14Y8C6, 14Y8C7, 14YLMF, 14YLMH, 14YLMJ, 14YLMK, 14YLML, 14YLMM, 14YLMN, 14YLMP, 14YXHY, 14YXJ0, 14YXJ1, 14YXJ3, 14YXJ4, 14YXJ5, 14YXJ6, 14YXJ7, 150HDA, 150HDC, 150HDD, 150HDE, 150HDF, 150HDH, 150HDJ, 150HDK, 150HDL, 150HDM, 150HDN, 150HDP, 150HDR, 151V90, 151V91, 151V92, 151V93, 151V94, 151V95, 151V96, 151V97, 152F5H, 152F5K, 152F5L, 152F5M, 152F5N, 152F5P, 152F5R, 152F5T, 152F5V, 152F5W, 152F5X, 1539WP, 1539WR, 1539WT, 1539WU, 1539WV, 1539WW, 1539WX, 1539WY, 1540J9, 1540JA, 1540JC, 1540JD, 1540JE, 1540JF, 1540JH, 1540JJ, 1547AV, 1547AW, 1547AX, 1547AY, 1547C0, 1547C1, 1547C2, 1547C3, 1547C4, 1547C5, 1547C6, 1547C7, 1547C8, 1547C9, 154YDE, 154YDF, 154YDH, 154YDJ, 154YDK, 154YDL, 154YDM, 154YDN, 154YDP, 154YDR, 155HJP, 155HJR, 155HJT, 155HJU, 155HJV, 155HJW, 1565PX, 1565PY, 1565R0, 1565R1, 1565R2, 1571HV, 1571HW, 1571HX, 1571HY, 1571J0, 1571J1, 1571J2, 1571J3, 1571J4, 1571J5, 1571J6, 157DN2, 157DN3, 157DTR, 157DTT, 157DTU, 157DTW, 157DU2, 157DU4, 157DU5, 157DU6, 157DU7, 157DU8, 157DU9, 158CXM, 158CXN, 158CXP, 158CXR, 158CXT, 158CXU, 158CXV, 158CXW, 158CXX, 158CXY, 158VLV, 158VLW, 158VLX, 158VLY, 158VM0, 158VM1, 158VM2, 158VM3, 1599H3, 1599H4, 1599H5, 1599H6, 1599H7, 1599H8, 1599HF, 1599HH, 1599HJ, 1599HK, 1599HL, 1599HM, 1599HN, 1599HP, 1599HT, 159PPH, 159PPJ, 159PPK, 159PPL, 159PPM, 159PPN, 159PPP, 159PPR, 159PPT, 159PPU, 159PPV, 159PPW, 159PPX, 159PPY, 159PR0, 159PR1, 15A8H0, 15A8H2, 15A8H3, 15A8H4, 15A8H5, 15A8H6, 15A8H7, 15A8H8, 15A8H9, 15A8HA, 15A8HC, 15A8HD, 15A8HE, 15A8HF, 15A8HH, 15A8HJ, 15AV8Y, 15AV90, 15AV91, 15AV92, 15AV93, 15AV94, 15C9UU, 15C9UV, 15C9UW, 15C9UX, 15C9UY, 15C9V0, 15CRV8, 15CRV9, 15CRVA, 15CRVC, 15CRVD, 15CRVE, 15CRVF, 15CRVH, 15CRVJ, 15CRVK, 15CRVL, 15CRVM, 15CRVN, 15CRVP, 15D86V, 15D86W, 15D86X, 15D86Y, 15D870, 15D871, 15DRRV, 15DRRW, 15DRRX, 15DRRY, 15DRT0, 15DRT1, 15DRT2, 15E7KD, 15E7KE, 15E7KF, 15E7KH, 15E7KJ, 15E7KK, 15ET22, 15ET23, 15ET24, 15ET25, 15ET26, 15ET27, 15F8J1, 15F8J3, 15F8J4, 15H004, 15H005, 15H006, 15H007, 15H008, 15HK4E, 15HK4F, 15HK4H, 15HK4J, 15HK4K, 15HK4L, 15HK4M, 15J10N, 15J10V, 15J116, 15JPYK, 15JPYL, 15JPYM, 15JPYP, 15JPYR, 15JPYT, 15KYKE, 15KYKH, 15LTWU, 15LTX0, 15LTX1, 15M91N, 15M91Y, 15M923, 15M928, 15METN, 15METP, 15PX3M, 14WW1L, 14WW1M, 14WW1R, 14WW1T, 14X6CY, 14X6D0, 14X6D2, 14X6D3, 14X6D4, 14X6D5, 14X6D6, 14XPEJ, 14XPEK, 14XPEL, 14XPEM, 14XPEN, 14XPEP, 14XPER, 14XPET, 14XPEU, 14Y8CA, 14Y8CC, 14Y8CE, 14Y8CF, 14Y8CH, 14Y8CK, 14Y8CL, 14YLMT, 14YLMU, 14YLMV, 14YLMW, 14YLMX, 14YLMY, 14YLN0, 14YLN1, 14YLN2, 14YLN3, 14YXJ8, 14YXJ9, 14YXJA, 14YXJC, 14YXJD, 14YXJE, 14YXJF, 14YXJH, 14YXJJ, 14YXJK, 14YXJL, 150HDV, 150HDW, 150HDX, 150HDY, 150HE0, 150HE1, 150HE2, 150HE3, 150HE4, 150HE6, 151VAD, 151VAE, 151VAF, 151VAH, 151VAJ, 151VAK, 151VAL, 152F60, 152F61, 152F62, 152F63, 152F64, 152F65, 152F66, 1539X0, 1539X1, 1539X2, 1539X3, 1539X4, 1539X5, 1539X6, 1539X7, 1540JN, 1540JP, 1540JR, 1540JT, 1540JV, 1540JW, 1540JX, 1540JY, 1547CF, 1547CH, 1547CJ, 1547CK, 1547CL, 1547CM, 1547CN, 1547CP, 154YDT, 154YDU, 155HJY, 155HK0, 155HK1, 155HK2, 155HK3, 1565R6, 1565R7, 1565R8, 1565R9, 1571JA, 1571JC, 1571JD, 1571JE, 1571JF, 1571JH, 1571JJ, 1571JK, 1571JL, 1571JN, 1571JP, 157DWT, 157DWV, 157DWX, 157DWY, 157DX1, 157DX3, 157DX5, 157DX7, 157DX9, 157DXC, 157DXE, 158CY0, 158CY1, 158CY2, 158CY3, 158CY4, 158CY5, 158CY6, 158CY7, 158CY8, 158CY9, 158CYA, 158VM4, 158VM5, 158VM6, 158VM7, 158VM8, 158VM9, 158VMA, 158VMC, 1599JL, 1599JM, 1599JN, 1599JP, 1599JT, 1599JU, 1599LK, 1599LL, 1599LM, 1599LN, 1599LP, 1599LR, 1599LT, 1599LU, 1599LV, 1599LW, 159PR5, 159PR6, 159PR7, 159PR8, 159PR9, 159PRA, 159PRC, 159PRD, 159PRE, 159PRF, 159PRH, 159PRJ, 159PRK, 159PRL, 159PRM, 159PRN, 15A8HM, 15A8HN, 15A8HP, 15A8HR, 15A8HT, 15A8HU, 15A8HV, 15A8HW, 15A8HX, 15A8HY, 15A8J0, 15A8J1, 15A8J2, 15A8J3, 15A8J4, 15A8J5, 15AV97, 15AV98, 15AV99, 15AV9A, 15C9V2, 15C9V3, 15C9V4, 15C9V5, 15CRWF, 15CRWH, 15CRWJ, 15CRWK, 15CRWL, 15CRWM, 15CRWN, 15CRWP, 15CRWR, 15CRWT, 15D87D, 15D87E, 15D87F, 15D87H, 15F8J5, 15F8JA, 15F8JM, 15F8JU, 15F8JV, 15H009, 15H00A, 15H00C, 15H00D, 15H00E, 15HK4N, 15HK4P, 15HK4R, 15HK4T, 15HK4U, 15HK4V, 15HK4W, 15HK4X, 15J12H, 15J12L, 15J12P, 15J12V, 15J130, 15J134, 15J135, 15J136, 15JR05, 15JR06, 15JR07, 15JR08, 15JR09, 15JR0A, 15JR0C, 15JR0D, 15KYKN, 15KYKP, 15KYKR, 15KYKT, 15LU00, 15LU01, 15M92C, 15M92D, 15M92E, 15M92H, 15M92J, 15M92K, 15M92L, 15M92M, 15METX, 15MEU0, 15MEU1, 14WW1V, 14WW1W, 14X6D7, 14X6D8, 14X6D9, 14X6DA, 14XPEW, 14XPEX, 14XPEY, 14XPF0, 14Y8CP, 14Y8CT, 14Y8CU, 14YLPR, 14YLPT, 14YLPU, 14YXJP, 14YXJR, 14YXJT, 14YXJU, 150HE9, 150HEA, 150HEC, 150HED, 150HEE, 151VC3, 151VC4, 151VC5, 151VC6, 151VC7, 151VC8, 152F6R, 1539XF, 1539XH, 1539XJ, 1540K0, 1540K1, 1540K2, 1547CW, 1547CY, 1547D0, 1547D1, 1547D2, 155HK5, 155HK6, 1565RC, 1565RD, 1565RE, 1565RF, 1565RH, 156X16, 156X17, 156X18, 156X1D, 1571JU, 1571JV, 1571JW, 157DY8, 157DYA, 158VMF, 158VMH, 158VMJ, 158VMK, 1599M2, 1599M3, 1599M4, 1599M5, 1599M6, 1599M7, 1599M8, 1599M9, 1599MA, 159PRP, 159PRR, 15A8J6, 15A8J7, 15A8J8, 15A8J9, 15A8JA, 15A8JC, 15A8JD, 15A8JE, 15D87W, 15D87X, 15D87Y, 15D880, 15DRTC, 15DRTD, 15E7KL, 15E7KM, 15E7KN, 15E7KP, 15E7KR, 15ET28, 15ET29, 15ET2A, 15ET2C, 15ET2D, 15ET2E, 15F8KN, 15F8KP, 15H01Y, 15H020, 15H021, 15KYL9, 15KYLA, 15KYLC, 15LU03, 15LU06, 15LU07, 15MEU9, 15MEUA, 15MEUD, 14X6DC, 14X6DE, 14XPF5, 14XPF7, 14XPF8, 14Y8E9, 14Y8EA, 14YLPX, 14YLPY, 14YLR0, 14YLR1, 150HEF, 150HEH, 150HEJ, 151VC9, 151VCA, 152F6U, 1539XK, 1540K3, 155HK7, 1565RJ, 1565RK, 1565RL, 1565RM, 1571JX, 1571JY, 157M7P, 158VML, 158VMM, 1599MC, 1599MD, 1599ME, 1599MF, 159PRT, 15A8JH, 15A8JJ, 15A8JL, 15A8JM, 15A8JN, 15A8JP, 15A8JR, 15A8JT, 15D883, 15D884, 15E7KW, 15E7KX, 15E7KY, 15E7L0, 15H022, 15KYLD, 15KYLE, 15MEUH, 15PWXF, 14XPAM, 14XPAN, 14XPAP, 150HA2, 150HA3, 151V6V, 151V6W, 152F44, 1539TX, 1539U0, 155HFY, 157LXH, 157LXJ, 158VL8, 159PMX, 159PMY, 15C9NH, 15D858, 15D859, 15DRPK, 15DRPL, 15DRPM, 15DRPN, 15F8CV, 15FYY0, 15KYJ4, 15LTVJ, 14Y8AK, 14Y8AL, 14YLLU, 150HAK, 150HAT, 151V6X, 151V6Y, 152F45, 1539XU, 1539XV, 1540H0, 155HHN, 155HHP, 1565JW, 1577YD, 157LXK, 157LXM, 157LXN, 158VL9, 159PN2, 159PN5, 15CRNT, 15CRNV, 15CRNY, 15CRP1, 15D85C, 15D85D, 15DRPR, 15ET09, 15F8D3, 15FYY4, 158419, 14WW0H, 14WW0J, 14WW0K, 14WW0P, 14WW0R, 14X6A9, 14X6AA, 14X6AC, 14X6AD, 14X6AE, 14XPAV, 14XPAW, 14XPAX, 14Y8AM, 14Y8AN, 14YLLV, 14YLLW, 14YXF2, 14YXF3, 14YXF4, 150HAV, 150HAW, 150HAX, 150HAY, 150HC0, 150HC1, 150HC2, 150HC3, 150HC4, 150HC5, 150HC6, 150HC7, 150HCH, 150HCJ, 150HCK, 150HCM, 151V70, 151V71, 151V72, 151V73, 151V74, 152F48, 152F49, 152F4A, 152F4C, 152F4D, 152F4E, 152F4F, 152F4J, 152F4L, 152F4N, 152F4P, 1539U2, 1539U3, 1539U4, 1539U5, 1539U6, 1539U7, 1539U8, 1539U9, 1539UA, 154YD5, 154YD6, 154YD7, 154YD8, 154YD9, 154YDA, 154YDC, 154YDD, 155HJ7, 155HJ8, 155HJ9, 155HJA, 155HJC, 1565KA, 1565KD, 1565KF, 1565KJ, 1565KL, 1565KM, 157LXR, 157LXT, 157LXU, 157LXV, 157LXX, 157LXY, 15840X, 15840Y, 158415, 158416, 158418, 158419, 1585WV, 1585WW, 1585WX, 1585WY, 1585X0, 1585X1, 158CV1, 158CV2, 158CV3, 158CV4, 158CV5, 158CV6, 158CV7, 158CV8, 158CV9, 158CVA, 158VLC, 158VLD, 158VLE, 15AV8J, 15AV8K, 15AV8L, 15AV8M, 15AV8P, 15C9NL, 15C9NM, 15C9NN, 15C9NP, 15C9NR, 15C9NT, 15C9NU, 15C9NV, 15CRPC, 15CRPE, 15CRPF, 15CRPV, 15CRPX, 15CRPY, 15CRR0, 15D85P, 15D85R, 15D85T, 15D85U, 15D85V, 15D85W, 15D85X, 15DRPT, 15DRPU, 15DRPV, 15DRPW, 15DRPX, 15DRPY, 15E7JU, 15E7JV, 15E7JW, 15E7JY, 15E7K0, 15E7K1, 15ET0A, 15ET0C, 15ET0D, 15ET0E, 15ET0F, 15ET0H, 15ET0J, 15F8D5, 15F8D6, 15F8D7, 15F8D9, 15FYY5, 15FYY6, 15FYY7, 15FYY8, 15FYY9, 15HK3Y, 15HK40, 15HK41, 15HK42, 15HK43, 15HK44, 15J0Y6, 15J0Y7, 15JPXU, 15JPXX, 15KYJ9, 15LTW5, 15LTWC, 14WW0V, 14WW0W, 14WW0X, 14WW17, 14XPAY, 14XPC0, 14XPC1, 14XPC2, 14XPC3, 14XPC4, 14XPC5, 14Y8AP, 14Y8AR, 14Y8AT, 14Y8AU, 14Y8AV, 14YLLX, 14YLLY, 14YLM0, 14YLM1, 14YLM2, 14YLM3, 14YLM4, 14YLM5, 14YXF5, 14YXF6, 14YXF7, 14YXF8, 14YXF9, 14YXFA, 14YXFC, 150HCN, 150HCP, 150HCR, 150HCT, 150HCU, 150HCV, 150HCW, 150HCX, 150HCY, 150HD0, 151V76, 151V77, 151V7D, 151V7E, 151V7F, 154YE4, 154YE5, 154YE6, 154YE7, 154YE8, 155HJE, 155HJF, 155HJH, 155HJJ, 155HJK, 1565L8, 1565L9, 1565LA, 1565LD, 1565LF, 1565LJ, 157M7L, 15841A, 15841C, 15841D, 15841E, 15841F, 15841H, 15841J, 15841K, 1585X3, 1585X4, 1585X5, 1585X6, 1585X8, 1585X9, 1585XD, 1585XE, 158CVD, 158CVE, 158CVF, 158CVH, 158CVJ, 158CVK, 158CVL, 158CVM, 158VLF, 158VLH, 158VLJ, 15AV8R, 15AV8T, 15AV8U, 15AV8V, 15C9P7, 15C9P8, 15C9P9, 15C9PA, 15CRT1, 15CRT6, 15CRT7, 15CRTE, 15CRTJ, 15CRTL, 15CRTN, 15D861, 15D862, 15D863, 15D864, 15D865, 15D866, 15D867, 15D868, 15D869, 15DRR3, 15DRR4, 15DRR5, 15DRR6, 15DRR7, 15DRR8, 15DRR9, 15DRRA, 15DRRR, 15E7K4, 15E7K5, 15E7K6, 15E7K7, 15E7K8, 15E7K9, 15E7KC, 15ET0K, 15ET0M, 15ET0N, 15ET0P, 15ET0R, 15ET0T, 15ET0U, 15F8HN, 15F8HT, 15F8HU, 15FYYU, 15FYYV, 15HK46, 15HK47, 15HK48, 15HK49, 15JPY1, 15JPY4, 15M91F, 15M91M, 14XPCJ, 14Y8AW, 14YLMC, 150HD2, 151V7H, 1539VE, 1540H6, 155HJL, 157LY0, 157LY1, 15CRTX, 15CRU0, 15DRRP, 15F8HV, 15KYJY, 14XPCK, 14Y8AX, 14YLMD, 14YLME, 150HD3, 151V7K, 1539VY, 1540HE, 157LY4, 159PN6, 15D86N, 15D86P, 15DRRT, 15F8HW, 15FYYY, 15H000, 14X6C6, 14X6C7, 14X6C8, 14X6C9, 14X6CA, 14X6CC, 14X6CD, 14YXH8, 14YXH9, 14YXHA, 14YXHC, 14YXHD, 14YXHE, 14YXHF, 151V89, 151V8A, 151V8C, 151V8D, 151V8E, 1540HJ, 1540HK, 1540HL, 1540HM, 1540HN, 1540HP, 1540HR, 1540HT, 1540HU, 1540HV, 1565NE, 1565NH, 1565NK, 1565NL, 1565NN, 1565NR, 1583W8, 1583W9, 1583WC, 159PNE, 159PNF, 159PNH, 15CRUV, 15CRUW, 15CRUX, 15CRUY, 15CRV0, 15ET1K, 15ET1L, 15ET1N, 15J0YX, 15J100, 15J101, 15J107, 15J108, 15J109, 15LTY8, 15LTYA, 14X6CE, 14X6CF, 14X6CH, 14X6CJ, 14X6CK, 14X6CL, 14X6CM, 14YXHJ, 14YXHK, 14YXHL, 14YXHM, 14YXHN, 14YXHP, 14YXHR, 14YXHT, 151V8F, 151V8H, 151V8J, 151V8K, 151V8L, 151V8M, 1540HW, 1540HX, 1540HY, 1540J0, 1540J1, 1540J2, 1540J3, 1540J4, 1565P7, 1565P9, 1565PA, 1565PD, 1565PF, 1565PH, 1565PK, 1565PL, 15840V, 15840W, 159PP0, 159PP1, 159PP2, 15CRV2, 15CRV4, 15CRV5, 15CRV6, 15CRV7, 15ET1R, 15ET1T, 15ET1U, 15ET1V, 15ET1W, 15ET1X, 15ET1Y, 15ET20, 15ET21, 15J10A, 15J10C, 15J10D, 15J10J, 15LTYJ, 15LTYK, 15LTYL, 15LTYM, 150HD4, 1539WH, 1565PT, 15AV8X, 15C9PX, 14XPCM, 14Y8AY, 14Y8C0, 14YXHV, 14YXHW, 150HD5, 150HD6, 1539WJ, 1539WK, 1539WL, 1540J8, 155HJM, 1565PV, 1565PW, 157LY6, 158VLU, 15C9R5, 15D86R, 15D86T, 15F8HX, 15H001, 15H002, 15JPYC, 15JPYE, 14XPE8, 14Y8C1, 150HD7, 151V8W, 1539WM, 1539WN, 155HJN, 157LY8, 15F8HY, 1545JX, 1565JK, 1585WM, 1585WR, 15AV9F, 15D885, 15H023, 15H024, 1585WT, 151PR4.
Devices named in this recall
19 devices in the FDA device registry whose identifier appears in the recall's code information.
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Alcon 8065752917Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall94 reports
- Alcon 8065752918Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall41 reports
- Alcon 8065752217Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall31 reports
- Alcon 00380650001344Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall30 reports
- Alcon 8065752216Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall27 reports
- Alcon 00380650001337Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall23 reports
- Alcon 8065752203Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall17 reports
- Alcon 8065752193Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall14 reports
- Alcon 8065752202Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall10 reports
- Alcon 8065752192Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall8 reports
- Alcon 00380650001382Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall7 reports
- CENTURION® INTREPID® 8065753134Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall4 reports
- CENTURION®, INTREPID® 8065753135Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall4 reports
- Alcon 00380650001399Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall3 reports
- Alcon 8065752245Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall0 reports
- CENTURION® INTREPID® 8065753136Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall0 reports
- CENTURION®, INTREPID® 8065753159Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92698).
- Z-2406-2023 · Open, ClassifiedConstellation Procedure Pak
- Z-2408-2023 · Open, ClassifiedInfinity FMS Pak
- Z-2409-2023 · Open, ClassifiedLegion FMS Pak
- Z-2410-2023 · Open, ClassifiedPhaco Standalone Tips
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page