Skip to content

Olympus Tracheal Intubation Fiberscope, Models LF-P & LF-V recallZ-2501-2023 · Olympus Corporation of the Americas · initiated Jul 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

FDA root cause category: Labeling Change Control

What the company is doing

Consignees were mailed an Urgent Medical Device Correction notification dated 7/27/2023 via UPS. The notification asks that consignees inspect their inventory for recalled devices and check clinical areas to see if any remain in use. Consignees are to ensure that all personnel, including clinical staff, are completely knowledgeable of this recall notification and addendum, which instructs users to slide the endotracheal tube along the entire length of the tracheal intubation scope insertion section to confirm compatibility before using on a patient. Consignees are asked to complete and return the provided response form to Sedgwick via email at [email protected] or by fax at 866-552-4917. Any questions about the acknowledgement form may be directed by phone to 866-737-4762. If product has been further distributed, please forward the notification to those secondary consignees and maintain records of your documentation process. Consignees are to report any complaints to Olympus Technical Assistance Center at 1-800-848-9024 (Option 1) and FDA Medwatch. Any further questions are to be directed to Tara Safi by email at [email protected].

Recall details

Recall number
Z-2501-2023
Status
Open, Classified
Initiated by the company
Jul 27, 2023
Posted by the FDA
Aug 31, 2023
Recalling firm
Olympus Corporation of the Americas
Quantity
2127 units
Product code
EOQ
FDA submissions
K221631
Distribution
US Nationwide distribution.

Product description

Olympus Tracheal Intubation Fiberscope, Models LF-P & LF-V.

Code information (models, lots, serial numbers)

UDI-DI: 04953170051098, 04953170452932, & 04953170156250; All Serial Numbers.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92779).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page