Skip to content

nanoDot D2DXS, Model Numbers recallZ-2548-2023 · Landauer · initiated Jul 12, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Landauer or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

FDA root cause category: Under Investigation by firm

What the company is doing

Landauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.

Recall details

Recall number
Z-2548-2023
Status
Completed
Initiated by the company
Jul 12, 2023
Posted by the FDA
Sep 12, 2023
Recalling firm
Landauer
Quantity
68007 units
Product code
IYE
FDA submissions
K192196
Distribution
Worldwide

Product description

nanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers

Code information (models, lots, serial numbers)

UDI/DI 0860003399903, all batch numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92647).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page