QC (Cs-137) nanoDot D2DNS, Model Numbers recallZ-2560-2023 · Landauer · initiated Jul 12, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Landauer or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
FDA root cause category: Under Investigation by firm
What the company is doing
Landauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.
Recall details
- Recall number
- Z-2560-2023
- Status
- Completed
- Initiated by the company
- Jul 12, 2023
- Posted by the FDA
- Sep 12, 2023
- Recalling firm
- Landauer
- Quantity
- 1440 units
- Product code
- IYE
- FDA submissions
- K192196
- Distribution
- Worldwide
Product description
QC (Cs-137) nanoDot D2DNS, Model Numbers: a) 04225-000 (EMEA number VINLNAN005); b) 04225-KIT; c) 04225-SET; radiation monitoring dosimeter used with the microSTAR readers
Code information (models, lots, serial numbers)
UDI/DI 0860003399903, all batch numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LANDAUER microSTAR® ii Dosimetry SystemLandauer, Inc. · Accelerator, Linear, Medical38 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92647).
- Z-2538-2023 · CompletednanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2539-2023 · CompletednanoDot D2DNS, for custom calibrate, OTO, Model Numbers 03034-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2540-2023 · CompletednanoDot D2DNN, Model Numbers: a) 03051-1MO; b) 03051-3MO; c) 03051-OTO; d) 03051-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2541-2023 · CompletednanoDot D2DNN, Model Numbers: a) 03053-1MO; b) 03053-1WK ; c) 03053-3MO; d) 03053-OTO; e) 03053-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2542-2023 · CompletednanoDot D2DNS, Model Numbers: a) 03055-1MO; b) 03055-3MO; c) 03055-OTO; d) 03055-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2543-2023 · CompletednanoDot D2DNS, Model Numbers: a) 03056-000 (EMEA number VINLNAN002); b) 03056-1MO; c) 03056-3MO; d) 03056-KIT; radiation monitoring dosimeter used with the microSTAR readers
- Z-2544-2023 · CompletednanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2545-2023 · CompletednanoDot D2DXN, Model Numbers: a) 03060-1MO; b) 03060-3MO; c) 03060-OTO; d) 03060-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2546-2023 · CompletednanoDot D2DXN, Model Numbers: a) 03061-1MO; b) 03061-3MO; c) 03061-OTO; d) 03061-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2547-2023 · CompletednanoDot D2DXS, Model Numbers: a) 03062-1MO; b) 03062-3MO; c) 03062-OTO; d) 03062-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2548-2023 · CompletednanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2549-2023 · CompletedCalibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2550-2023 · CompletedCalibrate (Cs-137) screened nanoDot D2DNS, Model Number 03065-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2551-2023 · CompletedQC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2552-2023 · CompletedQC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers
- Z-2553-2023 · CompletednanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
- Z-2554-2023 · CompletednanoDot A - item 1 of 3 for 03500-000, Model Number 03501-000; radiation monitoring dosimeter used with the microSTAR readers
- Z-2555-2023 · CompletednanoDot B - item 2 of 3 for 03500-000, Model Number 03502-000; radiation monitoring dosimeter used with the microSTAR readers
- Z-2556-2023 · CompletednanoDot C - item 3 of 3 for 03500-000, Model Number 03503-000; radiation monitoring dosimeter used with the microSTAR readers
- Z-2557-2023 · CompletedCalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR reade…
- Z-2558-2023 · CompletedQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 04218-000 (EMEA number VINLNAN004); b) 04218-KIT; c) 04218-SET; radiation monitoring dosimeter used with the microSTAR readers
- Z-2559-2023 · CompletedCalibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 04224-000 (EMEA number VINLNAN006); b) 04224-KIT; c) 04224-SET; radiation monitoring dosimeter used with the microSTAR reade…
- Z-2561-2023 · CompletednanoDot D2DNN , Model Numbers: a) 04293-000; b) 04293-KIT; radiation monitoring dosimeter used with the microSTAR readers
- Z-2562-2023 · CompletednanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the microSTAR readers
- Z-2563-2023 · CompletednanoDot D2DNN, Model Numbers: a) 18105-000; b) 18105-KIT; radiation monitoring dosimeter used with the microSTAR readers
- Z-2564-2023 · CompletedCalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with the microSTAR readers
- Z-2565-2023 · CompletedCalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18121-000; b)18121-KIT; c) 18121-SET; radiation monitoring dosimeter used with the microSTAR readers
- Z-2566-2023 · CompletedCalibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 18125-000; b) 18125-KIT; c) 18125-SET; radiation monitoring dosimeter used with the microSTAR readers
- Z-2567-2023 · CompletedQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18130-000; b) 18130-KIT; c) 18130-SET; radiation monitoring dosimeter used with the microSTAR readers
- Z-2568-2023 · CompletedQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18131-000; b) 18131-KIT; c) 18131-SET; radiation monitoring dosimeter used with the microSTAR readers
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page