HAMILTON-T1 recallZ-2526-2023 · Hamilton Medical, Inc. · initiated Jul 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.

Recall details

Recall number
Z-2526-2023
Status
Open, Classified
Initiated by the company
Jul 26, 2023
Posted by the FDA
Sep 20, 2023
Recalling firm
Hamilton Medical, Inc.
Quantity
76 units
Product code
CBK
FDA submissions
K181216
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided

Product description

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

Code information (models, lots, serial numbers)

UDI: 07630002801850 & 07630002813532 / Affected Serial Numbers: 15929 17600 17661 17677 17698 17699 17741 17850 17877 17913 17925 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18166 18183 18189 18191 18199 18224 18244 18251 18255 18256 18260 18263 18272 18273 18276 18288 18289 18305 18308 18333 18340 18349 18360 18363 18377 18378 18383 18414 18415 18419 18441 18451 18458 18461 18502 18537 18544 18550 18778 18784 19237 19290 19291 19292 19293 19294 19298 19304 19305 19470 19483 19490 26117

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92873).