HAMILTON-MR1 recallZ-2528-2023 · Hamilton Medical, Inc. · initiated Jul 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.

Recall details

Recall number
Z-2528-2023
Status
Open, Classified
Initiated by the company
Jul 26, 2023
Posted by the FDA
Sep 20, 2023
Recalling firm
Hamilton Medical, Inc.
Quantity
12 units
Product code
CBK
FDA submissions
K153046
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided

Product description

Hamilton-MR1, REF: 161010, Intensive Care Ventilator

Code information (models, lots, serial numbers)

UDI: 07630002800761 & 07630002813556 / Affected Serial Numbers: 3422 3423 3424 3425 3426 3427 3428 3429 3469 3503 3510 3800

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92873).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page