Palindrome recallZ-2632-2023 · Covidien LP · initiated Aug 22, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT: MEDICAL DEVICE RECALL notification letter was sent to customers on 8/22/23. Required Actions: 1. Immediately quarantine and discontinue use of all unused affected lots of the Palindrome" 14.5 Fr/Ch (4.8 mm) x 23 cm Chronic Catheter Kits. Please note: This recall does not include Palindrome Precision Catheter Kits. 2. Complete the online Customer Confirmation Form using the QR code located at the top of the first page of this letter or by using the link below. We request your response within 3 business days of receiving this notice. Replying promptly will confirm your receipt of this notification and prevent you from receiving additional notices. Customer Confirmation Link: https://www.novasyte.com/medtronic/fa1355 Note: Please complete the online Customer Confirmation Form even if you DO NOT have affected product. 3. If you have affected product, it must be returned for processing. Upon completion of the Customer Confirmation Form, Customer Service will contact you with an RGA number to return all unused affected Palindrome" 14.5 Fr/Ch (4.8 mm) x 23 cm Chronic Catheter Kits from your inventory to Medtronic. 4. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 5. This notice should be passed on to those who need to be aware within your organization or to any organization including, but not limited to, nephrologists, intensivists, implanting and managing physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Medtronic has partnered with Novasyte, an IQVIA company, to assist in this recall. For assistance regarding online response processing or product return for this recall, please contact Novasyte using the information below: " Phone number: (855) 566-3445 " Email: [email protected] We regret any inconvenience this may cause. We a
Recall details
- Recall number
- Z-2632-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 22, 2023
- Posted by the FDA
- Sep 25, 2023
- Recalling firm
- Covidien LP
- Quantity
- 3050 devices
- Product code
- MSD
- FDA submissions
- K111372
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, China, Czech Republic, Finland, France, Germany, Italy, Netherlands, Norway, Poland, Portugal, Puerto Rico, Spain, Thailand, United Kingdom.
Product description
Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
Code information (models, lots, serial numbers)
UDI-DI: 10884521013162, 20884521013169; Lot Numbers: 2221700131, 2224200233, 2230400271, 2230400272
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
Product lines
Other devices under the same FDA submission
42 devices cleared under K111372 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- MAHURKAR 8888145001Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall45 reports
- Palindrome 8888145016Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death38 reports
- MAHURKAR 8888145251Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death33 reports
- Palindrome 8888145017Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall22 reports
- Palindrome 8888145058Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall15 reports
- Palindrome 8888145018Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall12 reports
- MAHURKAR 8888145002Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall12 reports
- Palindrome 8888145040Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death12 reports
- MAHURKAR 8888123435Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall10 reports
- MAHURKAR 8888145252Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall9 reports
- Palindrome 8888145039Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.