Tempus Pro recallZ-2625-2023 · Remote Diagnostic Technologies Ltd. · initiated Aug 24, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Remote Diagnostic Technologies Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.
FDA root cause category: Device Design
What the company is doing
On 8/24/23 Correction Notices were mailed to customers and distributors who were asked to do the following: 1) If monitor has Trizeps-7 Hardware, remove the laryngoscope from service with this monitor. Users must use an alternative laryngoscope not connected to the Tempus Pro Monitor to manage the patient s airway to avoid interruption in patient care. The Tempus Pro Monitor can remain in service if the Tempus Pro Video Laryngoscope is not connected to the monitor. 2) If monitor has Trizeps-6 Hardware, the laryngoscope can continue to be used with the monitor. 3) Post this Urgent Medical Device Correction letter on or near the affected device. 4) Pass this notice on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred (if appropriate). 5) Complete and return the Urgent Medical Device Correction response form. Distributors were asked to take the following additional actions: If you have any Tempus Pro Video Laryngoscopes or any Tempus Pro Monitors in stock that have not been distributed, do not distribute them. Modify the response form to include your firm's contact information and then send the response form along with the customer correction notice to customers. If you need any further information or support contact the firm at (800) 722-9377 or email [email protected]. On 11/6/23, updated correction notices were distributed adding two additional models that could become affected if their hardware is updated. Additionally, the notices asked customers to install software (V07.34) on Tempus Pro Monitors according to installation instructions.
Recall details
- Recall number
- Z-2625-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 24, 2023
- Posted by the FDA
- Sep 21, 2023
- Recalling firm
- Remote Diagnostic Technologies Ltd.
- Quantity
- 3,524
- Product code
- MHX
- FDA submissions
- K201746
- Distribution
- US:AZ, AL, CA, CO, DC, DE, FL , GA, HI, IA, ID, IL, IN , KS, KY, MA , MD , ME, MI, MN, MO, MS, MT, NC , ND, NE, NJ, NM , NV, NY, OH, OK, OR, PA, TN, TX, UT, VA , WA, WI, WV, WY. OUS: AE, AT, AU, BE, BN, BO, CA, CH, CZ, DE, DK, ES, FR, GB, HK, IE, IL, IS, IT, KY, LU, MY, NL, NO, OM, PT, RO, SA, SE, SG, SI, TH, YT, ZA
Product description
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044
Code information (models, lots, serial numbers)
Monitors with hardware version: Trizeps-7, and software versions v07.26 through v07.30. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
- Tempus Pro 00-1026-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls1 death324 reports
- Tempus Pro 00-1026-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls233 reports
- Tempus Pro 00-1007-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls114 reports
- Tempus Pro 00-1004-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)6 recalls36 reports
- Tempus Pro 00-1007-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls64 reports
- Tempus Pro 00-1004-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls23 reports
- Tempus Pro 00-1007Remote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls2 reports
- Tempus Pro 00-1004Remote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page