Ivenix, Inc. recallZ-2635-2023 · Fresenius Kabi USA, LLC · initiated Apr 22, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

FDA root cause category: Device Design

What the company is doing

Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is unaffected and still within specifications, and the clinical practice of monitoring patients while administering high alert drugs mitigates the risk. Ivenix is working on a software update to correct this condition. It will be available in approximately 60 days. We will inform you when that update is ready to install. If you have additional questions, please contact your Ivenix representative or support below. Customer support - email Customer support - phone [email protected] I-978-775-8TAA

Recall details

Recall number
Z-2635-2023
Status
Open, Classified
Initiated by the company
Apr 22, 2022
Posted by the FDA
Sep 25, 2023
Recalling firm
Fresenius Kabi USA, LLC
Quantity
1344 pumps
Product code
FRN
FDA submissions
K183311
Distribution
US Nationwide distribution in the states of NJ, WI.

Product description

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Code information (models, lots, serial numbers)

(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page