Ivenix, Inc. recallZ-2635-2023 · Fresenius Kabi USA, LLC · initiated Apr 22, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
FDA root cause category: Device Design
What the company is doing
Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is unaffected and still within specifications, and the clinical practice of monitoring patients while administering high alert drugs mitigates the risk. Ivenix is working on a software update to correct this condition. It will be available in approximately 60 days. We will inform you when that update is ready to install. If you have additional questions, please contact your Ivenix representative or support below. Customer support - email Customer support - phone [email protected] I-978-775-8TAA
Recall details
- Recall number
- Z-2635-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 22, 2022
- Posted by the FDA
- Sep 25, 2023
- Recalling firm
- Fresenius Kabi USA, LLC
- Quantity
- 1344 pumps
- Product code
- FRN
- FDA submissions
- K183311
- Distribution
- US Nationwide distribution in the states of NJ, WI.
Product description
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Code information (models, lots, serial numbers)
(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ivenix, Inc. 2.0Fresenius Kabi AG · Pump, Infusion10 recalls0 reports
Product lines
Other devices under the same FDA submission
4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Ivenix Infusion System LVP-0004Fresenius Kabi AG · Pump, Infusion7 recalls2 deaths4,328 reports
- Ivenix Infusion System 030-0028-08Fresenius Kabi AG · Pump, Infusion7 reports
- Ivenix Infusion System 030-0027-06Fresenius Kabi AG · Infusion Safety Management Software3 reports
- Ivenix Infusion System 030-0028-07Fresenius Kabi AG · Pump, Infusion2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.