Skip to content

Incisive CT, software version 5.0 recallZ-2520-2023 · Philips Healthcare · initiated Aug 2, 2023

Class II Open, Classified All recalls posted in September 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Philips will inform customers of the issue via an Electronic Product Radiation Defect notification. Philips will schedule an appointment with customers to install the software update.

Recall details

Recall number
Z-2520-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Aug 2, 2023
Posted by the FDA
Sep 30, 2023
Recalling firm
Philips Healthcare
Quantity
140 systems
Product code
JAK
FDA submissions
K211168, K212441, K180015
Distribution
US Nationwide Distribution

Product description

Incisive CT, software version 5.0

Code information (models, lots, serial numbers)

software version 5.0

Other devices under the same FDA submission

1 device cleared under K211168, K212441, K180015 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page