SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO recallZ-0126-2024 · SynCardia Systems LLC · initiated Aug 1, 2023
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from SynCardia Systems LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Artificial hearts contain a epoxy resin that has not been reviewed by the FDA for safety and efficacy.
FDA root cause category: Other
What the company is doing
On August 22, 2023, SynCardia issued an Urgent Medical Device Correction Notification to all affected consignees via Email. SynCardia asked consignees to take the following actions: 1. You may continue to use the device while the change undergoes FDA review for safety and effectiveness. 2.Ensure that all medical professionals in your institution are informed of this Urgent Medical Device Correction. 3. Please complete the Customer Acknowledgement Form and return it via email to SynCardia Systems, LLC, within five business days of receipt of this letter. 4.For patients with implanted devices affected by this correction, monitor patients closely and take appropriate actions you believe necessary if any issues arise. 5. Adverse reactions or quality problems experienced with this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online: www.fda.gov/medwatch/report.htm " Regular Mail or Fax: Download the form from www.fda.gov/medwatch/getforms.htm or call 1-800-322-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Recall details
- Recall number
- Z-0126-2024
- Status
- Completed
- Initiated by the company
- Aug 1, 2023
- Posted by the FDA
- Oct 20, 2023
- Recalling firm
- SynCardia Systems LLC
- Quantity
- 114 units
- Product code
- LOZ
- FDA submissions
- P030011
- Distribution
- US: FL, WI, AZ, TX, CA, OH, NC, IN, OK, IL, VA, NY, NE, PA, WA, OUS: Saudi Arabia, Austria, Italy, Macedonia, Turkey, Germany, France, Kuwait, Serbia, United Kingdom, Canada, Lithuania, Slovenia Australia
Product description
SynCardia 50cc TAH-t, Clinical, REF 570500, Sterile EO
Code information (models, lots, serial numbers)
All Lots/ UDI: 00858000003053
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SynCardia 50cc TAH-t, Clinical 570500-001Syncardia Systems, LLC · Artificial Heart1 recall5 deaths5 reports
Other devices under the same FDA submission
8 devices cleared under P030011 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SynCardia Freedom® Driver 595000-001Syncardia Systems, LLC · Artificial Heart1 recall7 deaths809 reports
- SynCardia Companion 2 Driver 397002-001Syncardia Systems, LLC · Artificial Heart2 deaths414 reports
- SynCardia temporary Total Artificial Heart System 500101-001Syncardia Systems, LLC · Artificial Heart141 deaths215 reports
- SynCardia Freedom® Onboard Battery 295025-001Syncardia Systems, LLC · Artificial Heart62 reports
- SynCardia Companion Hospital Cart 397003-001Syncardia Systems, LLC · Artificial Heart9 reports
- SynCardia Companion Rechargeable LI-ION Battery 293001-001Syncardia Systems, LLC · Artificial Heart8 reports
- SynCardia Companion Driver Caddy 397001-001Syncardia Systems, LLC · Artificial Heart3 reports
- SynCardia Hand Pump 397004-001Syncardia Systems, LLC · Artificial Heart2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.