Alinity m recallZ-0407-2024 · Abbott Molecular, Inc. · initiated Oct 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Molecular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

FDA root cause category: Other

What the company is doing

Abbott issued an Urgent Field Safety Notice to its consignees on 10/30/2023 via letter delivered using FedEx priority overnight express. The notice explained the problem, the potential impact/risk to patient, and requested the following actions: -discard all affected product -notify all those to which the products were transferred -follow internal laboratory procedures in the case of suspected incorrect results Credit or replacement will be issued for units discarded/destroyed. Abbott expanded the recall and notified consignees on 07/23/2024 via letter. The letter instructed consignees of an additional lot number included in the recall, but otherwise contained the same instructions.

Recall details

Recall number
Z-0407-2024
Status
Open, Classified
Initiated by the company
Oct 30, 2023
Posted by the FDA
Nov 29, 2023
Recalling firm
Abbott Molecular, Inc.
Quantity
1,881 units
Product code
MZP
FDA submissions
P190025
Distribution
Worldwide

Product description

Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for use with the automated Alinity m System to detect and quantitate hepatitis C virus (HCV) RNA in human serum or plasma.

Code information (models, lots, serial numbers)

a) 08N50-095: UDI/DI 00884999047648, Lot Numbers: 381582, 382448, 383204; b) 08N50-090: UDI/DI 00884999047761, Lot Numbers: 383203, 385720, 381581, 382447