Skip to content

Alinity m 08N50-095Recalls & reported problems

Abbott Molecular Inc. · Alinity m HCV AMP Kit · Class 2

All models of Alinity m (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884999047761), received between Nov 7, 2023 and Nov 7, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Positive Result” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 30, 2023Open, Classified

    Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

    Z-0407-2024 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • False Positive Result1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2023False Positive Result; Unable to Obtain Readings; False Positive Result; Unable to Obtain Readings

    THROUGH COMPLAINT INVESTIGATION, ABBOTT HAS IDENTIFIED AN INCREASE IN POTENTIAL FOR REACTIVE NEGATIVE CONTROLS (NC) AS WELL AS SEPARATE INSTANCES OF FALSE POSITIVE RESULTS IN PATIENT SAMPLES WHEN USING SPECIFIC LOTS OF ALINITY M HCV AMP KIT. IT WAS DISCOVERED THAT THESE SPECIFIC LOTS OF ALINITY M HCV AMP KIT HAVE LOWER THAN THE EXPECTED RANGE OF HCV REFERENCE DYE FLUORESCENCE. A PRODUCT DEFICIENCY WAS IDENTIFIED FOR THESE SPECIFIC LOTS OF ALINITY M HCV ASSAY. THE POTENTIAL HAZARDS ASSOCIATED WITH THIS ISSUE ARE INCORRECT RESULTS (FALSE POSITIVE) AND DELAY OF RESULTS (REACTIVE NC REQUIRING A RE-TEST). FIELD ACTION FA-AM-OCT2023-293, WAS TAKEN TO ADDRESS THIS ISSUE. IT WAS DEEMED REPORTABLE PER 21CFR806 AND THEREFORE WILL BE REPORTED UNDER 21CFR803.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18089017 ↗

Other Assay, hybridization and/or nucleic acid amplification for detection of hepatitis c rna,hepatitis c virus devices

All Assay, hybridization and/or nucleic acid amplification for detection of hepatitis c rna,hepatitis c virus devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page